Applications of two-dimensional ion chromatography for analytes determination in environmental matrix: A review

被引:7
作者
Yang, Yang [1 ]
Pan, Huimei [1 ]
Li, Xiao [1 ]
Luo, Wang [1 ]
Bharti, Bandna [1 ,2 ]
机构
[1] Harbin Inst Technol Shenzhen, State Key Lab Urban Water Resource & Environm, Shenzhen Key Lab Organ Pollut Prevent & Control, Shenzhen 518055, Peoples R China
[2] DAV Univ, Dept Chem, Jalandhar 144001, Punjab, India
基金
中国国家自然科学基金;
关键词
Two-dimensional ion chromatography; One -pump column -switching ion; chromatography; Coupling columns; Various matrix; TRACE INORGANIC ANIONS; STATIONARY-PHASE; HALOACETIC ACIDS; SIMULTANEOUS SEPARATION; LIQUID-CHROMATOGRAPHY; ORGANIC MATRICES; EXCHANGE COLUMNS; DRINKING-WATER; COMMON ANIONS; FATTY-ACIDS;
D O I
10.1016/j.chroma.2023.463908
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ion chromatography (IC) has grown in usage rapidly since its first introduction in 1975. However, IC is still sometimes unable to separate target analytes from coexisting components well with identical elution time, due to the limited resolution and column capacity, especially in the presence of high-level salt ma-trix. These limitations hence drive IC to develop two-dimensional IC (2D-IC). In this review, we capture the 2D-IC applications in environmental samples via the perspective of coupling different IC columns, which aim to summarize where these 2D-IC methods fit in. In sequence, we firstly review the principles of 2D-IC and emphasize one-pump column-switching IC (OPCS IC) because it is a simplified 2D-IC that only uses one set of IC system. We then compare typical 2D-IC and OPCS IC performances in terms of application scope, method detection limit, drawbacks, and expectations. Finally, we propose some chal-lenges of current methods and opportunities for future research. For instance, it is challenging to couple anion exchange column and capillary column in OPCS IC due to the incompatibility between flow path dimensions and suppressor; coupling ion exclusion column and mixed-bed column may be promising to simultaneously determine anions and cations in weak acids or salts. The details of this study may help practitioners to better understand and implement 2D-IC methods and meanwhile motivate researchers to fill in the knowledge gap in the future.(c) 2023 Elsevier B.V. All rights reserved.
引用
收藏
页数:16
相关论文
共 110 条
[41]   Microbial life under extreme energy limitation [J].
Hoehler, Tori M. ;
Jorgensen, Bo Barker .
NATURE REVIEWS MICROBIOLOGY, 2013, 11 (02) :83-94
[42]   DETERMINATION OF TRACE ANIONS IN WATER BY MULTIDIMENSIONAL ION CHROMATOGRAPHY [J].
HOOVER, TB ;
YAGER, GD .
ANALYTICAL CHEMISTRY, 1984, 56 (02) :221-225
[43]   Simultaneous separation of inorganic cations and anions by ion chromatography using strong anionic/zwitterionic mixed micelles as the stationary phase [J].
Hu, WZ ;
Hasebe, K ;
Reynolds, DM ;
Haraguchi, H .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1997, 20 (08) :1221-1232
[44]   Analysis of trace inorganic anions in weak acid salts by single pump cycling-column-switching ion chromatography [J].
Huang, Zhongping ;
Ni, Chengzhu ;
Zhu, Zhuyi ;
Pan, Zaifa ;
Wang, Lili ;
Zhu, Yan .
JOURNAL OF SEPARATION SCIENCE, 2015, 38 (08) :1294-1300
[45]   Simultaneous determination of peroxydisulfate and conventional inorganic anions by ion chromatography with the column-switching technique [J].
Huang, Zhongping ;
Ni, Chengzhu ;
Wang, Fengli ;
Zhu, Zuoyi ;
Subhani, Qamar ;
Wang, Muhua ;
Zhu, Yan .
JOURNAL OF SEPARATION SCIENCE, 2014, 37 (03) :198-203
[46]   A single pump cycling-column-switching technique coupled with homemade high exchange capacity columns for the determination of iodate in iodized edible salt by ion chromatography with UV detection [J].
Huang, Zhongping ;
Subhani, Qamar ;
Zhu, Zuoyi ;
Guo, Weiqiang ;
Zhu, Yan .
FOOD CHEMISTRY, 2013, 139 (1-4) :144-148
[47]   Simultaneous determination of iodide and iodate in povidone iodine solution by ion chromatography with homemade and exchange capacity controllable columns and column-switching technique [J].
Huang, Zhongping ;
Zhu, Zuoyi ;
Subhani, Qamar ;
Yan, Wenwu ;
Guo, Weiqiang ;
Zhu, Yan .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1251 :154-159
[48]   Simultaneous detection of monovalent anions and cations using all solid-state contact PVC membrane anion and cation-selective electrodes as detectors in single column ion chromatography [J].
Isildak, I ;
Asan, A .
TALANTA, 1999, 48 (04) :967-978
[49]   Two-dimensional ion chromatography using tandem ion-exchange columns with gradient-pulse column switching [J].
Johns, Cameron ;
Shellie, Robert A. ;
Pohl, Christopher A. ;
Haddad, Paul R. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (41) :6931-6937
[50]   Formation of Bromate and Chlorate during Ozonation and Electrolysis in Seawater for Ballast Water Treatment [J].
Jung, Youmi ;
Hong, Eunkyung ;
Yoon, Yeojoon ;
Kwon, Minhwan ;
Kang, Joon-Wun .
OZONE-SCIENCE & ENGINEERING, 2014, 36 (06) :515-525