摘要
Infrared (IR) or thermal cameras are being increasingly used in livestock research and management. An IR camera's accuracy is specified over its entire surface temperature measurement range, whereas in livestock research and management, a narrower range suffices. A camera's accuracy could be higher in a narrower range of temperatures. Hence, a novel low-cost method was used to calculate the FLIR E8 camera's accuracy in a range of 24 degrees C to 37 degrees C, representative of surface temperature of poultry birds. Sensitivity analyses were also performed to evaluate the impact of three user spec-ified parameters, namely, emissivity (epsilon), distance between camera and surface (d), and reflected air temperature (tair). A linear regression model was used to correct the camera's absolute error of 2.8 degrees C (greater than its published error). How-ever, the camera possessed precision and hence, repeatability. The IR camera was moderately sensitive to epsilon, and slightly sensitive to tair and d, but its error could increase with the difference between the measured and assumed tair values. Attention is required to accurately characterize epsilon and tair. This simple calibration method can reduce cost and could improve accuracy in a narrower temperature range than the IR camera's published range, which could be useful for applied research.
引用
收藏
页码:529 / 534
页数:6
相关论文
共 50 条
  • [1] The Conception of Test Fields for Fast Geometric Calibration of the FLIR VUE PRO Thermal Camera for Low-Cost UAV Applications
    Fryskowska-Skibniewska, Anna
    Delis, Paulina
    Kedzierski, Michal
    Matusiak, Dominik
    SENSORS, 2022, 22 (07)
  • [2] A Low-Cost Terahertz Camera
    Blanchard, Francois
    Nkeck, Joel Edouard
    Matte, Dominique
    Nechache, Riad
    Cooke, David G.
    APPLIED SCIENCES-BASEL, 2019, 9 (12):
  • [3] Laboratory Calibration and Performance Evaluation of Low-Cost Capacitive and Very Low-Cost Resistive Soil Moisture Sensors
    Adla, Soham
    Rai, Neeraj Kumar
    Karumanchi, Sri Harsha
    Tripathi, Shivam
    Disse, Markus
    Pande, Saket
    SENSORS, 2020, 20 (02)
  • [4] Approaches to low-cost infrared sensing
    Reyner, Charles J.
    Ariyawansa, Gamini
    Claflin, Bruce
    Duran, Joshua M.
    Grzybowski, Gordon J.
    APPLIED OPTICS, 2021, 60 (25) : G162 - G169
  • [5] Low-Cost Reflectance-Based Method for the Radiometric Calibration of Kinect 2
    Rodriguez-Gonzalvez, Pablo
    Gonzalez-Aguilera, Diego
    Gonzalez-Jorge, Higinio
    Hernandez-Lopez, David
    IEEE SENSORS JOURNAL, 2016, 16 (07) : 1975 - 1985
  • [6] Assessing the Performance of a Low-Cost Thermal Camera in Proximal and Aerial Conditions
    Acorsi, Matheus Gabriel
    Gimenez, Leandro Maria
    Martello, Mauricio
    REMOTE SENSING, 2020, 12 (21) : 1 - 24
  • [7] A Combined Physical and Mathematical Calibration Method for Low-Cost Cameras in the Air and Underwater Environment
    Ma, Zhenling
    Zhong, Xu
    Xie, Hong
    Zhou, Yongjun
    Chen, Yuan
    Wang, Jiali
    SENSORS, 2023, 23 (04)
  • [8] The syringe potentiometer: a low-cost device for pneumotachograph calibration
    Cross, Troy J.
    Kelley, Eli F.
    Hardy, Tim A.
    Isautier, Jennifer M. J.
    Johnson, Bruce D.
    JOURNAL OF APPLIED PHYSIOLOGY, 2019, 127 (04) : 1150 - 1162
  • [9] A low-cost towed video camera system for underwater surveys: comparative performance with standard methodology
    Trobbiani, G. A.
    Irigoyen, A.
    Venerus, L. A.
    Fiorda, P. M.
    Parma, A. M.
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2018, 190 (11)
  • [10] LOW-COST CLOCK SYNCHRONIZATION
    DRUMMOND, R
    BABAOGLU, O
    DISTRIBUTED COMPUTING, 1993, 6 (04) : 193 - 203