Feasibility of His bundle pacing facilitated by EASI derived 12-lead ECG

被引:0
作者
Langanke, Alexander [1 ]
Andreas, Klaus [1 ]
机构
[1] Med Klin 1, Varisano Klinikum Bad Soden, Kronberger Str 36, D-65812 Bad Soden, Germany
关键词
Electrocardiogram; Vector-ECG; His bundle pacing; Conduction system pacing; EASI; ELECTROCARDIOGRAM; IMPLANTATION; STANDARD; SOCIETY; SYSTEM;
D O I
10.1016/j.jelectrocard.2023.10.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: His bundle pacing (HBP) has become popular in recent years as a more physiological alternative to conventional right ventricular pacing. Implantation requires 12-lead ECG during surgery, which is not readily available in a standard operating room. Often but not always HBP is performed in an electrophysiology lab. EASI is a reduced lead system which enables derived 12-lead ECG. EASI derived 12-lead ECGs on modern tablet computers offer a more mobile and lightweight ECG solution which does not obstruct fluoroscopy during implantation. This case series aims to compare standard 12-lead ECG to EASI derived 12-lead ECG in patients undergoing HBP implantation. Methods and results: A total of 11 patients received permanent HBP guided only by fluoroscopy, a pacing system analyzer (Medtronic CareLink SmartSync Device Manager) and EASI derived 12-lead ECG (CardioSecur Pro). During the first postoperative device interrogation HBP criteria, as defined in the EHRA consensus paper on conduction system pacing, were evaluated with the EASI derived system as well as a standard 12-lead ECG and compared to each other. There was perfect agreement with regards to these criteria which lead to identical conclusions in all cases. Conclusion: HBP implantation can be performed with EASI derived 12-lead ECG instead of conventional 12-lead ECG. Criteria for discriminating between selective His bundle, non-selective His bundle or myocardial capture alone are clearly visible in the EASI derived ECG leading to the same conclusion when compared to standard 12-lead ECG. Compared to a conventional 12-lead ECG the EASI system offers a leaner setup with less visual obstruction on fluoroscopy.
引用
收藏
页码:272 / 276
页数:5
相关论文
共 10 条
  • [1] Burri H, 2023, EUROPACE, V25, P1208, DOI 10.1093/europace/euad043
  • [2] Electrocardiographic Analysis for His Bundle Pacing at Implantation and Follow-Up
    Burri, Haran
    Jastrzebski, Marek
    Vijayaraman, Pugazhendhi
    [J]. JACC-CLINICAL ELECTROPHYSIOLOGY, 2020, 6 (07) : 883 - 900
  • [3] DERIVING THE 12-LEAD ELECTROCARDIOGRAM FROM 4 (EASI) ELECTRODES
    DOWER, GE
    YAKUSH, A
    NAZZAL, SB
    JUTZY, RV
    RUIZ, CE
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 : S182 - S187
  • [4] Comparison of standard and derived 12-lead electrocardiograms for diagnosis of coronary angioplasty-induced myocardial ischemia
    Drew, BJ
    Adams, MG
    Pelter, MM
    Wung, SF
    Caldwell, MA
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (05) : 639 - 644
  • [5] Comparison of a new reduced lead set ECG with the standard ECG for diagnosing cardiac arrhythmias and myocardial ischemia
    Drew, BJ
    Pelter, MM
    Brodnick, DE
    Yadav, AV
    Dempel, D
    Adams, MG
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2002, 35 : 13 - 21
  • [6] COMPARISON OF A VECTORCARDIOGRAPHICALLY DERIVED 12-LEAD ELECTROCARDIOGRAM WITH THE CONVENTIONAL ELECTROCARDIOGRAM DURING WIDE QRS COMPLEX TACHYCARDIA, AND ITS POTENTIAL APPLICATION FOR CONTINUOUS BEDSIDE MONITORING
    DREW, BJ
    SCHEINMAN, MM
    EVANS, GT
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (06) : 612 - 618
  • [7] 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA)
    Glikson, Michael
    Nielsen, Jens Cosedis
    Kronborg, Mads Brix
    Michowitz, Yoav
    Auricchio, Angelo
    Barbash, Israel Moshe
    Barrabes, Jose A.
    Boriani, Giuseppe
    Braunschweig, Frieder
    Brignole, Michele
    Burri, Haran
    Coats, Andrew J. S.
    Deharo, Jean-Claude
    Delgado, Victoria
    Diller, Gerhard-Paul
    Israel, Carsten W.
    Keren, Andre
    Knops, Reinoud E.
    Kotecha, Dipak
    Leclercq, Christophe
    Merkely, Bela
    Starck, Christoph
    Thylen, Ingela
    Tolosana, Maria
    [J]. EUROPACE, 2022, 24 (01): : 71 - +
  • [8] Smartphone-derived multichannel electrocardiogram for exercise stress testing
    Langanke, Alexander
    Reifart, Nicolaus
    Reifart, Joerg
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2021, 69 : 74 - 81
  • [9] PERVENOUS PACING OF SPECIALIZED CONDUCTING SYSTEM IN MAN . HIS BUNDLE AND A-V NODAL STIMULATION
    NARULA, OS
    SCHERLAG, BJ
    SAMET, P
    [J]. CIRCULATION, 1970, 41 (01) : 77 - &
  • [10] Zhu L, 2023, medRxiv, DOI [10.1101/2023.05.22.23290368.2023.05.22.23290368, DOI 10.1101/2023.05.22.23290368.2023.05.22.23290368]