Canceling Notch Improves the Mechanical Safety of Clavicle Locking Plate: A 3D Finite Element Study

被引:5
作者
Han, Zhihua [1 ,3 ]
Luo, Qian [2 ]
Deng, Guoying [1 ]
Bi, Chun [1 ]
Yin, Gang [1 ]
Lin, Haodong [1 ]
Wu, Jianhong [1 ]
Wu, Xiaoming [1 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Gen Hosp, Trauma Ctr, Dept Orthopaed & Traumatol, Shanghai 201620, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Gen Hosp, Dept Radiol, Shanghai, Peoples R China
[3] Sino Euro Orthopaed Network, Berlin, Germany
关键词
Biomechanics; Clavicle fracture; Finite element; Locking plate; NONOPERATIVE TREATMENT; OPERATIVE FIXATION; FRACTURES; OSTEOSYNTHESIS; COMPLICATIONS; MULTICENTER;
D O I
10.1111/os.13558
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective Implant failure is a disastrous complication of the operative treatment of midshaft clavicle fractures, and improving the osteosynthesis plate is a strategy for preventing this. We aimed to investigate whether canceling the notch and adding screw-hole inserts enhanced the mechanical properties of the plate. Methods A clavicle model was generated based on the CT images of six adult volunteers (age range, 20-40 years; three males and three females; height range 160-175) using dedicated software, and a midshaft fracture model was created. The domestically made seven-hole locking plate commonly used for midshaft clavicle fractures was simulated (Model I); modifications were made to the plate (Model II). Using 3D finite element analysis, we simulated the fracture construct under three different load conditions-downward cantilever bending, axial compression, and axial torsion-and compared the stress distribution. Results We found that under axial compression, Model II experienced its maximum stress on the plate at 551.9MPa, which was less than that in Model I (790.4 MPa). Moreover, a greater stress concentration at the fracture site was observed under axial torsion, despite the maximum stress of both the models being similar. Conclusion Canceling the notch and filling the screw holes near the fracture can ameliorate stress concentration on the internal fixation construct and enhance its reliability under axial compression. This improvement has substantial effects on the mechanical properties of implants and potentially prevents implant failure. Modern osteosynthesis anatomical implants need to be improved.
引用
收藏
页码:2152 / 2156
页数:5
相关论文
共 24 条
[1]   An intramedullary Echidna pin for fixation of comminuted clavicle fractures: a biomechanical study [J].
Ackland, David ;
Griggs, Ian ;
Hislop, Patrick ;
Wu, Wen ;
Patel, Minoo ;
Richardson, Martin .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2017, 12
[2]   Complications associated with operative fixation of acute midshaft clavicle fractures [J].
Asadollahi, Saeed ;
Hau, Raphael C. ;
Page, Richard S. ;
Richardson, Martin ;
Edwards, Elton R. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (06) :1248-1252
[3]   Mechanical failure of plate breakage after open reduction and plate fixation of displaced midshaft clavicle fracture - a possible new risk factor: a case report [J].
Batash, Ron ;
Debi, Ronen ;
Grinberg, Dan ;
Shema, Maayan ;
Elbaz, Avi ;
Benedict, Yeshaiau .
JOURNAL OF MEDICAL CASE REPORTS, 2019, 13 (01)
[4]   Threaded Screw Head Inserts Improve Locking Plate Biomechanical Properties [J].
Bellapianta, Joseph ;
Dow, Kyle ;
Pallotta, Nicholas A. ;
Hospodar, Paul P. ;
Uhl, Richard L. ;
Ledet, Eric H. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (02) :65-71
[5]   Comparison of implant failure rates of different plates for midshaft clavicular fractures based on fracture classifications [J].
Chiu, Yung-Cheng ;
Huang, Kui-Chou ;
Shih, Cheng-Min ;
Lee, Kun-Tsan ;
Chen, Kun-Hui ;
Hsu, Cheng-En .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)
[6]   Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation [J].
Eden, Lars ;
Doht, Stefanie ;
Frey, Soenke P. ;
Ziegler, Dirk ;
Stoyhe, Jan ;
Fehske, Kai ;
Blunk, Torsten ;
Meffert, Rainer H. .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (12) :2537-2543
[7]   Balser Plate Stabilization for Traumatic Sternoclavicular Instabilities or Medial Clavicle Fractures: A Case Series and Literature Review [J].
Feng, Wei-lou ;
Cai, Xiao ;
Li, Shu-hao ;
Li, Zi-jun ;
Zhang, Kun ;
Wang, Hao ;
Zhang, Jun ;
Zhu, Yang-jun ;
Feng, Dong-xu .
ORTHOPAEDIC SURGERY, 2020, 12 (06) :1627-1634
[8]   Locking plate osteosynthesis of clavicle fractures: complication and reoperation rates in one hundred and five consecutive cases [J].
Fridberg, Marie ;
Ban, Ilija ;
Issa, Zaid ;
Krasheninnikoff, Michael ;
Troelsen, Anders .
INTERNATIONAL ORTHOPAEDICS, 2013, 37 (04) :689-692
[9]   Clavicle fractures in adults; current concepts [J].
Frima, Herman ;
van Heijl, Mark ;
Michelitsch, Christian ;
van Der Meijden, Olivier ;
Beeres, Frank J. P. ;
Houwert, Roderick M. ;
Sommer, Christoph .
EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2020, 46 (03) :519-529
[10]   Comparison of the Effectiveness of Simple Plate Fixation and Plate Combined with Local Fixation of Broken Ends in the Treatment of Oblique Fracture of Midshaft Clavicle [J].
Gao, Gong-ming ;
Zhang, Yi ;
Li, Hai-bo ;
Nong, Lu-ming ;
Zhou, Xin-die ;
Jiang, Wei ;
Han, Long .
ORTHOPAEDIC SURGERY, 2022, 14 (07) :1331-1339