Comparison of the optical behaviour of five different multifocal diffractive intraocular lenses in a model eye

被引:11
作者
Can, Efe [1 ]
Senel, Esat Can [1 ]
Holmstroem, Sven T. S. [1 ]
Pinero, David P. [2 ,3 ]
机构
[1] VSY Biotechnol GMBH, R&D Dept, Esslinger Str 7, D-70771 Leinfelden Echterdingen, Germany
[2] Univ Alicante, Dept Opt Pharmacol & Anat, Grp Opt & Visual Percept, Crta San Vicente Del Raspeig S-N, Alicante 03690, Spain
[3] Vithas Medimar Int Hosp, Dept Ophthalmol, Alicante, Spain
来源
SCIENTIFIC REPORTS | 2023年 / 13卷 / 01期
关键词
EXTENDED DEPTH; BENCH PERFORMANCE; TRIFOCAL IOL; FOCUS IOL; DECENTRATION; TILT; QUALITY; DESIGN; SIZE;
D O I
10.1038/s41598-023-47102-y
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The purpose of this study was to investigate and compare the optical performance of five trifocal intraocular lenses (IOLs) following the ISO 11979-2 standards, analysing the impact of tilt and decentration. Five different diffractive trifocal IOLs were evaluated in this experimental study: Acriva Trinova (VSY-Biotechnology) (AT), FineVision HP (PhysIOL) (FVHP), AT LISA tri 839 MP (Zeiss) (ATLT), PanOptix TFNT00 IOL (Alcon) (PO), and Tecnis Synergy (J&J Vision) (TS). In-vitro optical quality analysis of them was performed with the Lambda PMTF system that has an aberration neutral cornea model (Lambda-X Ophthalmics). Measurements were performed on-axis, with 5 degrees of IOL tilt and with 0.5 mm of IOL decentration using 543-nm monochromatic light. Finally, IOL dimensions and diffractive disk profile inspection was performed using the VisIOLA system (Rotlex). On-axis measurements showed a far through-focus MTF > 0.3 at 3 mm aperture, except for TS. FVHP and PO showed better far MTFs for larger apertures (3.75 mm and 4.5 mm) while AT showed good intermediate and near vision for such apertures. With 5 degrees of IOL tilt, the better optical performance at all distances was found with AT for medium-sized pupils (3 mm) and an important reduction of MTF was found for ATLT and PO, especially in the intermediate focus. The induction of 0.5 mm of IOL decentration especially affected the intermediate focus of ATLT and TS and the far focus of FVHP and PO. IOL dimensions and diffractive profile were consistent with those described by the manufacturer. In conclusion, there are differences in the optical performance according to the pupil aperture of the five trifocal IOLs evaluated and this should be considered in clinical practice when selecting the most appropriate implant in each specific case. IOL tilt and decentration can affect significantly in most of the designs evaluated the performance of the IOL at intermediate vision range. It should be noted that measurements were made with an aberration-free cornea, being necessary future studies analysing the impact of different levels of corneal aberrations.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Comparatively analysing the postoperative optical performance of different intraocular lenses: a prospective observational study
    Guo, Shuanglin
    Huang, Hao
    Li, Bowen
    Huang, Mansha
    Gao, Lu
    Chen, Jingyi
    Zeng, Yuying
    Yang, Ye
    Liu, Lin
    Cheng, Lu
    Yao, Siyang
    Cheng, Hao
    BMC OPHTHALMOLOGY, 2024, 24 (01)
  • [42] Comparison of visual outcomes after bilateral implantation of two intraocular lenses with distinct diffractive optics
    de Medeiros, Andre Lins
    Saraiva, Flavio Jones
    Iguma, Camila Ishii
    Kniggendorf, Danilo Varela
    Alves, Guilherme
    Pereira Dias Chaves, Mario Augusto
    Vilar, Cesar
    Pimenta Motta, Antonio Francisco
    Carricondo, Pedro Carlos
    Nakano, Celso Takashi
    Nose, Walton
    Hida, Wilson Takashi
    CLINICAL OPHTHALMOLOGY, 2019, 13 : 1657 - 1663
  • [43] Optical Quality Differences Between Three Multifocal Intraocular Lenses: Bifocal Low Add, Bifocal Moderate Add, and Trifocal
    Madrid-Costa, David
    Ruiz-Alcocer, Javier
    Ferrer-Blasco, Teresa
    Garcia-Lazaro, Santiago
    Montes-Mico, Robert
    JOURNAL OF REFRACTIVE SURGERY, 2013, 29 (11) : 749 - +
  • [44] Prospective analysis of visual outcomes using apodized, diffractive multifocal intraocular lenses following phacoemulsification for cataract or clear lens extraction
    Altaie, Rasha
    Ring, C. Peter
    Morarji, Jasha
    Patel, Dipika V.
    McGhee, Charles N. J.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (02) : 148 - 154
  • [45] Comparison of Five Presbyopia-Correcting Intraocular Lenses: Optical-Bench Assessment with Visual-Quality Simulation
    Labuz, Grzegorz
    Yan, Weijia
    Baur, Isabella D. D.
    Khoramnia, Ramin
    Auffarth, Gerd U. U.
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (07)
  • [46] Comparison between OPD-Scan results and visual outcomes of monofocal and multifocal intraocular lenses
    Hida, Wilson Takashi
    Pimenta Motta, Antonio Francisco
    Kara-Jose Junior, Newton
    Alves, Emerson
    Tadeu, Marcel
    Cordeiro, Livio Neiva
    Nakano, Celso Takashi
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2009, 72 (04) : 526 - 532
  • [47] Laboratory Evaluation of Halos and Through-Focus Performance of Three Different Multifocal Intraocular Lenses
    Liu, Xiaomin
    Wu, Xiaoming
    Huang, Yusen
    JOURNAL OF REFRACTIVE SURGERY, 2022, 38 (09) : 552 - +
  • [48] Effects of decentration and tilt at different orientations on the optical performance of a rotationally asymmetric multifocal intraocular lens
    Liu, Xiaomin
    Xie, Lixin
    Huang, Yusen
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (04) : 507 - 514
  • [49] Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses
    Lasta, Michael
    Mihaltz, Kata
    Kovacs, Illes
    Vecsei-Marlovits, Pia Veronika
    JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [50] Effect of misalignment at different orientations associated with angle κ on optical performance of aspheric intraocular lenses with different surface designs
    Yuan, Bo
    Li, Jun
    Song, Hui
    APPLIED OPTICS, 2021, 60 (20) : 5917 - 5924