Rapid deployment versus its conventional counterpart in aortic valve replacement: comparison of early hemodynamic outcomes

被引:1
作者
Sohn, Suk Ho [1 ]
Kim, Kyung Hwan [1 ]
Kang, Yoonjin [1 ]
Kim, Ji Seong [1 ]
Choi, Jae Woong [1 ]
Hwang, Ho Young [1 ]
机构
[1] Seoul Natl Univ, Seoul Natl Univ Hosp, Dept Thorac & Cardiovasc Surg, Coll Med, 101 Daehak Ro, Seoul 03080, South Korea
关键词
Aortic valve replacement (AVR); bioprosthetic valve; rapid deployment valve; hemodynamics; PROSTHESIS-PATIENT MISMATCH; PROSPECTIVE MULTICENTER; SUTURELESS VALVES; PERFORMANCE; ROOTS; TRIAL;
D O I
10.21037/jtd-23-318
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Edwards Intuity is designed for rapid deployment based on the structure of Magna Ease. This study was conducted to compare early hemodynamic performance between the two valves. Methods: Patients who underwent aortic valve replacement (AVR) using Edwards Intuity or CarpentierEdwards PERIMOUNT Magna Ease in our institution from June 2016 to July 2021 were enrolled. Intuity valve was used in 215 patients, and Magna Ease valve was used in 198 patients, respectively. Early postoperative echocardiographic data were available in 99.0% (409/413) of the patients. The transvalvular mean pressure gradient, effective orifice area, and effective orifice area index were compared between the valves stratified by prosthesis size. Results: There were no differences in the proportion of female patients or body surface area between the groups. Mean pressure gradient on early postoperative echocardiography was significantly lower in Intuity valve than Magna Ease valve for 19, 21, 23, and 25 mm valves (15.5 & PLUSMN;5.0 vs. 20.8 & PLUSMN;9.1 mmHg, P=0.004; 12.7 & PLUSMN;4.2 vs. 15.6 & PLUSMN;5.3 mmHg, P=0.001; 11.5 & PLUSMN;3.3 vs. 13.4 & PLUSMN;5.8 mmHg, P=0.034; and 9.9 & PLUSMN;3.1 vs. 12.3 & PLUSMN;4.0 mmHg, P=0.029; respectively). Effective orifice area was larger in Intuity valve than Magna Ease valve for 19 mm valve (1.45 & PLUSMN;0.38 vs. 1.19 & PLUSMN;0.28 cm2, P=0.002), and effective orifice area index was also larger in Intuity valve than Magna Ease valve for 19 mm valve (0.96 & PLUSMN;0.26 vs. 0.80 & PLUSMN;0.20 cm2/m2, P=0.005). Early clinical outcomes, including operative mortality and postoperative complications, demonstrated no significant differences between the groups. Conclusions: Edwards Intuity demonstrated superior early hemodynamic performance compared with Magna Ease in a size-by-size comparison, and this superiority was more definite for small prostheses.
引用
收藏
页码:3673 / +
页数:13
相关论文
共 35 条
[1]  
Ai LS, 2017, INNOVATIONS, V12, P338, DOI 10.1097/IMI.0000000000000407
[2]   Intermediate-term outcome of 500 consecutive rapid-deployment surgical aortic valve procedures [J].
Andreas, Martin ;
Coti, Iuliana ;
Rosenhek, Raphael ;
Shabanian, Shiva ;
Mahr, Stephane ;
Uyanik-Uenal, Keziban ;
Wiedemann, Dominik ;
Binder, Thomas ;
Kocher, Alfred ;
Laufer, Guenther .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2019, 55 (03) :527-533
[3]   Conventional versus rapid-deployment aortic valve replacement: a single-centre comparison between the Edwards Magna valve and its rapid-deployment successor [J].
