Shoulder Range of Motion Measurement Using Inertial Measurement Unit-Validation with a Robot Arm

被引:5
作者
Bialecka, Martyna [1 ,2 ]
Gruszczynski, Kacper [3 ]
Cisowski, Pawel [1 ,4 ]
Kaszynski, Jakub [1 ]
Baka, Cezary [1 ]
Lubiatowski, Przemyslaw [1 ,5 ]
机构
[1] Rehasport Clin, Gorecka 30, PL-60201 Poznan, Poland
[2] Poznan Univ Tech, Inst Appl Mech, Fac Mech Engn, PL-60965 Poznan, Poland
[3] RSQ Technol Co, 27 Grudnia 3, PL-61737 Poznan, Poland
[4] Poznan Univ Med Sci, Spine Disorders & Pediat Orthoped Dept, PL-61545 Poznan, Poland
[5] Poznan Univ Med Sci, Orthopaed Traumatol & Hand Surg Dept, 28 Czerwca 1956, 135-147, PL-61545 Poznan, Poland
关键词
inertial measurement unit; sensor accuracy; validation; shoulder; range of motion;
D O I
10.3390/s23125364
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The invention of inertial measurement units allowed the construction of sensors suitable for human motion tracking that are more affordable than expensive optical motion capture systems, but there are a few factors influencing their accuracy, such as the calibration methods and the fusion algorithms used to translate sensor readings into angles. The main purpose of this study was to test the accuracy of a single RSQ Motion sensor in comparison to a highly precise industrial robot. The secondary objectives were to test how the type of sensor calibration affects its accuracy and whether the time and magnitude of the tested angle have an impact on the sensor's accuracy. We performed sensor tests for nine repetitions of nine static angles made by the robot arm in eleven series. The chosen robot movements mimicked shoulder movements in a range of motion test (flexion, abduction, and rotation). The RSQ Motion sensor appeared to be very accurate, with a root-mean-square error below 0.15 & DEG;. Furthermore, we found a moderate-to-strong correlation between the sensor error and the magnitude of the measured angle but only for the sensor calibrated with the gyroscope and accelerometer readings. Although the high accuracy of the RSQ Motion sensors was demonstrated in this paper, they require further study on human subjects and comparisons to the other devices known as the gold standards in orthopedics.
引用
收藏
页数:12
相关论文
共 19 条
[1]   Inertial Measurement Units for Clinical Movement Analysis: Reliability and Concurrent Validity [J].
Al-Amri, Mohammad ;
Nicholas, Kevin ;
Button, Kate ;
Sparkes, Valerie ;
Sheeran, Liba ;
Davies, Jennifer L. .
SENSORS, 2018, 18 (03)
[2]  
[Anonymous], NOR ULT MOT WEB PAG
[3]  
[Anonymous], XSENS MTW AW WEB PAG
[4]  
[Anonymous], DAT SHEET KUKA KR 3
[5]   Wearable systems for shoulder kinematics assessment: a systematic review [J].
Carnevale, Arianna ;
Longo, Umile Giuseppe ;
Schena, Emiliano ;
Massaroni, Carlo ;
Lo Presti, Daniela ;
Berton, Alessandra ;
Candela, Vincenzo ;
Denaro, Vincenzo .
BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (01)
[6]  
Cutti Andrea Giovanni, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P5912
[7]   Ambulatory measurement of shoulder and elbow kinematics through inertial and magnetic sensors [J].
Cutti, Andrea Giovanni ;
Giovanardi, Andrea ;
Rocchi, Laura ;
Davalli, Angelo ;
Sacchetti, Rinaldo .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2008, 46 (02) :169-178
[8]   9-DOF IMU-Based Attitude and Heading Estimation Using an Extended Kalman Filter with Bias Consideration [J].
Farahan, Sajjad Boorghan ;
Machado, Jose J. M. ;
de Almeida, Fernando Gomes ;
Tavares, Joao Manuel R. S. .
SENSORS, 2022, 22 (09)
[9]   Survey of Motion Tracking Methods Based on Inertial Sensors: A Focus on Upper Limb Human Motion [J].
Filippeschi, Alessandro ;
Schmitz, Norbert ;
Miezal, Markus ;
Bleser, Gabriele ;
Ruffaldi, Emanuele ;
Stricker, Didier .
SENSORS, 2017, 17 (06)
[10]   Proprioception after rehabilitation and reconstruction in knees with deficiency of the anterior cruciate ligament - A prospective, longitudinal study [J].
Fremerey, RW ;
Lobenhoffer, P ;
Zeichen, J ;
Skutek, M ;
Bosch, U ;
Tscherne, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (06) :801-806