Comparison of the optical behaviour of five different multifocal diffractive intraocular lenses in a model eye

被引:10
|
作者
Can, Efe [1 ]
Senel, Esat Can [1 ]
Holmstroem, Sven T. S. [1 ]
Pinero, David P. [2 ,3 ]
机构
[1] VSY Biotechnol GMBH, R&D Dept, Esslinger Str 7, D-70771 Leinfelden Echterdingen, Germany
[2] Univ Alicante, Dept Opt Pharmacol & Anat, Grp Opt & Visual Percept, Crta San Vicente Del Raspeig S-N, Alicante 03690, Spain
[3] Vithas Medimar Int Hosp, Dept Ophthalmol, Alicante, Spain
来源
SCIENTIFIC REPORTS | 2023年 / 13卷 / 01期
关键词
EXTENDED DEPTH; BENCH PERFORMANCE; TRIFOCAL IOL; FOCUS IOL; DECENTRATION; TILT; QUALITY; DESIGN; SIZE;
D O I
10.1038/s41598-023-47102-y
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The purpose of this study was to investigate and compare the optical performance of five trifocal intraocular lenses (IOLs) following the ISO 11979-2 standards, analysing the impact of tilt and decentration. Five different diffractive trifocal IOLs were evaluated in this experimental study: Acriva Trinova (VSY-Biotechnology) (AT), FineVision HP (PhysIOL) (FVHP), AT LISA tri 839 MP (Zeiss) (ATLT), PanOptix TFNT00 IOL (Alcon) (PO), and Tecnis Synergy (J&J Vision) (TS). In-vitro optical quality analysis of them was performed with the Lambda PMTF system that has an aberration neutral cornea model (Lambda-X Ophthalmics). Measurements were performed on-axis, with 5 degrees of IOL tilt and with 0.5 mm of IOL decentration using 543-nm monochromatic light. Finally, IOL dimensions and diffractive disk profile inspection was performed using the VisIOLA system (Rotlex). On-axis measurements showed a far through-focus MTF > 0.3 at 3 mm aperture, except for TS. FVHP and PO showed better far MTFs for larger apertures (3.75 mm and 4.5 mm) while AT showed good intermediate and near vision for such apertures. With 5 degrees of IOL tilt, the better optical performance at all distances was found with AT for medium-sized pupils (3 mm) and an important reduction of MTF was found for ATLT and PO, especially in the intermediate focus. The induction of 0.5 mm of IOL decentration especially affected the intermediate focus of ATLT and TS and the far focus of FVHP and PO. IOL dimensions and diffractive profile were consistent with those described by the manufacturer. In conclusion, there are differences in the optical performance according to the pupil aperture of the five trifocal IOLs evaluated and this should be considered in clinical practice when selecting the most appropriate implant in each specific case. IOL tilt and decentration can affect significantly in most of the designs evaluated the performance of the IOL at intermediate vision range. It should be noted that measurements were made with an aberration-free cornea, being necessary future studies analysing the impact of different levels of corneal aberrations.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Comparison of the optical behaviour of five different multifocal diffractive intraocular lenses in a model eye
    Efe Can
    Esat Can Senel
    Sven T. S. Holmström
    David P. Piñero
    Scientific Reports, 13 (1)
  • [2] Assessing the optical performance of multifocal (diffractive) intraocular lenses
    Eppig, Timo
    Scholz, Katja
    Langenbucher, Achim
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2008, 28 (05) : 467 - 474
  • [3] Wavefront measurements of diffractive and refractive multifocal intraocular lenses in an artificial eye
    Campbell, Charles E.
    JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (03) : 308 - 311
  • [4] Different reading ability with diffractive and refractive multifocal intraocular lenses
    Richter-Mueksch, S
    Weghaupt, H
    Skorpik, C
    Velikay-Parel, M
    Stifter, E
    Radner, W
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U88 - U88
  • [5] Comparison of Diffractive and Refractive Multifocal Intraocular Lenses in Presbyopia Treatment
    Barisic, Ante
    Dekaris, Iva
    Gabric, Nikica
    Bohac, Maja
    Romac, Ivana
    Mravicic, Ivana
    Lazic, Ratimir
    COLLEGIUM ANTROPOLOGICUM, 2008, 32 : 27 - 31
  • [6] Visualization of the retinal image in an eye model with spherical and aspheric, diffractive, and refractive multifocal intraocular lenses
    Terwee, Thom
    Weeber, Henk
    van der Mooren, Marrie
    Piers, Patricia
    JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (03) : 223 - 232
  • [7] Visual and Optical Performance With Two Different Diffractive Multifocal Intraocular Lenses Compared to a Monofocal Lens
    Alio, Jorge L.
    Pinero, David P.
    Plaza-Puche, Ana B.
    Amparo, Francisco
    Jimenez, Ramon
    Rodriguez-Prats, Jose L.
    Javaloy, Jaime
    JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (08) : 570 - 581
  • [8] Straylight Measurements in Two Different Apodized Diffractive Multifocal Intraocular Lenses
    Lapid-Gortzak, Ruth
    Labuz, Grzegorz
    van der Meulen, Ivanka J.
    van der Linden, Jan Willem
    Mourits, M. P.
    van den Berg, Tom J.
    JOURNAL OF REFRACTIVE SURGERY, 2015, 31 (11) : 746 - 751
  • [9] Comparison of Visual Performance of four types of Diffractive Multifocal Intraocular Lenses
    Doroodgar, Farideh
    Karimian, Farid
    Niazi, Sana
    Sanginabadi, Azad
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [10] Design and Optimization of Diffractive Multifocal Intraocular Lenses
    Dong Bo
    Yang Ying
    Xue Changxi
    ACTA OPTICA SINICA, 2023, 43 (19)