Andreas, Martin ;
Wallner, Stephanie ;
Habertheuer, Andreas ;
Rath, Claus ;
Schauperl, Martin ;
Binder, Thomas ;
Beitzke, Dietrich ;
Rosenhek, Raphael ;
Loewe, Christian ;
Wiedemann, Dominik ;
Kocher, Alfred ;
Laufer, Guenther .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2016, 22 (06) :799-805
[4]   TRANSFORM (Multicenter Experience With Rapid Deployment Edwards INTUITY Valve System for Aortic Valve Replacement) US clinical trial: Performance of a rapid deployment aortic valve [J].
Barnhart, Glenn R. ;
Accola, Kevin D. ;
Grossi, Eugene A. ;
Woo, Y. Joseph ;
Mumtaz, Mubashir A. ;
Sabik, Joseph F. ;
Slachman, Frank N. ;
Patel, Himanshu J. ;
Borger, Michael A. ;
Garrett, H. Edward, Jr. ;
Rodriguez, Evelio ;
McCarthy, Patrick M. ;
Ryan, William H. ;
Duhay, Francis G. ;
Mack, Michael J. ;
Chitwood, W. Randolph, Jr. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2017, 153 (02) :241-+
[5]   Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method? [J].
Bleiziffer, Sabine ;
Eichinger, Walter B. ;
Hettich, Ina ;
Guenzinger, Ralf ;
Ruzicka, Daniel ;
Bauernschmitt, Robert ;
Lange, Ruediger .
HEART, 2007, 93 (05) :615-620
[6]   Haemodynamic benefits of rapid deployment aortic valve replacement via a minimally invasive approach: 1-year results of a prospective multicentre randomized controlled trialaEuro [J].
Borger, Michael A. ;
Dohmen, Pascal M. ;
Knosalla, Christoph ;
Hammerschmidt, Robert ;
Merk, Denis R. ;
Richter, Markus ;
Doenst, Torsten ;
Conradi, Lenard ;
Treede, Hendrik ;
Moustafine, Vadim ;
Holzhey, David M. ;
Duhay, Francis ;
Strauch, Justus .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2016, 50 (04) :713-720
[7]   A Randomized Multicenter Trial of Minimally Invasive Rapid Deployment Versus Conventional Full Sternotomy Aortic Valve Replacement [J].
Borger, Michael A. ;
Moustafine, Vadim ;
Conradi, Lenard ;
Knosalla, Christoph ;
Richter, Markus ;
Merk, Denis R. ;
Doenst, Torsten ;
Hammerschmidt, Robert ;
Treede, Hendrik ;
Dohmen, Pascal ;
Strauch, Justus T. .
ANNALS OF THORACIC SURGERY, 2015, 99 (01) :17-24
[8]   Very Long-Term Outcomes of the Carpentier-Edwards Perimount Valve in Aortic Position [J].
Bourguignon, Thierry ;
Bouquiaux-Stablo, Anne-Lorraine ;
Candolfi, Pascal ;
Mirza, Alain ;
Loardi, Claudia ;
May, Marc-Antoine ;
El-Khoury, Rym ;
Marchand, Michel ;
Aupart, Michel .
ANNALS OF THORACIC SURGERY, 2015, 99 (03) :831-837
[9]   Early failures of Trifecta aortic bioprosthesis [J].
Campisi, Salvatore ;
Camilleri, Lionel ;
Innorta, Andrea ;
Azarnoush, Kasra .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 148 (01) :E133-E134
[10]   Pledget-Armed Sutures Affect the Haemodynamic Performance of Biologic Aortic Valve Substitutes: A Preliminary Experimental and Computational Study [J].
Capelli, Claudio ;
Corsini, Chiara ;
Biscarini, Dario ;
Ruffini, Francesco ;
Migliavacca, Francesco ;
Kocher, Alfred ;
Laufer, Guenther ;
Taylor, Andrew M. ;
Schievano, Silvia ;
Andreas, Martin ;
Burriesci, Gaetano ;
Rath, Claus .
CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2017, 8 (01) :17-29