Blood flow effects in a patient with a thoracic aortic endovascular prosthesis

被引:0
|
作者
Tello, Juan P. [1 ]
Velez, Juan C. [1 ]
Cadena, Alberto [2 ]
Jutinico, Andres [3 ]
Pardo, Mauricio [1 ]
Percybrooks, Winston [1 ]
机构
[1] Univ Norte, Km 5 Via Puerto Colombia, Barranquilla, Colombia
[2] Clin Costa, Barranquilla, Colombia
[3] Univ Distrital Francisco Jose Caldas, Bogota, Colombia
关键词
Blood flow; Velocity profiles; Pressure distributions; Image processing; Aorta; Cross section; Endovascular prosthesis; STENT-GRAFT MODELS; SHEAR-STRESS; NUMERICAL-SIMULATION; HEMODYNAMICS; RESTENOSIS; STENOSIS; ARTERIES; REPAIR;
D O I
10.1016/j.heliyon.2024.e26355
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This work analyzes hemodynamic phenomena within the aorta of two elderly patients and their impact on blood flow behavior, particularly affected by an endovascular prosthesis in one of them (Patient II). Computational Fluid Dynamics (CFD) was utilized for this study, involving measurements of velocity, pressure, and wall shear stress (WSS) at various time points during the third cardiac cycle, at specific positions within two cross sections of the thoracic aorta. The first cross-section (Cross -Section 1, CS1) is located before the initial fluid bifurcation, just before the right subclavian artery. The second cross-section (Cross -Section 2, CS2) is situated immediately after the left subclavian artery. The results reveal that, under regular aortic geometries, velocity and pressure magnitudes follow the principles of fluid dynamics, displaying variations. However, in Patient II, an endoprosthesis near the CS2 and the proximal border of the endoprosthesis significantly disrupts fluid behavior owing to the pulsatile flow. The cross-sectional areas of Patient I are smaller than those of Patient II, leading to higher flow magnitudes. Although in CS1 of Patient I, there is considerable variability in velocity magnitudes, they exhibit a more uniform and predictable transition. In contrast, CS2 of Patient II, where magnitude variation is also high, displays irregular fluid behavior due to the endoprosthesis presence. This cross-section coincides with the border of the fluid bifurcation. Additionally, the irregular geometry caused by endovascular aneurysm repair contributes to flow disruption as the endoprosthesis adjusts to the endothelium, reshaping itself to conform with the vessel wall. In this context, significant alterations in velocity values, pressure differentials fluctuating by up to 10%, and low wall shear stress indicate the pronounced influence of the endovascular prosthesis on blood flow behavior. These flow disturbances, when compounded by the heart rate, can potentially lead to changes in vascular anatomy and displacement, resulting in a disruption of the prosthesis -endothelium continuity and thereby causing clinical complications in the patient.
引用
收藏
页数:19
相关论文
共 50 条
  • [21] Influence of Type of Aortic Valve Prosthesis on Coronary Blood Flow Velocity
    Jelenc, Matija
    Juvan, Katja Azman
    Medvescek, Nadja Tatjana Ruzic
    Gersak, Borut
    HEART SURGERY FORUM, 2013, 16 (01) : E8 - E14
  • [22] Endovascular surgery for thoracic aortic pathologies involving the aortic arch
    Lu, Heng
    Huang, Ling-chen
    Chen, Liang-wan
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [23] Thoracic endovascular aortic repair for avulsion of aortic branches in a trauma patient requiring resuscitative thoracotomy: a case report
    Tanikawa, Atsushi
    Sato, Takeaki
    Fujita, Motoo
    Tsuchiya, Chieri
    Katsuta, Ken
    Suzuki, Yusuke
    Kumagai, Kiichiro
    Saiki, Yoshikatsu
    Kushimoto, Shigeki
    SURGICAL CASE REPORTS, 2022, 8 (01)
  • [24] Endovascular Stent in Traumatic Thoracic Aortic Dissection
    Jang, Mi Ok
    Kim, Ju Han
    Oh, Sang Ki
    Lee, Min Goo
    Park, Keun Ho
    Sim, Doo Sun
    Hong, Young Joon
    Ahn, Youngkeun
    Jeong, Myung Ho
    KOREAN CIRCULATION JOURNAL, 2012, 42 (05) : 341 - 344
  • [25] Thoracic endovascular aortic repair for avulsion of aortic branches in a trauma patient requiring resuscitative thoracotomy: a case report
    Atsushi Tanikawa
    Takeaki Sato
    Motoo Fujita
    Chieri Tsuchiya
    Ken Katsuta
    Yusuke Suzuki
    Kiichiro Kumagai
    Yoshikatsu Saiki
    Shigeki Kushimoto
    Surgical Case Reports, 8
  • [26] Endovascular Repair of Blunt Thoracic Aortic Injuries
    Brinster, Derek R.
    SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2009, 21 (04) : 393 - 398
  • [27] Thoracic Aortic Aneurysms and Dissections: Endovascular Treatment
    Baril, Donald T.
    Cho, Jae S.
    Chaer, Rabih A.
    Makaroun, Michel S.
    MOUNT SINAI JOURNAL OF MEDICINE, 2010, 77 (03): : 256 - 269
  • [28] Endovascular Treatment of Traumatic Thoracic Aortic Injuries
    Feezor, Robert J.
    Hess, Philip J., Jr.
    Martin, Tomas D.
    Klodell, Charles T.
    Beaver, Thomas M.
    Lottenberg, Lawrence
    Martin, Larry C.
    Lee, W. Anthony
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2009, 208 (04) : 510 - 516
  • [29] Endovascular treatment of thoracic aortic aneurysms:: a review
    Thurnher, SA
    Grabenwöger, M
    EUROPEAN RADIOLOGY, 2002, 12 (06) : 1370 - 1387
  • [30] A Review of the Endovascular Management of Thoracic Aortic Pathology
    Bell, David
    Bassin, Levi
    Neale, Michael
    Brady, Peter
    HEART LUNG AND CIRCULATION, 2015, 24 (12) : 1211 - 1215