摘要
BackgroundConventional x-ray imaging and fluoroscopy have limitations in quantitation due to several challenges, including scatter, beam hardening, and overlapping tissues. Dual-energy (DE) imaging, with its capability to quantify area density of specific materials, is well-suited to address such limitations, but only if the dual-energy projections are acquired with perfect spatial and temporal alignment and corrected for scatter.PurposeIn this work, we propose single-shot quantitative imaging (SSQI) by combining the use of a primary modulator (PM) and dual-layer (DL) detector, which enables motion-free DE imaging with scatter correction in a single exposure.MethodsThe key components of our SSQI setup include a PM and DL detector, where the former enables scatter correction for the latter while the latter enables beam hardening correction for the former. The SSQI algorithm allows simultaneous recovery of two material-specific images and two scatter images using four sub-measurements from the PM encoding. The concept was first demonstrated using simulation of chest x-ray imaging for a COVID patient. For validation, we set up SSQI geometry on our tabletop system and imaged acrylic and copper slabs with known thicknesses (acrylic: 0-22.5 cm; copper: 0-0.9 mm), estimated scatter with our SSQI algorithm, and compared the material decomposition (MD) for different combinations of the two materials with ground truth. Second, we imaged an anthropomorphic chest phantom containing contrast in the coronary arteries and compared the MD with and without SSQI. Lastly, to evaluate SSQI in dynamic applications, we constructed a flow phantom that enabled dynamic imaging of iodine contrast.ResultsOur simulation study demonstrated that SSQI led to accurate scatter correction and MD, particularly for smaller focal blur and finer PM pitch. In the validation study, we found that the root mean squared error (RMSE) of SSQI estimation was 0.13 cm for acrylic and 0.04 mm for copper. For the anthropomorphic phantom, direct MD resulted in incorrect interpretation of contrast and soft tissue, while SSQI successfully distinguished them quantitatively, reducing RMSE in material-specific images by 38%-92%. For the flow phantom, SSQI was able to perform accurate dynamic quantitative imaging, separating contrast from the background.ConclusionsWe demonstrated the potential of SSQI for robust quantitative x-ray imaging. The integration of SSQI is straightforward with the addition of a PM and upgrade to a DL detector, which may enable its widespread adoption, including in techniques such as radiography and dynamic imaging (i.e., real-time image guidance and cone-beam CT).
引用
收藏
页码:2621 / 2632
页数:12
相关论文
共 50 条
  • [1] Single-Shot Quantitative X-ray Imaging Using a Primary Modulator and Dual-Layer Detector: Simulation and Phantom Studies
    Shi, Linxi
    Bennett, N. Robert
    Wang, Adam S.
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [2] Experimental Validation of Single-Shot Quantitative X-Ray Imaging Using Primary Modulation and a Dual-Layer Detector
    Shi, L.
    Bennett, N.
    Wang, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E609 - E609
  • [3] Single-Shot Quantitative X-Ray Imaging Using Primary Modulation and Dual-Layer Detector: A Simulation Study
    Shi, L.
    Bennett, N.
    Wang, A.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Linear modeling of single-shot dual-energy x-ray imaging using a sandwich detector
    Kim, J.
    Kim, D. W.
    Kim, S. H.
    Yun, S.
    Youn, H.
    Jeon, H.
    Kim, H. K.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [5] Single-shot X-ray scattering imaging
    Wali, Faiz
    Kun, Gao
    Zhao, Wu
    Lei, Yaohu
    Huang, Jianheng
    Li, Ji
    Liu, Jinyuan
    Zhu, Peiping
    IMAGING SCIENCE JOURNAL, 2020, 68 (04): : 195 - 200
  • [6] Single-shot dual-energy x-ray imaging with a flat-panel sandwich detector for preclinical imaging
    Han, Jong Chul
    Kim, Ho Kyung
    Kim, Dong Woon
    Yun, Seungman
    Youn, Hanbean
    Kam, Soohwa
    Tanguay, Jesse
    Cunningham, Ian A.
    CURRENT APPLIED PHYSICS, 2014, 14 (12) : 1734 - 1742
  • [7] X-ray imaging - Single-shot X-ray diffraction aims at imaging macromolecules
    Brinkmann, Uwe
    LASER FOCUS WORLD, 2007, 43 (01): : 26 - +
  • [8] Single-Shot Quantitative X-ray Imaging from Simultaneous Scatter and Dual Energy Measurements: A Simulation Study
    Wang, Adam S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [9] Quantitative imaging of single-shot liquid distributions in sprays using broadband flash x-ray radiography
    Halls, B. R.
    Roy, S.
    Gord, J. R.
    Kastengren, A. L.
    Meyer, T. R.
    INTERNATIONAL JOURNAL OF MULTIPHASE FLOW, 2016, 87 : 241 - 249
  • [10] Effects of X-Ray Scatter in Quantitative Dual-Energy Imaging Using Dual-Layer Flat Panel Detectors
    Zhao, Chumin
    Liu, Stephen Z.
    Wang, Wenying
    Herbst, Magdalena
    Weber, Thomas
    Vogt, Sebastian
    Ritschl, Ludwig
    Kappler, Steffen
    Stayman, J. Webster
    Siewerdsen, Jeffrey H.
    Zbijewski, Wojciech
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595