Comparison of the Discovery A and Stratos DR densitometers for assessing whole-body and regional bone mineral density and body composition

被引:1
作者
Maimoun, Laurent [1 ,2 ]
Mahadea, Krishna Kunal [3 ]
Alonso, Sandrine [4 ]
Chevallier, Thierry [4 ,5 ]
Kotzki, Pierre-Olivier [6 ,7 ]
Mura, Thibault [4 ]
Boudousq, Vincent [6 ,7 ]
机构
[1] CHU Montpellier, Hop Lapeyronie, Nucl Med Serv, Montpellier, France
[2] Univ Montpellier, Physiol & Med Expt Coeur & Muscles PhyMedEx, INSERM, CNRS, Montpellier, France
[3] Univ Montpellier, Montpellier, France
[4] Univ Montpellier, Nimes Univ Hosp, Dept Biostat Clin Epidemiol Publ Hlth & Innovat M, Nimes, France
[5] Univ Montpellier, Inst Desbrest Epidemiol & Sante Publ, UMR, INSERM, Montpellier, France
[6] CHU Nimes, Hop Caremeau, Nucl Med Serv, Nimes, France
[7] Univ Montpellier, Inst Rech Cancerol Montpellier, Inst Reg Canc Montpellier, IRCM,INSERM U1194, Montpellier, France
关键词
body composition; bone mineral density (BMD); cross-calibration; Discovery A; dual X-ray absorptiometry (DXA); Stratos DR; visceral adipose tissue; X-RAY ABSORPTIOMETRY; GE-LUNAR IDXA; IN-VIVO PRECISION; VISCERAL ADIPOSE-TISSUE; CROSS-CALIBRATION; FAT DISTRIBUTION; DXA; MASS; REPRODUCIBILITY; POPULATION;
D O I
10.1111/cpf.12836
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
PurposeThe agreement between the Stratos DR and Discovery A densitometers was assessed for measurements of whole-body (WB) and regional fat mass (FM), fat-free soft tissue (FFST) and bone mineral density (BMD). Moreover, the precision of the Stratos DR was also evaluated. MethodsFifty participants (35 women, 70%) were measured consecutively, once on the Discovery A and once on the Stratos DR. In a subgroup of participants (n = 29), two successive measurements with the Stratos DR were also performed. ResultsFM, FFST and BMD measured with the two devices were highly correlated, with a coefficient of correlation ranging from 0.80 to 0.99. Bland-Altman analyses indicated significant bias between the two devices for all measurements. Thus, compared to the Discovery A, the Stratos DR underestimated WB BMD and WB and regional FM and FFST, with the exception of trunk FM and visceral adipose tissue (VAT), which were overestimated. Precision error for the Stratos DR, when expressed as root mean square-coefficient of variation (RMS-CV%) for FM, was 1.4% for WB, 3.0% for the gynoid and android regions, and 15.9% for VAT. The RMS-CV% for FFST was 1.0% for WB. The root mean square of standard deviation for WB BMD was 0.018 g/cm(2), corresponding to a 1.4% CV. The least significant change was 0.050 g/cm(2) (SD), and 4.0% was considered to be a significant biological change. ConclusionsDifferences between the Stratos DR and Discovery A measurements are significant and require the use of translational cross-calibration equations. For most of the BMD and body composition parameters, our results demonstrated good Stratos DR precision.
引用
收藏
页码:382 / 392
页数:11
相关论文
共 45 条
[1]   Clinical applications of body composition measurements using DXA [J].
Albanese, CV ;
Diessel, E ;
Genant, HK .
JOURNAL OF CLINICAL DENSITOMETRY, 2003, 6 (02) :75-85
[2]   Precision Error in Dual-Energy X-Ray Absorptiometry Body Composition Measurements in Elite Male Rugby League Players [J].
Barlow, Matthew J. ;
Oldroyd, Brian ;
Smith, Debbie ;
Lees, Matthew J. ;
Brightmore, Amy ;
Till, Kevin ;
Jones, Benjamin ;
Hind, Karen .
JOURNAL OF CLINICAL DENSITOMETRY, 2015, 18 (04) :546-550
[3]   DXA: Technical aspects and application [J].
Bazzocchi, Alberto ;
Ponti, Federico ;
Albisinni, Ugo ;
Battista, Giuseppe ;
Guglielmi, Giuseppe .
EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (08) :1481-1492
[4]   Visceral Fat and Body Composition Changes in a Female Population After RYGBP: a Two-Year Follow-Up by DXA [J].
Bazzocchi, Alberto ;
Ponti, Federico ;
Cariani, Stefano ;
Diano, Danila ;
Leuratti, Luca ;
Albisinni, Ugo ;
Marchesini, Giulio ;
Battista, Giuseppe .
OBESITY SURGERY, 2015, 25 (03) :443-451
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Canchola J., 2022, PAPER SA 004
[7]   In Vivo Precision of the GE iDXA for the Assessment of Total Body Composition and Fat Distribution in Severely Obese Patients [J].
Carver, Tamara E. ;
Christou, Nicholas V. ;
Andersen, Ross E. .
OBESITY, 2013, 21 (07) :1367-1369
[8]   Cross-Calibration of iDXA and Prodigy on Spine and Femur Scans in Korean Adults [J].
Choi, Yong Jun ;
Lee, Byung Joo ;
Lim, Hyun Chae ;
Chung, Yoon-Sok .
JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (04) :450-455
[9]  
CHRISTIANSEN C, 1991, AM J MED, V90, P107
[10]   Regional Body Composition: Cross-calibration of DXA Scanners-QDR4500W and Discovery Wi [J].
Covey, Margaret K. ;
Berry, Jean K. ;
Hacker, Eileen D. .
OBESITY, 2010, 18 (03) :632-637