A hybrid Bayesian network for medical device risk assessment and management

被引:5
作者
Hunte, Joshua L. [1 ]
Neil, Martin [1 ,2 ]
Fenton, Norman E. [1 ,2 ]
机构
[1] Queen Mary Univ London, Sch Elect Engn & Comp Sci, Risk & Informat Management Res Grp, London E1 4NS, England
[2] Agena Ltd, Cambridge, England
关键词
Medical device; ISO; 14971; Medical device risk management; Bayesian networks; Risk assessment; FAULT-TREE ANALYSIS; OF-THE-ART; SAFETY;
D O I
10.1016/j.ress.2023.109630
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Risk analysis methods for medical devices, including fault tree analysis, have limitations such as handling uncertainty and providing reasonable risk estimates with limited or no testing data. To address these limitations, this paper proposes a novel systematic method for medical device risk management using hybrid Bayesian networks (BNs). We apply the method to a Defibrillator device to demonstrate the process involved for risk management during production and post-production using 4 different scenarios: (1) where there are available testing data; (2) where there are limited or no testing data; (3) where it is a completely new device with no testing data; (4) where we are reassessing the risk of a previous model on the market based on reported hazards and injuries. In each scenario, the BN model, for the available data, provides the full probability of failure per demand distribution for each category of injury severity (fatal, critical, major, minor, negligible) and the probabilities associated with various risk acceptability criteria. The model results are validated using publicly available data for the LIFEPAK 1000 Defibrillator (PN: 320371500XX), which was recalled by Physio-Control in 2017. The results show that the device would fail the acceptability criteria for probability of fatal injury.
引用
收藏
页数:15
相关论文
共 75 条
[21]  
Hunte J, 2022, arXiv, DOI DOI 10.48550/ARXIV.2206.02144
[22]   A causal Bayesian network approach for consumer product safety and risk assessment [J].
Hunte, Joshua L. ;
Neil, Martin ;
Fenton, Norman E. .
JOURNAL OF SAFETY RESEARCH, 2022, 80 :198-214
[23]  
Hyman W. A., 2002, Journal of Clinical Engineering, V27, P134
[24]  
Hyman W.A., 2008, Journal of Clinical Engineering, V33, P85, DOI DOI 10.1097/01.JCE.0000305872.86942.66
[25]  
IEC, 2015, IEC 62366-1:2015 Medical devices - Part 1: application of usability engineering to medical devices
[26]  
IEC, 2015, IEC 60601-1-11
[27]  
IMDRF, 2013, INT MED DEV REG FOR
[28]  
ISO, 2020, ISO 24971 medical devices - guidance on the application of ISO 14971
[29]   Probability Analysis of Construction Risk based on Noisy-or Gate Bayesian Networks [J].
Ji, Chenyi ;
Su, Xing ;
Qin, Zhongfu ;
Nawaz, Ahsan .
RELIABILITY ENGINEERING & SYSTEM SAFETY, 2022, 217
[30]   Applications of Bayesian networks and Petri nets in safety, reliability, and risk assessments: A review [J].
Kabir, Sohag ;
Papadopoulos, Yiannis .
SAFETY SCIENCE, 2019, 115 :154-175