An image-based metal artifact reduction technique utilizing forward projection in computed tomography

被引:0
|
作者
Ichikawa, Katsuhiro [1 ]
Kawashima, Hiroki [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, 5-11-80 Kodatsuno, Kanazawa 9200942, Japan
[2] Kanazawa Univ Hosp, Dept Diagnost Radiol, 13-1 Takara Machi, Kanazawa 9208641, Japan
关键词
Computed tomography; Metal artifact; Forward projection; Projection data; Parallel computation; RECONSTRUCTION;
D O I
10.1007/s12194-024-00790-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The projection data generated via the forward projection of a computed tomography (CT) image (FP-data) have useful potentials in cases where only image data are available. However, there is a question of whether the FP-data generated from an image severely corrupted by metal artifacts can be used for the metal artifact reduction (MAR). The aim of this study was to investigate the feasibility of a MAR technique using FP-data by comparing its performance with that of a conventional robust MAR using projection data normalization (NMARconv). The NMAR(conv) was modified to make use of FP-data (FPNMAR). A graphics processing unit was used to reduce the time required to generate FP-data and subsequent processes. The performances of FPNMAR and NMAR(conv) were quantitatively compared using a normalized artifact index (AI(n)) for two cases each of hip prosthesis and dental fillings. Several clinical CT images with metal artifacts were processed by FPNMAR. The AI(n) values of FPNMAR and NMAR(conv) were not significantly different from each other, showing almost the same performance between these two techniques. For all the clinical cases tested, FPNMAR significantly reduced the metal artifacts; thereby, the images of the soft tissues and bones obscured by the artifacts were notably recovered. The computation time per image was similar to 56 ms. FPNMAR, which can be applied to CT images without accessing the projection data, exhibited almost the same performance as that of NMAR(conv), while consuming significantly shorter processing time. This capability testifies the potential of FPNMAR for wider use in clinical settings.
引用
收藏
页码:402 / 411
页数:10
相关论文
共 50 条
  • [41] Metal artifact reduction for practical dental computed tomography by improving interpolation-based reconstruction with deep learning
    Liang, Kaichao
    Zhang, Li
    Yang, Hongkai
    Yang, Yirong
    Chen, Zhiqiang
    Xing, Yuxiang
    MEDICAL PHYSICS, 2019, 46 (12) : E823 - E834
  • [42] Imaging of Arthroplasties: Improved Image Quality and Lesion Detection With Iterative Metal Artifact Reduction, a New CT Metal Artifact Reduction Technique
    Subhas, Naveen
    Polster, Joshua M.
    Obuchowski, Nancy A.
    Primak, Andrew N.
    Dong, Frank F.
    Herts, Brian R.
    Iannotti, Joseph P.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 378 - 385
  • [43] Metal artifact correction in head computed tomography based on a homographic adaptation convolution neural network
    Xie, Shipeng
    Song, Zhenrong
    MULTIMEDIA TOOLS AND APPLICATIONS, 2022, 81 (09) : 13045 - 13064
  • [44] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) : R49 - R80
  • [45] A metal artifact reduction method for a dental CT based on adaptive local thresholding and prior image generation
    Hegazy, Mohamed A. A.
    Cho, Min Hyoung
    Lee, Soo Yeol
    BIOMEDICAL ENGINEERING ONLINE, 2016, 15
  • [46] Nonlinearly scaled prior image-controlled frequency split for high-frequency metal artifact reduction in computed tomography
    Anhaus, Julian A.
    Killermann, Philipp
    Sedlmair, Martin
    Winter, Jonas
    Mahnken, Andreas H.
    Hofmann, Christian
    MEDICAL PHYSICS, 2022, 49 (09) : 5870 - 5885
  • [47] A robust index for metal artifact quantification in computed tomography
    Cammin, Jochen
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (08):
  • [48] Prior-Guided Metal Artifact Reduction for Iterative X-Ray Computed Tomography
    Chang, Zhiqian
    Ye, Dong Hye
    Srivastava, Somesh
    Thibault, Jean-Baptiste
    Sauer, Ken
    Bouman, Charles
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2019, 38 (06) : 1532 - 1542
  • [49] Evaluation of the effects of differences in metal artifact type and location on image quality in computed tomography scans
    Lim, Se-Won
    Choi, Woo-Keun
    Kim, Sungchul
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (01) : 80 - 84
  • [50] Framework of metal insertion in the projection domain for image quality optimization in interventional computed tomography
    Ren, Liqiang
    Ferrero, Andrea
    Favazza, Christopher
    JOURNAL OF MEDICAL IMAGING, 2022, 9 (03)