Evaluation of a planar diode matrix for SRS patient-specific QA in comparison with GAFchromic films

被引:6
作者
Infusino, Erminia [1 ,3 ]
Ianiro, Anna [1 ]
Luppino, Stefano [1 ]
Nocentini, Sandro [1 ]
Pugliatti, Cristina [2 ]
Soriani, Antonella [1 ]
机构
[1] IRCCS Regina Elena Natl Canc Inst, Dept Med Phys, Rome, Italy
[2] Michele & Pietro Ferrero Hosp, Dept Med Phys, Verduno, Italy
[3] IRCCS Regina Elena Natl Canc Inst, Med Phys Dept, Rome, Italy
关键词
Cyberknife; pre-treatment verification; quality assurance; radiosurgery; DELIVERY QUALITY-ASSURANCE; ION-CHAMBER ARRAY; RADIOSURGERY SYSTEM; IONIZATION-CHAMBER; RADIATION-THERAPY; IMRT; CYBERKNIFE; COMPLEXITY; VERIFICATION; RESOLUTION;
D O I
10.1002/acm2.13947
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe validate the routine use of a two-dimensional (2D) diode matrix for patient specific pre-treatment verification for Cyberknife (CK) stereotactic radiosurgery and to compare it with film dosimetry. Materials and methodA total of 46 patients were selected according to the most frequent diseases treated at our institution with the CK system, that is, brain metastases, meningiomas, spine metastases, and prostate tumors. All cases were evaluated with GAFChromic EBT-3 films and SRS MapCHECK for Fixed cone, IRIS, and MLC collimators of the CK. ResultsThe highest mean passing rate was observed for the SRS MapCHECK system compared to films. In order to assess if the two techniques provide statistically different results, a Wilcoxon Signed-Rank non-parametric test was performed (p < 0.05) and we found gamma values significantly lower for EBT-3 films with respect to the SRS MapCHECK. We noticed a moderately significant association between the two techniques using Spearman's rank correlation coefficient (rs > 0.4). We also performed the Bland-Altman statistical method: less than 5% of the differences resulted outside the range (mean & PLUSMN; 1.96 x SD), so the two methods can be considered interchangeable within the combined inaccuracy. ConclusionsThe use of SRS MapCHECK for CK patient specific quality assurance (QA) is feasible for a variety of clinical districts and could be reliably used as a replacement for radiochromic films.
引用
收藏
页数:10
相关论文
共 35 条
[1]   The cyberknife: A frameless robotic system for radiosurgery [J].
Adler, JR ;
Chang, SD ;
Murphy, MJ ;
Doty, J ;
Geis, P ;
Hancock, SL .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 69 (1-4) :124-128
[2]   Comprehensive evaluation of the high-resolution diode array for SRS dosimetry [J].
Ahmed, Saeed ;
Zhang, Geoffrey ;
Moros, Eduardo G. ;
Feygelman, Vladimir .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10) :13-23
[3]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[4]  
[Anonymous], GAFChromicEBT3 film specifications
[5]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[6]   Plan delivery quality assurance for CyberKnife: Statistical process control analysis of 350 film-based patient-specific QAs [J].
Bellec, J. ;
Delaby, N. ;
Jouyaux, F. ;
Perdrieux, M. ;
Bouvier, J. ;
Sorel, S. ;
Henry, O. ;
Lafond, C. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 39 :50-58
[7]   Stereotactic body radiation therapy: The report of AAPM Task Group 101 [J].
Benedict, Stanley H. ;
Yenice, Kamil M. ;
Followill, David ;
Galvin, James M. ;
Hinson, William ;
Kavanagh, Brian ;
Keall, Paul ;
Lovelock, Michael ;
Meeks, Sanford ;
Papiez, Lech ;
Purdie, Thomas ;
Sadagopan, Ramaswamy ;
Schell, Michael C. ;
Salter, Bill ;
Schlesinger, David J. ;
Shiu, Almon S. ;
Solberg, Timothy ;
Song, Danny Y. ;
Stieber, Volker ;
Timmerman, Robert ;
Tome, Wolfgang A. ;
Verellen, Dirk ;
Wang, Lu ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2010, 37 (08) :4078-4101
[8]   High resolution ion chamber array delivery quality assurance for robotic radiosurgery: Commissioning and validation [J].
Blanck, Oliver ;
Masi, Laura ;
Chan, Mark K. H. ;
Adamczyk, Sebastian ;
Albrecht, Christian ;
Damme, Marie-Christin ;
Loutfi-Krauss, Britta ;
Alraun, Manfred ;
Fehr, Roman ;
Ramm, Ulla ;
Siebert, Frank-Andre ;
Stelljes, Tenzin Sonam ;
Poppinga, Daniela ;
Poppe, Bjoern .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (06) :838-846
[9]   Film-based delivery quality assurance for robotic radiosurgery: Commissioning and validation [J].
Blanck, Oliver ;
Masi, Laura ;
Damme, Marie-Christin ;
Hildebrandt, Guido ;
Dunst, Juergen ;
Siebert, Frank-Andre ;
Poppinga, Daniela ;
Poppe, Bjoern .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (05) :476-483
[10]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310