Isolated subcutaneous implantable cardioverter-defibrillator generator displacement causing inappropriate shocks despite preserved lead tip and coil position: a case report

被引:0
作者
Chow, Ji-Jian [1 ]
Bilham, James [2 ]
Lane, Rebecca [2 ]
机构
[1] Hammersmith Hosp, London W12 0NN, England
[2] Harefield Hosp, Cardiol Dept, Uxbridge UB9 6JH, England
关键词
Implantable cardioverter-defibrillator; Dilated cardiomyopathy; Primary prevention; Inappropriate shock; Radiography; Case report;
D O I
10.1093/ehjcr/ytad023
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Subcutaneous implantable cardioverter-defibrillators (S-ICDs) are increasingly used in patients without a pacing indication, and may reduce venous complications, endocarditis and extraction morbidity. Subcutaneous implantable cardioverter-defibrillator displacements may be less obvious than their transvenous counterparts. Case summary A 59-year-old man was found to have dilated cardiomyopathy associated with heavy alcohol intake following investigation for a stroke. Despite 2 years of alcohol cessation and optimal drug therapy, his ejection fraction remained severely impaired, and he received an S-ICD using the manufacturer recommended screening and implant method, and by an experienced operator. Ten months later, inappropriate shocks were delivered despite optimal programming. Device displacement was demonstrated by lateral chest radiography on the second instance of inappropriate therapy. On the first admission, a lateral film was not performed, and simple device programming was undertaken which failed to prevent the second occurrence. The patient requested an explant; as ventricular function had improved following initiation of sacubutril/valsartan, the clinical team opted to remove the device. Discussion Careful inspection of lateral chest films and review of device indication are needed to reduce the risk of inappropriate shocks. This is the first description of inappropriate device activity following lone generator displacement-lead displacement is well described.
引用
收藏
页数:5
相关论文
共 6 条
[1]   Inappropriate shocks in single-chamber and subcutaneous implantable cardioverter-defibrillators: a systematic review and meta-analysis [J].
Auricchio, Angelo ;
Hudnall, J. Harrison ;
Schloss, Edward J. ;
Sterns, Laurence D. ;
Kurita, Takashi ;
Meijer, Albert ;
Fagan, Dedra H. ;
Rogers, Tyson .
EUROPACE, 2017, 19 (12) :1973-1980
[2]   Implantation of the Subcutaneous Implantable Cardioverter-Defibrillator An Evaluation of 4 Implantation Techniques [J].
Brouwer, Tom F. ;
Miller, Marc A. ;
Quast, Anne-Floor B. E. ;
Palaniswamy, Chandrasekar ;
Dukkipati, Srinivas R. ;
Reddy, Vivek ;
Wilde, Arthur A. ;
Willner, Jonathan M. ;
Knops, Reinoud E. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (01)
[3]   Twiddler's syndrome with a subcutaneous implantable cardioverter-defibrillator presenting with an inappropriate shock: a case report [J].
Camm, C. Fielder ;
Rajappan, Kim ;
Curson, Mark ;
Tilling, Lindsey .
EUROPEAN HEART JOURNAL-CASE REPORTS, 2019, 3 (04)
[4]   The Entirely Subcutaneous Defibrillator (S-Icd): State of the Art and Selection of the Ideal Candidate [J].
De Maria, Elia ;
Olaru, Alina ;
Cappelli, Stefano .
CURRENT CARDIOLOGY REVIEWS, 2015, 11 (02) :180-186
[5]   Inappropriate shocks in the subcutaneous ICD: Incidence, predictors and management [J].
Nordkamp, Louise R. A. Olde ;
Brouwer, Tom F. ;
Barr, Craig ;
Theuns, Dominic A. M. J. ;
Boersma, Lucas V. A. ;
Johansen, Jens B. ;
Neuzil, Petr ;
Wilde, Arthur A. M. ;
Carter, Nathan ;
Husby, Michael ;
Lambiase, Pier D. ;
Knops, Reinoud E. .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2015, 195 :126-133
[6]   Intermuscular technique for implantation of the subcutaneous implantable cardioverter defibrillator: long-term performance and complications [J].
Winter, Joachim ;
Siekiera, Markus ;
Shin, Dong-In ;
Meyer, Christian ;
Kroepil, Patric ;
Clahsen, Harald ;
O'Connor, Stephen .
EUROPACE, 2017, 19 (12) :2036-2041