Comparison of the TEMPO binocular perimeter and Humphrey field analyzer

被引:2
作者
Nishida, Takashi [1 ,2 ]
Weinreb, Robert N. [1 ,2 ]
Arias, Juan [1 ,2 ]
Vasile, Cristiana [1 ,2 ]
Moghimi, Sasan [1 ,2 ]
机构
[1] Univ Calif San Diego, Shiley Eye Inst, Hamilton Glaucoma Ctr, 9500 Campus Point Dr, La Jolla, CA 92093 USA
[2] Univ Calif San Diego, Viterbi Family Dept Ophthalmol, 9500 Campus Point Dr, La Jolla, CA 92093 USA
基金
美国国家卫生研究院;
关键词
VISUAL-FIELD; RELIABILITY;
D O I
10.1038/s41598-023-48105-5
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study compared between TEMPO, a new binocular perimeter, with the Humphrey Field Analyzer (HFA). Patients were tested with both TEMPO 24-2 Ambient Interactive Zippy Estimated by Sequential Testing (AIZE)-Rapid and HFA 24-2 Swedish Interactive Threshold Algorithm (SITA)-Fast in a randomized sequence on the same day. Using a mixed-effects model, visual field (VF) parameters and reliability indices were compared. Retinal nerve fiber layer (RNFL) thickness was measured using Cirrus optical coherence tomography (OCT), and coefficient of determinations for VF and OCT parameters were calculated and compared using Akaike information criteria. 740 eyes (including 68 healthy, 262 glaucoma suspects, and 410 glaucoma) of 370 participants were evaluated. No significant differences were seen in mean deviation and visual field index between the two perimeters (P > 0.05). A stronger association between VF mean sensitivity (dB or 1/L) and circumpapillary RNFL was found for TEMPO (adjusted R-2 = 0.25; Akaike information criteria [AIC] = 5235.5 for dB, and adjusted R-2 = 0.29; AIC = 5200.8 for 1/L, respectively) compared to HFA (adjusted R-2 = 0.22; AIC = 5263.9 for dB, and adjusted R-2 = 0.22; AIC = 5262.7 for 1/L, respectively). Measurement time was faster for TEMPO compared to HFA (261 s vs. 429 s, P < 0.001). Further investigations are needed to assess the long-term monitoring potential of this binocular VF test.
引用
收藏
页数:8
相关论文
共 50 条
[41]   Comparison of the structure-function relationship between compass microperimetry and Humphrey field analyser in myopic open-angle glaucoma eyes [J].
Shin, Joong Won ;
Song, Min Kyung ;
Won, Hun Jae ;
Jo, Younhye ;
Kook, Michael S. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2022, 106 (04) :485-490
[42]   Development of Visual Field Screening Procedures: A Case Study of the Octopus Perimeter [J].
Turpin, Andrew ;
Myers, Jonathan S. ;
McKendrick, Allison M. .
TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2016, 5 (03)
[43]   Comparison of the Humphrey Swedish interactive thresholding algorithm (SITA) and full threshold strategies [J].
Sharma, AK ;
Goldberg, I ;
Graham, SL ;
Mohsin, M .
JOURNAL OF GLAUCOMA, 2000, 9 (01) :20-27
[44]   Comparison of Peristat Online Perimetry with the Humphrey Perimetry in a Clinic-Based Setting [J].
Lowry, Eugene A. ;
Hou, Jing ;
Hennein, Lauren ;
Chang, Robert T. ;
Lin, Shan ;
Keenan, Jeremy ;
Wang, Sean K. ;
Ianchulev, Sean ;
Pasquale, Louis R. ;
Han, Ying .
TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2016, 5 (04)
[45]   Modified method for assessment of the binocular fusional field in patients with suppression [J].
Yagasaki, Teiji ;
Oya, Yoshimi ;
Maeda, Mariko ;
Tsukui, Makiko .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2009, 53 (02) :138-144
[46]   Development of a reliability criterion for the binocular Esterman visual field test [J].
Hirasawa, Kazunori ;
Nagayama-Ito, Akari ;
Maruta-Akaishizawa, Chiho ;
Shoji, Nobuyuki .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2025,
[47]   Validation and reproducibility of the Heidelberg Edge Perimeter in the detection of glaucomatous visual field defects [J].
Cui, Qi N. ;
Gogt, Priyanka ;
Lam, Jonathan M. ;
Siraj, Safa ;
Hark, Lisa A. ;
Myers, Jonathan S. ;
Katz, L. Jay ;
Waisbourd, Michael .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2019, 12 (04) :577-581
[48]   Validation and reproducibility of the Heidelberg Edge Perimeter in the detection of glaucomatous visual field defects [J].
Qi N.Cui ;
Priyanka Gogt ;
Jonathan M.Lam ;
Safa Siraj ;
Lisa A.Hark ;
Jonathan S.Myers ;
L.Jay Katz ;
Michael Waisbourd .
International Journal of Ophthalmology, 2019, (04) :577-581
[49]   Glaucoma Progression Detection and Humphrey Visual Field Prediction Using Discriminative and Generative Vision Transformers [J].
Tian, Ye ;
Zang, Mingyang ;
Sharma, Anurag ;
Gu, Sophie Z. ;
Leshno, Ari ;
Thakoor, Kaveri A. .
OPHTHALMIC MEDICAL IMAGE ANALYSIS, OMIA 2023, 2023, 14096 :62-71
[50]   Relationships of Retinal Structure and Humphrey 24-2 Visual Field Thresholds in Patients With Glaucoma [J].
Bogunovic, Hrvoje ;
Kwon, Young H. ;
Rashid, Adnan ;
Lee, Kyungmoo ;
Critser, Douglas B. ;
Garvin, Mona K. ;
Sonka, Milan ;
Abramoff, Michael D. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (01) :259-271