Effect of Sacropelvic Hardware on Axis and Center of Rotation of the Sacroiliac Joint: A Finite Element Study

被引:0
作者
Panico, Matteo [1 ,2 ]
Chande, Ruchi D. [3 ]
Polly, David W. [4 ]
Lindsey, Derek P. [3 ]
Villa, Tomaso M. T. [1 ,2 ]
Yerby, Scott A. [3 ]
Brayda-Bruno, Marco [1 ]
Galbusera, Fabio [5 ,6 ]
机构
[1] IRCCS Ist Ortoped Galeazzi, Milan, Italy
[2] Politecn Milan, Dept Chem Mat & Chem Engn Giulio Natta, Milan, Italy
[3] SI BONE Inc, Santa Clara, CA USA
[4] Univ Minnesota, Dept Orthopaed Surg, Minneapolis, MN USA
[5] Schulthess Clin, Spine Ctr, Zurich, Switzerland
[6] Schulthess Clin, Lengghalde 2, CH-8008 Zurich, Switzerland
关键词
sacropelvic fixation; triangular implants; S2 alar-iliac screws; ADULT SPINAL DEFORMITY; FIXATION; FUSION; PAIN; ANATOMY; LUMBAR; INSTRUMENTATION; COMPLICATIONS; MOVEMENTS; DIAGNOSIS;
D O I
10.14444/8387
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The sacroiliac joint (SIJ) transfers the load of the upper body to the lower extremities while allowing a variable physiological movement among individuals. The axis of rotation (AoR) and center of rotation (CoR) of the SIJ can be evaluated to analyze the stability of the SIJ, including when the sacrum is fixed. The purpose of this study was to determine how load intensity affects the SIJ for the intact model and to characterize how sacropelvic fixation performed with different techniques affects this joint.Methods: Five T10-pelvis models were used: (1) intact model; (2) pedicle screws and rods in T10 -S1; (3)pedicle screws and rods in T10 -S1, and bilateral S2 alar- iliac screws (S2AI); (4) pedicle screws and rods in T10 -S1, bilateral S2AI screws, and triangular implants inserted bilaterally in a sacral alar- iliac trajectory ; and (5) pedicle screws and rods in T10 -S1, bilateral S2AI screws, and 2 bilateral triangular implants inserted in a lateral trajectory. Outputs of these models under flexion-extension were compared: AoR and CoR of the SIJ at incremental steps from 0 to 7.5 Nm for the intact model and AoR and CoR of the SIJ for the instrumented models at 7.5 Nm.Results: The intact model was validated against an in vivo study by comparing range of motion and displacement of the sacrum. Increasing the load intensity for the intact model led to an increase of the rotation of the sacrum but did not change the CoR. Comparison among the instrumented models showed that sacropelvic fixation techniques reduced the rotation of the sacrum and stabilized the SIJ, in particular with triangular implants.Conclusion: The study outcomes suggest that increasing load intensity increases the rotation of the sacrum but does not influence the CoR, and use of sacropelvic fixation increases the stability of the SIJ, especially when triangular implants are employed.Clinical Relevance: The choice of the instrumentation strategy for sacropelvic fixation affects the stability of the construct in terms of both range of motion and axes of rotation, with direct consequences on the risk of failure and mobilization. Clinical studies should be performed to confirm these biomechanical findings.
引用
收藏
页码:122 / 131
页数:11
相关论文
共 50 条
  • [41] Effect of side holes in cervical fusion cages: a finite element analysis study
    Aslani, F. J.
    Hukins, D. W. L.
    Shepherd, D. E. T.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H10) : 986 - 992
  • [42] Effect of Percutaneous Endoscopic Lumbar Foraminoplasty of Different Facet Joint Portions on Lumbar Biomechanics: A Finite Element Analysis
    Yu, Yang
    Zhou, Qun
    Xie, Yi-zhou
    Wang, Xin-ling
    Fan, Xiao-hong
    Gu, Dang-wei
    Huang, Xue
    Wu, Wei-dong
    ORTHOPAEDIC SURGERY, 2020, 12 (04) : 1277 - 1284
  • [43] Effect of clenching on biomechanical response of human mandible and temporomandibular joint to traumatic force analyzed by finite element method
    Murakami, Kazuhiro
    Yamamoto, Kazuhiko
    Sugiura, Tsutomu
    Kawakami, Masayoshi
    Kang, Yu-Bong
    Tsutsumi, Sadami
    Kirita, Tadaaki
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2013, 18 (03): : E473 - E478
  • [44] Study protocol for a randomised controlled trial on the effect of local analgesia for pain relief after minimal invasive sacroiliac joint fusion: the ARTEMIS study
    Hermans, Sem M. M.
    Nellensteijn, Jorm M.
    van Santbrink, Henk
    Knoef, Rob
    Reinders, Mattheus K.
    Hoofwijk, Daisy M. N.
    Potters, Jan W.
    Movig, Kris L. L.
    Curfs, Inez
    van Hemert, Wouter L. W.
    BMJ OPEN, 2021, 11 (12):
  • [45] The implant thread as a retention element in cortical bone: the effect of thread size and thread profile: a finite element study
    Hansson, S
    Werke, M
    JOURNAL OF BIOMECHANICS, 2003, 36 (09) : 1247 - 1258
  • [46] The Influence of a Positioning Hex on Abutment Rotation in Tapered Internal Implants: A 3D Finite Element Model Study
    Hyun, Dong-Geun
    Kwon, Ho-Beom
    Lim, Young-Jun
    Koak, Jai-Young
    Kim, Myung-Joo
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (02) : 281 - 288
  • [47] Effect of vehicular vibrations on L-4 lumbar vertebrae - A finite element study
    Kishore, Y. S.
    Sreedhara, B. M.
    Manoj, A.
    Raveesh, R. M.
    Rakesh, B.
    Bhaskar, S.
    Kuntoji, Geetha
    Chethan, B. A.
    JOURNAL OF ORTHOPAEDICS, 2025, 63 : 109 - 115
  • [48] A finite element method study of the effect of vibration on the dynamic biomechanical response of the lumbar spine
    Zhu, Shuai
    Dong, RuiChun
    Liu, Zhong
    Liu, Hong
    Lu, ZhuangQi
    Guo, YunQiang
    CLINICAL BIOMECHANICS, 2024, 111
  • [49] Effect of posterolateral disc replacement on kinematics and stress distribution in the lumbar spine: A finite element study
    Wang, Wenhai
    Zhang, Henry
    Sadeghipour, Keya
    Baran, George
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (03) : 357 - 364
  • [50] Effect of a novel compressible artificial disk on biomechanical performance of cervical spine: A finite element study
    Chen, Jou-Wen
    Chen, Wen-Chuan
    Lai, Yu-Shu
    Chang, Chia-Ming
    Wang, Shih-Tien
    ADVANCES IN MECHANICAL ENGINEERING, 2015, 7 (08) : 1 - 5