Finite element analysis comparing a PEEK posterior fixation device versus pedicle screws for lumbar fusion

被引:5
作者
Eastlack, Robert K. [1 ]
Nunley, Pierce D. [2 ]
Poelstra, Kornelis A. [3 ]
Vaccaro, Alexander R. [3 ]
Stone, Marcus [2 ]
Miller, Larry E. [4 ]
Legay, Pierre [5 ]
Clin, Julien [5 ]
Agarwal, Aakash [6 ]
机构
[1] Scripps Clin, Dept Orthopaed Surg, San Diego, CA USA
[2] Spine Inst Louisiana, Shreveport, LA USA
[3] Thomas Jefferson Univ, Rothman Inst, Dept Orthopaed Surg, Philadelphia, PA USA
[4] Miller Sci, 3101 Browns Mill Rd,Ste 6,311, Johnson City, TN 37604 USA
[5] Numalog Inc, Montreal, PQ, Canada
[6] Univ Toledo, Dept Bioengn & Orthopaed Surg, Toledo, OH USA
关键词
Karma; Posterior fixation; Lumbar spine; Pedicle screw; Cortical fixation; BIOMECHANICAL EVALUATION; SPINE; STRENGTH; INSTRUMENTATION; BEHAVIOR; DENSITY; DISEASE;
D O I
10.1186/s13018-023-04349-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundPedicle screw loosening and breakage are common causes of revision surgery after lumbar fusion. Thus, there remains a continued need for supplemental fixation options that offer immediate stability without the associated failure modes. This finite element analysis compared the biomechanical properties of a novel cortico-pedicular posterior fixation (CPPF) device with those of a conventional pedicle screw system (PSS).MethodsThe CPPF device is a polyetheretherketone strap providing circumferential cortical fixation for lumbar fusion procedures via an arcuate tunnel. Using a validated finite element model, we compared the stability and load transfer characteristics of CPPF to intact conditions under a 415 N follower load and PSS conditions under a 222 N preload. Depending on the instrumented levels, two different interbody devices were used: a lateral lumbar interbody device at L4-5 or an anterior lumbar interbody device at L5-S1. Primary outcomes included range of motion of the functional spinal units and anterior load transfer, defined as the total load through the disk and interbody device after functional motion and follower load application.ResultsAcross all combinations of interbody devices and lumbar levels evaluated, CPPF consistently demonstrated significant reductions in flexion (ranging from 90 to 98%), extension (ranging from 88 to 94%), lateral bending (ranging from 75 to 80%), and torsion (ranging from 77 to 86%) compared to the intact spine. Stability provided by the CPPF device was comparable to PSS in all simulations (range of motion within 0.5 degrees for flexion-extension, 0.6 degrees for lateral bending, and 0.5 degrees for torsion). The total anterior load transfer was higher with CPPF versus PSS, with differences across all tested conditions ranging from 128 to 258 N during flexion, 89-323 N during extension, 135-377 N during lateral bending, 95-258 N during torsion, and 82-250 N during standing.ConclusionUnder the modeled conditions, cortico-pedicular fixation for supplementing anterior or lateral interbody devices between L4 and S1 resulted in comparable stability based on range of motion measures and less anterior column stress shielding based on total anterior load transfer measures compared to PSS. Clinical studies are needed to confirm these finite element analysis findings.
引用
收藏
页数:10
相关论文
共 33 条
[1]   Pedicle Screws Loosening in Patients With Degenerative Diseases of the Lumbar Spine: Potential Risk Factors and Relative Contribution [J].
Bokov, Andrey ;
Bulkin, Anatoliy ;
Aleynik, Alexander ;
Kutlaeva, Marina ;
Mlyavykh, Sergey .
GLOBAL SPINE JOURNAL, 2019, 9 (01) :55-61
[2]   Lumbar decompression and fusion in elderly osteoporotic patients: A prospective study using less rigid titanium rod fixation [J].
Cavagna, Remi ;
Tournier, Clement ;
Aunoble, Stephane ;
Bouler, Jean M. ;
Antonietti, Pierre ;
Ronai, Martin ;
Le Huec, Jean-Charles .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2008, 21 (02) :86-91
[3]   Segment inertial properties of Chinese adults determined from magnetic resonance imaging [J].
Cheng, CK ;
Chen, HH ;
Chen, CS ;
Lee, CL ;
Chen, CY .
CLINICAL BIOMECHANICS, 2000, 15 (08) :559-566
[4]   Biomechanical Comparison of the Load-Sharing Capacity of High and Low Implant Density Constructs With Three Types of Pedicle Screws for the Instrumentation of Adolescent Idiopathic Scoliosis [J].
Clin, Julien ;
Le Naveaux, Franck ;
Driscoll, Mark ;
Mac-Thiong, Jean-Marc ;
Labelle, Hubert ;
Parent, Stefan ;
Shah, Suken A. ;
Lonner, Baron S. ;
Newton, Peter O. ;
Serhan, Hassan .
SPINE DEFORMITY, 2019, 7 (01) :2-10
[5]   Biomechanical assessment of reduction forces measured during scoliotic instrumentation using two different screw designs [J].
Driscoll M. ;
Mac-Thiong J.-M. ;
Labelle H. ;
Slivka M. ;
Stad S. ;
Parent S. .
Spine Deformity, 2013, 1 (2) :94-101
[6]   Development of a Detailed Volumetric Finite Element Model of the Spine to Simulate Surgical Correction of Spinal Deformities [J].
Driscoll, Mark ;
Mac-Thiong, Jean-Marc ;
Labelle, Hubert ;
Parent, Stefan .
BIOMED RESEARCH INTERNATIONAL, 2013, 2013
[7]  
Eberle S, 2007, ANSYS C 25 CADFEM US
[8]   Revision Rate of Misplaced Pedicle Screws of the Thoracolumbar Spine-Comparison of Three-Dimensional Fluoroscopy Navigation with Freehand Placement: A Systematic Analysis and Review of the Literature [J].
Fichtner, Jens ;
Hofmann, Nicole ;
Rienmueller, Anna ;
Buchmann, Niels ;
Gempt, Jens ;
Kirschke, Jan S. ;
Ringel, Florian ;
Meyer, Bernhard ;
Ryang, Yu-Mi .
WORLD NEUROSURGERY, 2018, 109 :E24-E32
[9]   The use of pedicle-screw internal fixation for the operative treatment of spinal disorders [J].
Gaines, RW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (10) :1458-1476
[10]   Pedicle screw loosening: a clinically relevant complication? [J].
Galbusera, Fabio ;
Volkheimer, David ;
Reitmaier, Sandra ;
Berger-Roscher, Nikolaus ;
Kienle, Annette ;
Wilke, Hans-Joachim .
EUROPEAN SPINE JOURNAL, 2015, 24 (05) :1005-1016