A Novel Miniature Spring-Based Continuum Manipulator for Minimally Invasive Surgery: Design and Evaluation

被引:21
作者
Wang, Jie [1 ]
Hu, Chengquan [1 ]
Ning, Guochen [1 ]
Ma, Longfei [1 ]
Zhang, Xinran [1 ]
Liao, Hongen [1 ]
机构
[1] Tsinghua Univ, Sch Med, Dept Biomed Engn, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
Manipulators; Surgery; Springs; Bending; Kinematics; Instruments; Aerospace electronics; Cable-driven robot; flexible surgical instrument; minimally invasive surgery (MIS); offset neutral bending plane; spring-based continuum manipulator; VALIDATION; ROBOT;
D O I
10.1109/TMECH.2023.3252279
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Minimally invasive surgery is gaining wide-spread attention in the laboratory and the clinic due to its advantages of less trauma, less pain, and faster recovery. Miniature continuum instruments with a large central channel are needed for reducing damage to healthy tissue and delivering multiple surgical tools in microsurgery such as fetal surgery and maxillary sinus surgery. In this article, we proposed a novel miniature spring-based cable-driven continuum manipulator characterized by its offset neutral bending plane and its simple and clever structure. These features enable the manipulator to have several advantages such as easy miniaturization, bending stability and uniformity, a large central channel, smaller cable tension required, and low cost. A piecewise constant curvature model, with the offset neutral bending plane taken into account, is presented for the kinematic control of the proposed manipulator. To explore the relationship between the cable tension and manipulator bending angle, the statics model based on the equivalent model principle, is exhaustively discussed. The accuracy of kinematics and statics models, as well as the design superiority, was verified experimentally. Furthermore, teleoperation experiments of peg-transfer tasks and porcine tissue laser ablation are conducted to verify the feasibility and effectiveness of the proposed manipulator in different forms. Final, design variations are showcased to enrich manipulator design and functionality.
引用
收藏
页码:2716 / 2727
页数:12
相关论文
共 46 条
[41]   Design of a multi-arm concentric-tube robot system for transnasal surgery [J].
Wang, Jie ;
Yang, Xing ;
Li, Peng ;
Song, Shuang ;
Liu, Li ;
Meng, Max Q. -H. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (03) :497-508
[42]   A Minimal POE-Based Model for Robotic Kinematic Calibration With Only Position Measurements [J].
Wu, Liao ;
Yang, Xiangdong ;
Chen, Ken ;
Ren, Hongliang .
IEEE TRANSACTIONS ON AUTOMATION SCIENCE AND ENGINEERING, 2015, 12 (02) :758-763
[43]   A Continuum Robotic Cannula With Tip Following Capability and Distal Dexterity for Intracerebral Hemorrhage Evacuation [J].
Yan, Junyan ;
Chen, Jibiao ;
Chen, Junfan ;
Yan, Wanquan ;
Ding, Qingpeng ;
Yan, Kim ;
Du, Jingxin ;
Lam, Chun Ping ;
Wong, George Kwok Chu ;
Cheng, Shing Shin .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2022, 69 (09) :2958-2969
[44]  
York PA, 2015, IEEE INT CONF ROBOT, P1776, DOI 10.1109/ICRA.2015.7139428
[45]   Workspace analysis of cable-driven continuum manipulators based on static model [J].
Yuan, Han ;
Li, Zheng .
ROBOTICS AND COMPUTER-INTEGRATED MANUFACTURING, 2018, 49 :240-252
[46]  
Zhang Y. Xian, 2022, MECH MACH THEORY, V171