Towards universal comparability of pericoronary adipose tissue attenuation: a coronary computed tomography angiography phantom study

被引:13
作者
Etter, Dominik [1 ,2 ]
Warnock, Geoff [1 ,2 ]
Koszarski, Frederic [1 ]
Niemann, Tilo [3 ]
Mikail, Nidaa [1 ,2 ]
Bengs, Susan [1 ,2 ]
Buechel, Ronny. R. R. [1 ]
Kaufmann, Philipp [1 ]
Gebhard, Catherine [1 ,2 ]
Rossi, Alexia [1 ,2 ]
机构
[1] Univ Hosp Zurich, Dept Nucl Med, Raemistr 100, CH-8091 Zurich, Switzerland
[2] Univ Zurich, Ctr Mol Cardiol, CH-8952 Schlieren, Switzerland
[3] Kantonsspital Baden, Dept Radiol, CH-5400 Baden, Switzerland
关键词
Adipose tissue; Computed tomography angiography; Coronary arteries; Image reconstruction; INFLAMMATION;
D O I
10.1007/s00330-022-09274-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesDifferent computed tomography (CT) scanners, variations in acquisition protocols, and technical parameters employed for image reconstruction may introduce bias in the analysis of pericoronary adipose tissue (PCAT) attenuation derived from coronary computed tomography angiography (CCTA). Therefore, the aim of this study was to establish the effect of tube voltage, measured as kilovoltage peak (kVp), and iterative reconstruction on PCAT mean attenuation (PCAT(MA)). MethodsTwelve healthy ex vivo porcine hearts were injected with iodine-enriched agar-agar to allow for ex vivo CCTA imaging on a 256-slice CT and a dual-source CT system. Images were acquired at tube voltages of 80, 100, 120, and 140 kVp and reconstructed by using both filtered back projection and iterative reconstruction algorithms. PCAT(MA) was measured semi-automatically on CCTA images in the proximal segment of coronary arteries. ResultsThe tube voltage showed a significant effect on PCAT(MA) measurements on both the 256-slice CT scanner (p < 0.001) and the dual-source CT system (p = 0.013), resulting in higher attenuation values with increasing tube voltage. Similarly, the use of iterative reconstructions was associated with a significant increase of PCAT(MA) (256-slice CT: p < 0.001 and dual-source CT: p = 0.014). Averaged conversion factors to correct PCAT(MA) measurements for tube voltage other than 120 kVp were 1.267, 1.080 and 0.947 for 80, 100, and 140 kVp, respectively. ConclusionPCAT(MA) values are significantly affected by acquisition and reconstruction parameters. The same tube voltage and reconstruction type are recommended when PCAT attenuation is used in multicenter and longitudinal studies.
引用
收藏
页码:2324 / 2330
页数:7
相关论文
共 24 条
[1]   Influence of slice thickness and reconstruction kernel on the computed tomographic attenuation of coronary atherosclerotic plaque [J].
Achenbach, Stephan ;
Boehmer, Kerstin ;
Pflederer, Tobias ;
Ropers, Dieter ;
Seltmann, Martin ;
Lell, Michael ;
Anders, Katharina ;
Kuettner, Axel ;
Uder, Michael ;
Daniel, Werner G. ;
Marwan, Mohamed .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2010, 4 (02) :110-115
[2]   Detecting human coronary inflammation by imaging perivascular fat [J].
Antonopoulos, Alexios S. ;
Sanna, Fabio ;
Sabharwal, Nikant ;
Thomas, Sheena ;
Oikonomou, Evangelos K. ;
Herdman, Laura ;
Margaritis, Marios ;
Shirodaria, Cheerag ;
Kampoli, Anna-Maria ;
Akoumianakis, Ioannis ;
Petrou, Mario ;
Sayeed, Rana ;
Krasopoulos, George ;
Psarros, Constantinos ;
Ciccone, Patricia ;
Brophy, Carl M. ;
Digby, Janet ;
Kelion, Andrew ;
Uberoi, Raman ;
Anthony, Suzan ;
Alexopoulos, Nikolaos ;
Tousoulis, Dimitris ;
Achenbach, Stephan ;
Neubauer, Stefan ;
Channon, Keith M. ;
Antoniades, Charalambos .
SCIENCE TRANSLATIONAL MEDICINE, 2017, 9 (398)
[3]   Quantification of perivascular inflammation does not provide incremental prognostic value over myocardial perfusion imaging and calcium scoring [J].
Bengs, Susan ;
Haider, Ahmed ;
Warnock, Geoffrey I. ;
Fiechter, Michael ;
Pargaetzi, Yves ;
Rampidis, Georgios ;
Etter, Dominik ;
Wijnen, Winandus J. ;
Portmann, Angela ;
Osto, Elena ;
Treyer, Valerie ;
Benz, Dominik C. ;
Meisel, Alexander ;
Fuchs, Tobias A. ;
Graeni, Christoph ;
Buechel, Ronny R. ;
Kaufmann, Philipp A. ;
Pazhenkottil, Aju P. ;
Gebhard, Catherine .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (06) :1806-1812
[4]   Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography [J].
Cademartiri, Filippo ;
La Grutta, Ludovico ;
Runza, Giuseppe ;
Palumbo, Alessandro ;
Maffei, Erica ;
Mollet, Nico R. A. ;
Bartolotta, Tommaso Vincenzo ;
Somers, Pamela ;
Knaapen, Michiel ;
Verheye, Stefan ;
Midiri, Massimo ;
Hamers, Ronald ;
Bruining, Nico .
EUROPEAN RADIOLOGY, 2007, 17 (07) :1842-1849
[5]   Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis [J].
Den Harder, Annemarie M. ;
Willemink, Martin J. ;
De Ruiter, Quirina M. B. ;
De Jong, Pim A. ;
Schilham, Arnold M. R. ;
Krestin, Gabriel P. ;
Leiner, Tim ;
Budde, Ricardo P. J. .
BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1058)
[6]  
Eun Sul Lee RNF., 2006, ANAL COMPLEX SURVEY
[7]   CT-based analysis of pericoronary adipose tissue density: Relation to cardiovascular risk factors and epicardial adipose tissue volume [J].
Hell, Michaela M. ;
Achenbach, Stephan ;
Schuhbaeck, Annika ;
Klinghammer, Lutz ;
May, Matthias S. ;
Marwan, Mohamed .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2016, 10 (01) :52-60
[8]   Matplotlib: A 2D graphics environment [J].
Hunter, John D. .
COMPUTING IN SCIENCE & ENGINEERING, 2007, 9 (03) :90-95
[9]   Inflammation and its Resolution as Determinants of Acute Coronary Syndromes [J].
Libby, Peter ;
Tabas, Ira ;
Fredman, Gabrielle ;
Fisher, Edward A. .
CIRCULATION RESEARCH, 2014, 114 (12) :1867-1879
[10]   Pericoronary adipose tissue computed tomography attenuation distinguishes different stages of coronary artery disease: a cross-sectional study [J].
Lin, Andrew ;
Nerlekar, Nitesh ;
Yuvaraj, Jeremy ;
Fernandes, Katrina ;
Jiang, Cathy ;
Nicholls, Stephen J. ;
Dey, Damini ;
Wong, Dennis T. L. .
EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2021, 22 (03) :298-306