Computed tomography-based device-sizing in Amplatzer Amulet left atrial appendage occlusion

被引:5
作者
Nirmalan, Jonatan Gerard [1 ]
Kramer, Anders [1 ]
Korsholm, Kasper [1 ]
Jensen, Jesper Moller [1 ]
Nielsen-Kudsk, Jens Erik [1 ]
机构
[1] Aarhus Univ Hosp, Dept Cardiol, Palle Juul Jensens Blvd 99, DK-8200 Aarhus N, Denmark
关键词
Amplatzer Amulet; Left atrial appendage; Left atrial appendage occlusion; Cardiac computed tomography; Peri-device leak; PERIDEVICE LEAK; CLOSURE; AGREEMENT;
D O I
10.1007/s10840-023-01665-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundAmplatzer Amulet is a frequently used device for left atrial appendage occlusion (LAAO). The current sizing protocol is based on the maximum diameter of the left atrial appendage (LAA) landing zone. However, mean, perimeter-, or area-derived diameter might be more accurate measures for device sizing.MethodsRetrospective analysis of 150 consecutive patients undergoing LAAO is guided by pre-procedural cardiac CT. A total of 117 patients were included; 7 were excluded due to renal failure and 26 due to closure with the sandwich technique. The maximum, mean, area-, and perimeter-derived diameters of the landing zone were derived from pre-procedural cardiac CT scans, and their accuracy to predict the implanted device size was investigated. The predicted device size was determined based on the currently recommended sizing algorithm. Peri-device leak (PDL) was assessed (grade 1-3) along with the underlying mechanism.ResultsDevice-sizing accuracy was superior for mean, area-, and perimeter derived diameters compared with the maximal diameter, especially for eccentric landing zones. Mean difference between predicted and actually implanted device size was 0.08 mm (+/- 2.77), 0.30 mm (+/- 2.40), - 0.39 mm (+/- 2.43), and - 2.55 mm (+/- 2.57) across mean, area-derived, perimeter-derived, and maximal diameter, respectively. Grade 3 peri-device leak was seen in 8.5% of implants without a significant association to the eccentricity of the landing zone. The leading mechanism for PDL was device malalignment.ConclusionOur results indicate mean, area-, and perimeter-derived diameters of the device landing zone to perform similar and superior in device-sizing accuracy compared with the maximum diameter.
引用
收藏
页码:785 / 795
页数:11
相关论文
共 23 条
[11]   EHRA/EAPCI expert consensus statement on catheter-based left atrial appendage occlusion - an update [J].
Glikson, Michael ;
Wolff, Rafael ;
Hindricks, Gerhard ;
Mandrola, John ;
Camm, A. John ;
Lip, Gregory Y. H. ;
Fauchier, Laurent ;
Betts, Tim R. ;
Lewalter, Thorsten ;
Saw, Jacqueline ;
Tzikas, Apostolos ;
Sternik, Leonid ;
Nietlispach, Fabian ;
Berti, Sergio ;
Sievert, Horst ;
Bertog, Stefan ;
Meier, Bernhard .
EUROPACE, 2020, 22 (02) :184
[12]  
Hindricks G, 2021, EUR HEART J, V42, P546, DOI [10.1093/eurheartj/ehaa945, 10.1093/eurheartj/ehaa612]
[13]   Peridevice Leak Following Amplatzer Left Atrial Appendage Occlusion Cardiac Computed Tomography Classification and Clinical Outcomes [J].
Korsholm, Kasper ;
Jensen, Jesper Moller ;
Norgaard, Bjarne Linde ;
Samaras, Athanasios ;
Saw, Jacqueline ;
Berti, Sergio ;
Tzikas, Apostolos ;
Nielsen-Kudsk, Jens Erik .
JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (01) :83-93
[14]   Expert Recommendations on Cardiac Computed Tomography for Planning Transcatheter Left Atrial Appendage Occlusion [J].
Korsholm, Kasper ;
Berti, Sergio ;
Iriart, Xavier ;
Saw, Jacqueline ;
Wang, Dee Dee ;
Cochet, Hubert ;
Chow, Danny ;
Clemente, Alberto ;
De Backer, Ole ;
Jensen, Jesper Moller ;
Nielsen-Kudsk, Jens Erik .
JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (03) :277-292
[15]  
Korsholm Kasper, 2018, Interv Cardiol Clin, V7, P229, DOI [10.1016/j.iccl.2017.12.004, 10.1016/j.iccl.2017.12.004]
[16]  
Osmancik Pavel, 2021, J Atr Fibrillation, V13, P20200449, DOI [10.4022/jafib.20200449, 10.4022/jafib.20200449]
[17]   "Chicken-Wing" Undersizing for "Sandwich" Left Atrial Appendage Closure Technique An Image of Caution [J].
Pracon, Radoslaw ;
Konka, Marek ;
Kepka, Cezary ;
Kaczmarska-Dyrda, Edyta ;
Michalowska, Ilona ;
Demkow, Marcin .
JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (11) :E127-E128
[18]   Peridevice Leak After Transcatheter Left Atrial Appendage Occlusion An Analysis of the Amulet IDE Trial [J].
Price, Matthew J. ;
Ellis, Christopher R. ;
Nielsen-Kudsk, Jens Erik ;
Thaler, David ;
Gupta, Nigel ;
Koulogiannis, Konstantinos ;
Anderson, Jordan A. ;
Gage, Ryan ;
Lakkireddy, Dhanunjaya .
JACC-CARDIOVASCULAR INTERVENTIONS, 2022, 15 (21) :2127-2138
[19]   5-Year Outcomes After Left Atrial Appendage Closure From the PREVAIL and PROTECT AF Trials [J].
Reddy, Vivek Y. ;
Doshi, Shephal K. ;
Kar, Saibal ;
Gibson, Douglas N. ;
Price, Matthew J. ;
Huber, Kenneth ;
Horton, Rodney P. ;
Buchbinder, Maurice ;
Neuzil, Petr ;
Gordon, Nicole T. ;
Holmes, David R., Jr. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (24) :2964-2975
[20]   Comparison of left atrial appendage parameters using computed tomography vs. transesophageal echocardiography for watchman device implantation: a systematic review & meta-analysis [J].
Sattar, Yasar ;
Kompella, Ritika ;
Ahmad, Bachar ;
Aamir, Muhammad ;
Suleiman, Abdul-Rahman M. ;
Zghouzi, Mohamed ;
Ullah, Waqas ;
Zafrullah, Fnu ;
Elgendy, Islam Y. ;
Balla, Sudharshan ;
Kawsara, Akram ;
Alraies, M. Chadi .
EXPERT REVIEW OF CARDIOVASCULAR THERAPY, 2022, 20 (02) :151-160