Technical note: Simulation of lung counting applications using Geant4

被引:1
|
作者
Jutila, Henri [1 ,2 ]
Greenlees, Paul [1 ,2 ]
Torvela, Tiina [3 ]
Muikku, Maarit [3 ]
机构
[1] Univ Jyvaskyla, Dept Phys, Accelerator Lab, FI-40014 Jyvaskyla, Finland
[2] Univ Helsinki, Helsinki Inst Phys, POB 64, FI-00014 Helsinki, Finland
[3] STUK Radiat & Nucl Safety Author, Jokiniemenkuja 1, Vantaa 01370, Finland
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2023年 / 108卷
关键词
Lung counting; Voxel phantom; Geant4; simulation; Low-energy gamma-ray spectra;
D O I
10.1016/j.ejmp.2023.102573
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A Geant4 simulation package has been developed to investigate and test detector configurations for lung counting applications. The objective of this study was to measure radiation emitted from the human body and to make a qualitative comparison of the results of the simulation with an experiment. Experimental data were measured from a plastic phantom with a set of lungs containing 241Am activity. For comparison, simulations in which 241Am activity was uniformly distributed inside the lungs of the ICRP adult reference computational phantom were made. The attenuation of photons by the chest wall was simulated and from this photopeak ef-ficiency and photon transmission were calculated as a function of photon energy. The transmission of 59.5 keV gamma rays, characteristic of the decay of 241Am, was determined from the computational phantom as a function of the angular position of the detector. It was found that the simulated detector response corresponds well with that from an experiment. The simulated count rate below 100 keV was 10.0(7) % greater compared to the experimental measurement. It was observed that 58.3(4) % of photons are attenuated for energies below 100 keV by the chest wall. In the simulation, the transmission of 59.5 keV gamma rays varied from 13.8(2) % to 38.0(4) % as a function of the angular position of the detector. The results obtained from the simulations show a satisfactory agreement with experimental data and the package can be used in the development of future body counting applications and enables optimization of the detection geometry.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Photon Elastic Scattering Simulation: Validation and Improvements to Geant4
    Batic, Matej
    Hoff, Gabriela
    Pia, Maria Grazia
    Saracco, Paolo
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (04) : 1636 - 1664
  • [22] GEANT4 simulation of cyclotron radioisotope production in a solid target
    Poignant, F.
    Penfold, S.
    Asp, J.
    Takhar, P.
    Jackson, P.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (05): : 728 - 734
  • [23] Validation of Geant4 Simulation of Proton Energy Straggling: First Results
    Han, Min Cheol
    Pia, Maria Grazia
    Saracco, Paolo
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [24] Validation of a GEANT4 simulation of neutron stimulated emission computed tomography
    Kapadia, Anuj J.
    Harrawood, Brian P.
    Tourassi, Georgia D.
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [25] FAST NEUTRON MULTIPLICITY SIMULATION OF NUCLEAR MATERIAL BASED ON GEANT4
    Li, Sufen
    Qiu, Suizheng
    Zhang, Quanhu
    Huo, Yonggang
    Liu, Limin
    PROCEEDINGS OF THE 24TH INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING, 2016, VOL 4, 2016,
  • [26] GEANT4 Simulation of Photoneutron Spectrum from Medical Linear Accelerator
    Chernyaev, Alexander
    Belikhin, Mikhail
    Lykova, Ekaterina
    Shcherbakov, Alexey
    QUANTUM BEAM SCIENCE, 2023, 7 (03)
  • [27] Geometry and physics of the Geant4 toolkit for high and medium energy applications
    Apostolakis, J.
    Asai, M.
    Bogdanov, A. G.
    Burkhardt, H.
    Cosmo, G.
    Elles, S.
    Folger, G.
    Grichine, V. M.
    Gumplinger, P.
    Heikkinen, A.
    Hrivnacova, I.
    Ivanchenko, V. N.
    Jacquemier, J.
    Koi, T.
    Kokoulin, R. P.
    Kossov, M.
    Kurashige, H.
    McLaren, I.
    Link, O.
    Maire, M.
    Pokorski, W.
    Sasaki, T.
    Starkov, N.
    Urban, L.
    Wright, D. H.
    RADIATION PHYSICS AND CHEMISTRY, 2009, 78 (10) : 859 - 873
  • [28] Design of a controllable Am-Be neutron source: A Geant4 simulation
    Hu, Qiang
    Jia, Wenbao
    Hei, Daqian
    Li, Jiatong
    Cheng, Can
    Zhao, Dong
    APPLIED RADIATION AND ISOTOPES, 2021, 174
  • [29] Simulation of spread-out bragg peaks in proton beams using Geant4/TOPAS
    Velten, Christian
    Tome, Wolfgang A.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (04):
  • [30] Calculating the Outdoor Dose Deposited by γ-Photons Using GEANT4
    Argintaru, Danut
    Baban, Valerica
    Constantinescu, Eliodor
    ADVANCED TOPICS IN OPTOELECTRONICS, MICROELECTRONICS, AND NANOTECHNOLOGIES V, 2010, 7821