Dose uncertainty due to energy dependence in dual-energy computed tomography

被引:1
作者
Kawahara, Daisuke [1 ,4 ]
Toyoda, Takatoshi [2 ]
Yokomachi, Kazushi [2 ]
Fujioka, Chikako [2 ]
Nagata, Yasushi [1 ,3 ]
机构
[1] Hiroshima Univ, Inst Biomed & Hlth Sci, Dept Radiat Oncol, Hiroshima, Japan
[2] Hiroshima Univ Hosp, Div Clin Support, Radiat Therapy Sect, Hiroshima, Japan
[3] Hiroshima High Precis Radiotherapy Canc Ctr, Hiroshima, Japan
[4] Hiroshima Univ Hosp, Dept Clin Support, Radiat Therapy Sect, 1 2 3 Kasumi, Minami ku, Hiroshima 7348551, Japan
关键词
dual-energy CT; energy absorption coefficient; CT; CONTRAST; HEAD; ANGIOGRAPHY; NECK;
D O I
10.5114/pjr.2023.128682
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the absolute dose uncertainty at 2 different energies and for the large and small bowtie filters in dual-energy computed tomography (DECT).Material and methods: Measurements were performed using DECT at 80 kV and 140 kilovoltage peak (kVp), and sin-gle-energy computed tomography (CT) at 120 kV. The absolute dose was calculated from the mass-energy absorption obtained from the half-value layer (HVL) of aluminium.Results: The difference in the water-to-air ratio of the mean mass energy-absorption coefficients at 80 kV and 140 kV was 2.0% for the small bow-tie filter and 3.0% for the large bow-tie filter. At lower tube voltages, the difference in the absorbed dose with the large and small bow-tie filters was larger.Conclusions: The absolute dose uncertainty due to energy dependence was 3.0%, which could be reduced with sin-gle-energy beams at 120 kV or by using the average effective energy measurement with dual-energy beams.
引用
收藏
页码:e270 / e274
页数:5
相关论文
共 21 条
[1]  
Aichinger H., 2004, Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology
[2]   Xenon ventilation CT with a dual-energy technique of dual-source CT: Initial experience [J].
Chae, Eun Jin ;
Seo, Joon Beom ;
Goo, Hyun Woo ;
Kim, Namkug ;
Song, Koun-Sik ;
Lee, Sang Do ;
Hong, Soo-Jong ;
Krauss, Bernhard .
RADIOLOGY, 2008, 248 (02) :615-624
[3]   Dual-energy CT angiography of the lungs: Comparison of test bolus and bolus tracking techniques for the determination of scan delay [J].
Henzler, Thomas ;
Meyer, Mathias ;
Reichert, Miriam ;
Krissak, Radko ;
Nance, John W., Jr. ;
Haneder, Stefan ;
Schoenberg, Stefan O. ;
Fink, Christian .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (01) :132-138
[4]   Dual-Energy Computed Tomography Angiography of the Head and Neck With Single-Source Computed Tomography: A New Technical (Split Filter) Approach for Bone Removal [J].
Kaemmerer, Nadine ;
Brand, Michael ;
Hammon, Matthias ;
May, Matthias ;
Wuest, Wolfgang ;
Krauss, Bernhard ;
Uder, Michael ;
Lell, Michael M. .
INVESTIGATIVE RADIOLOGY, 2016, 51 (10) :618-623
[5]  
Kawahara Daisuke, 2019, BJR Open, V1, P20180045, DOI 10.1259/bjro.20180045
[6]   Evaluation of metal artefact techniques with same contrast scale for different commercially available dual-energy computed tomography scanners [J].
Kawahara, Daisuke ;
Ozawa, Shuichi ;
Yokomachi, Kazushi ;
Higaki, Toru ;
Shiinoki, Takehiro ;
Fujioka, Chikako ;
Saito, Akito ;
Miki, Kentaro ;
Kimura, Tomoki ;
Murakami, Yuji ;
Awai, Kazuo ;
Nagata, Yasushi .
PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) :539-546
[7]   Improving automatic contrast agent extraction system using monochromatic CT number [J].
Kawahara, Daisuke ;
Ozawa, Shuichi ;
Yokomachi, Kazushi ;
Higaki, Toru ;
Fujioka, Chikako ;
Mori, Masayoshi ;
Nagata, Yasushi .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (03) :819-826
[8]   Evaluation of raw-data-based and calculated electron density for contrast media with a dual-energy CT technique [J].
Kawahara, Daisuke ;
Ozawa, Shuichi ;
Yokomachi, Kazushi ;
Higaki, Toru ;
Shiinoki, Takehiro ;
Ohno, Yoshimi ;
Murakami, Yuji ;
Awai, Kazuo ;
Nagata, Yasushi .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (05) :499-506
[9]   Automatic contrast medium extraction system using electron density data with dual-energy CT [J].
Kawahara, Daisuke ;
Ozawa, Shuichi ;
Tanaka, Sodai ;
Yokomachi, Kazushi ;
Higaki, Toru ;
Saito, Akito ;
Miki, Kentaro ;
Fujioka, Chikako ;
Ohno, Yoshimi ;
Kimura, Tomoki ;
Murakami, Yuji ;
Nagata, Yasushi .
BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1090)
[10]   Accuracy of the raw-data-based effective atomic numbers and monochromatic CT numbers for contrast medium with a dual-energy CT technique [J].
Kawahara, Daisuke ;
Ozawa, Shuichi ;
Yokomachi, Kazushi ;
Tanaka, Sodai ;
Higaki, Toru ;
Fujioka, Chikako ;
Suzuki, Tatsuhiko ;
Tsuneda, Masato ;
Nakashima, Takeo ;
Ohno, Yoshimi ;
Nagata, Yasushi .
BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1082)