A new breast phantom suitable for digital mammography, contrast-enhanced digital mammography and digital breast tomosynthesis

被引:5
作者
Zhang, Changsheng [1 ]
Fu, Jian [1 ,2 ,3 ]
机构
[1] Beihang Univ, Sch Mech Engn & Automat, Beijing 100190, Peoples R China
[2] Beihang Univ, Jiangxi Res Inst, Nanchang 330000, Peoples R China
[3] Beihang Univ, Ningbo Inst Technol, Ningbo 315000, Peoples R China
基金
中国国家自然科学基金;
关键词
breast test phantom; mammography; imaging quality-dose relationship; tissue equivalent; MODULATION TRANSFER-FUNCTION; IMAGE QUALITY; ACCREDITATION; PERFORMANCE; 2D;
D O I
10.1088/1361-6560/acb636
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our objective is to report a new breast phantom that provides the objective assessment for three types of clinical mammography, i.e. digital mammography (DM), contrast-enhanced digital mammography (CEDM), and digital breast tomosynthesis (DBT). The tissue-equivalent materials are used to represent the corresponding tissue, and the layer-by-layer structure with separate regions is designed for image quality assessment of different mammography modes. For DM imaging, substitutes for microcalcifications and fibroglandular tissue of different sizes are used to simulate the conventional breast. For CEDM imaging, the tumor module that can be injected with imaging contrast agents is adopted to distinguish normal tissue and diseased tissue in the dense breast. For DBT imaging, the overlapping breast mass module with multiple layers is designed to perform the layer-by-layer imaging of overlapping tissue. In addition, the quantitative assessment module of image quality is designed based on contrast-to-noise ratio, modulation transfer function and artifact spread function. This phantom allows image quality to be evaluated objectively for three different types of the clinical mammography, while it provides an effective tool for optimizing the dose-image quality relationship of patients.
引用
收藏
页数:17
相关论文
共 47 条
[1]   Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations [J].
Amini, Iman ;
Akhlaghi, Parisa ;
Sarbakhsh, Parvin .
RADIATION PHYSICS AND CHEMISTRY, 2018, 150 :51-57
[2]   MAMMOGRAPHY ACCREDITATION AND EQUIPMENT PERFORMANCE [J].
BARNES, GT ;
HENDRICK, RE .
RADIOGRAPHICS, 1994, 14 (01) :129-138
[3]   Digital breast tomosynthesis using an amorphous selenium flat panel detector [J].
Bissonnette, M ;
Hansroul, M ;
Masson, E ;
Savard, S ;
Cadieux, S ;
Warmoes, P ;
Gravel, D ;
Agopyan, J ;
Polischuk, B ;
Haerer, W ;
Mertelmeier, T ;
Lo, JY ;
Chen, Y ;
Dobbins, JT ;
Jesneck, JL ;
Singh, S .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :529-540
[4]   Effect of injection rate on contrast-enhanced MR angiography image quality: Modulation transfer function analysis [J].
Clark, Toshimasa J. ;
Wilson, Gregory J. ;
Maki, Jeffrey H. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (01) :357-369
[5]  
Commission I E, 2015, MED EL EQ CHAR DIG X
[6]  
Commission I E, 2003, MED EL EQ CHAR DIG X
[7]   A PRELIMINARY INVESTIGATION OF THE IMAGING PERFORMANCE OF PHOTOSTIMULABLE PHOSPHOR COMPUTED RADIOGRAPHY USING A NEW DESIGN OF MAMMOGRAPHIC QUALITY-CONTROL TEST OBJECT [J].
COWEN, AR ;
BRETTLE, DS ;
COLEMAN, NJ ;
PARKIN, GJS .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (774) :528-535
[8]   SIGNAL AND NOISE IN MODULATION TRANSFER-FUNCTION DETERMINATIONS USING THE SLIT, WIRE, AND EDGE TECHNIQUES [J].
CUNNINGHAM, IA ;
REID, BK .
MEDICAL PHYSICS, 1992, 19 (04) :1037-1044
[9]  
DeWerd LA., 2014, The Phantoms of Medical and Health Physics, DOI [DOI 10.1007/978-1-4614-8304-5, 10.1007/978-1-4614-8304-5]
[10]   A clinical audit programme for diagnostic radiology: the approach adopted by the International Atomic Energy Agency [J].
Faulkner, K. ;
Jarvinen, H. ;
Butler, P. ;
McLean, I. D. ;
Pentecost, M. ;
Rickard, M. ;
Abdullah, B. .
RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) :418-421