Hemodynamic performance evaluation of neonatal ECMO double lumen cannula using fluid-structure interaction

被引:2
|
作者
Ahmad, Faiq [1 ]
Cheema, Taqi Ahmad [1 ]
Rehman, Khawar [2 ,3 ]
Ullah, Minhaj [1 ]
Jamil, Muhammad [4 ,5 ]
Park, Cheol Woo [6 ,7 ]
机构
[1] GIK Inst Engn Sci & Technol, Fac Mech Engn, Topi 23460, Pakistan
[2] Hanyang Univ, Dept Civil & Environm Engn, Seoul 04763, South Korea
[3] GIK Inst Engn Sci & Technol, Dept Civil Engn, Topi 23460, Pakistan
[4] Anal Grp, Starfish Med, 455 Boleskine Rd, Victoria, BC, Canada
[5] KoC Univ, Dept Mech Engn, Sariyar, TR-34450 Istanbul, Turkiye
[6] Kyungpook Natl Univ, Sch Mech Engn, 80 Daehak Ro, Daegu 41566, South Korea
[7] Kyungpook Natl Univ, Sch Mech Engn, 80 Daehak Ro, Daegu 41566, South Korea
基金
新加坡国家研究基金会;
关键词
blood damage; double lumen cannula (DLC); extra corporeal membrane oxygenation (ECMO); fluid-structure interaction (FSI); hemodynamic performance; BLOOD DAMAGE; FLOW; OXYGENATION; DESIGN; IMPACT;
D O I
10.1002/cnm.3706
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Extra corporeal membrane oxygenation (ECMO) is an artificial oxygenation facility, employed in situations of cardio-pulmonary failure. Some diseases i.e., acute respiratory distress syndrome, pulmonary hypertension, corona virus disease (COVID-19) etc. affect oxygenation performance of the lungs thus requiring the need of artificial oxygenation. Critical care teams used ECMO technique during the COVID-19 pandemic to support the heart and lungs of COVID-19 patients who had an acute respiratory or cardiac failure. Double Lumen Cannula (DLC) is one of the most critical components of ECMO as it resides inside the patient and, connects patient with external oxygenation circuit. DLC facilitates delivery and drainage of blood from the patient's body. DLC is characterized by delicate balance of internal and external flows inside a limited space of the right atrium (RA). An optimal performance of the DLC necessitates structural stability under biological and hemodynamic loads, a fact that has been overlooked by previously published studies. In the past, many researchers experimentally and computationally investigated the hemodynamic performance of DLC by employing Eulerian approach, which evaluate instantaneous blood damage without considering blood shear exposure history (qualitative assessment only). The present study is an attempt to address the aforementioned limitations of the previous studies by employing Lagrangian (quantitative assessment) and incorporating the effect of fluid-structure interaction (FSI) to study the hemodynamic performance of neonatal DLC. The study was performed by solving three-dimensional continuity, momentum, and structural mechanics equation(s) by numerical methods for the blood flow through neonatal DLC. A two-way coupled FSI analysis was performed to analyze the effect of DLC structural deformation on its hemodynamic performance. Results show that the return lumen was the most critical section with maximum pressure drop, velocity, shear stresses, and blood damage. Recirculation and residence time of blood in the right atrium (RA) increases with increasing blood flow rates. Considering the structural deformation has led to higher blood damage inside the DLC-atrium system. Maximum Von-Mises stress was present on the side edges of the return lumen that showed direct proportionality with the blood flow rate.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] ASSESSMENT RESULTS OF FLUID-STRUCTURE INTERACTION NUMERICAL SIMULATION USING FUZZY LOGIC
    Markovic, Zoran J.
    Stupar, Slobodan N.
    Dinulovic, Mirko R.
    Pekovic, Ognjen M.
    Stefanovic, Predrag Lj.
    Cvetinovic, Dejan B.
    THERMAL SCIENCE, 2016, 20 : S235 - S250
  • [22] A Wind Tunnel Setup for Fluid-Structure Interaction Measurements Using Optical Methods
    Nietiedt, Simon
    Wester, Tom T. B.
    Langidis, Apostolos
    Kroeger, Lars
    Rofallski, Robin
    Goering, Martina
    Kuhn, Martin
    Guelker, Gerd
    Luhmann, Thomas
    SENSORS, 2022, 22 (13)
  • [23] Computational Modelling of Bileaflet Mechanical Valves Using Fluid-Structure Interaction Approach
    Yeh, Han Hung
    Grecov, Dana
    Karri, Satya
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2014, 34 (05) : 482 - 486
  • [24] Study of velocity effects on parachute inflation performance based on fluid-structure interaction method
    Han Cheng
    Xin-hua Zhang
    Li Yu
    Meng Chen
    Applied Mathematics and Mechanics, 2014, 35 : 1177 - 1188
  • [25] FLUID-STRUCTURE INTERACTION ANALYSIS OF A PERISTALTIC PUMP BY USING NUMERICAL AND EXPERIMENTAL METHODS
    Li, Qingye
    Zong, Chaoyong
    Zhang, Jianhua
    Zhang, Yanfeng
    Songa, Xueguan
    PROCEEDINGS OF ASME 2022 PRESSURE VESSELS AND PIPING CONFERENCE, PVP2022, VOL 3, 2022,
  • [26] Investigations on aerodynamic and mechanical performance of a transonic blade based on a fluid-structure interaction method
    Wang S.
    Li S.
    Song X.
    Li, Shaobin (lee_shaobin@buaa.edu.cn), 2016, Chinese Vibration Engineering Society (35): : 104 - 110
  • [27] Study of adaptive blades in extreme environment using fluid-structure interaction method
    Miao, Weipao
    Li, Chun
    Wang, Yuanbo
    Xiang, Bin
    Liu, Qingsong
    Deng, Yunhe
    JOURNAL OF FLUIDS AND STRUCTURES, 2019, 91
  • [28] Fluid-structure interaction computations for geometrically resolved rotor simulations using CFD
    Heinz, Joachim C.
    Sorensen, Niels N.
    Zahle, Frederik
    WIND ENERGY, 2016, 19 (12) : 2205 - 2221
  • [29] The dynamic performance analysis of the reciprocating continuous wave generator based on the fluid-structure interaction
    Fang, J.
    Jia, P.
    Liu, W.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2011, 225 (C12) : 2892 - 2899
  • [30] A novel modeling method to investigate the performance of aerostatic spindle considering the fluid-structure interaction
    Gao, Qiang
    Lu, Lihua
    Chen, Wanqun
    Chen, Guoda
    Wang, Guanglin
    TRIBOLOGY INTERNATIONAL, 2017, 115 : 461 - 469