Comparison of electrical impedance tomography and spirometry-based measures of airflow in healthy adult horses

被引:1
作者
Byrne, David P. [1 ]
Keeshan, Ben [2 ]
Hosgood, Giselle [1 ]
Adler, Andy [2 ]
Mosing, Martina [3 ]
机构
[1] Murdoch Univ, Sch Vet Med, Perth, WA, Australia
[2] Carleton Univ, Dept Syst & Comp Engn, Ottawa, ON, Canada
[3] Univ Vet Med, Dept Compan Anim & Horses, Anaesthesiol & Perioperat Intens Care, Vienna, Austria
关键词
asthma; equine; expiratory; inspiratory; peak flow; plethysmography; REGIONAL LUNG-FUNCTION; RESPIRATORY-FUNCTION; ANESTHETIZED HORSES; VENTILATION; OBSTRUCTION; ANESTHESIA; FREQUENCY; PRESSURE; HEAVES; MASK;
D O I
10.3389/fphys.2023.1164646
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Electrical impedance tomography (EIT) is a non-invasive diagnostic tool for evaluating lung function. The objective of this study was to compare respiratory flow variables calculated from thoracic EIT measurements with corresponding spirometry variables. Ten healthy research horses were sedated and instrumented with spirometry via facemask and a single-plane EIT electrode belt around the thorax. Horses were exposed to sequentially increasing volumes of apparatus dead space between 1,000 and 8,500 mL, in 5-7 steps, to induce carbon dioxide rebreathing, until clinical hyperpnea or a tidal volume of 150% baseline was reached. A 2-min stabilization period followed by 2 minutes of data collection occurred at each timepoint. Peak inspiratory and expiratory flow, inspiratory and expiratory time, and expiratory nadir flow, defined as the lowest expiratory flow between the deceleration of flow of the first passive phase of expiration and the acceleration of flow of the second active phase of expiration were evaluated with EIT and spirometry. Breathing pattern was assessed based on the total impedance curve. Bland-Altman analysis was used to evaluate the agreement where perfect agreement was indicated by a ratio of EIT:spirometry of 1.0. The mean ratio (bias; expressed as a percentage difference from perfect agreement) and the 95% confidence interval of the bias are reported. There was good agreement between EIT-derived and spirometry-derived peak inspiratory [-15% (-46-32)] and expiratory [10% (-32-20)] flows and inspiratory [-6% (-25-18)] and expiratory [5% (-9-20)] times. Agreement for nadir flows was poor [-22% (-87-369)]. Sedated horses intermittently exhibited Cheyne-Stokes variant respiration, and a breath pattern with incomplete expiration in between breaths (crown-like breaths). Electrical impedance tomography can quantify airflow changes over increasing tidal volumes and changing breathing pattern when compared with spirometry in standing sedated horses.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Deep learning based reconstruction enables high-resolution electrical impedance tomography for lung function assessment
    Zeng, Shihao
    Kwole, Wang Chun
    Cao, Peng
    Zouari, Fedi
    Lee, Philip Tin Yun
    Chan, Russell W.
    Touboul, Adrien
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [42] Electrical Impedance Tomography Predicts Weaning Success in Adult Patients With Delayed Upper Abdominal Surgery: A Single-Center Retrospective Study
    Li, Jiajia
    Zeng, Fan
    Yang, Fuxun
    Luo, Xiaoxiu
    Liu, Rongan
    Ren, Yinjie
    Lan, Yunping
    Lei, Yu
    Zhao, Gaoping
    Huang, Xiaobo
    FRONTIERS IN MEDICINE, 2021, 8
  • [43] Estimation of Alveolar Recruitment Potential Using Electrical Impedance Tomography Based on an Exponential Model of the Pressure-Volume Curve
    Turco, G. E.
    Moura, F. S.
    Camargo, E. D. L. B.
    XXVII BRAZILIAN CONGRESS ON BIOMEDICAL ENGINEERING, CBEB 2020, 2022, : 1949 - 1952
  • [44] A new current injection and voltage measurement strategy of 3D electrical impedance tomography based on scanning electrode
    Liu, Jinzhen
    Zhu, Zhaoqi
    Xiong, Hui
    Li, Chunchan
    Chen, Yunjun
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2022, 93 (09)
  • [45] Experimental Validations of In Vivo Human Musculoskeletal Tissue Conductivity Images Using MR-Based Electrical Impedance Tomography
    Jeong, Woo Chul
    Meng, Zi Jun
    Kim, Hyung Joong
    Kwon, Oh In
    Woo, Eung Je
    BIOELECTROMAGNETICS, 2014, 35 (05) : 363 - 372
  • [46] Electrical-impedance-tomography imaging based on a new three-dimensional thorax model for assessing the extent of lung injury
    Li, Xiuyan
    Chen, Xiaojing
    Wang, Qi
    Wang, Jianming
    Duan, Xiaojie
    Sun, Yukuan
    Wang, Huaxiang
    AIP ADVANCES, 2019, 9 (12)
  • [47] Electrical impedance tomography guided positive end-expiratory pressure titration in critically ill and surgical adult patients: a systematic review and meta-analysis
    Gao, Yelin
    He, Huaiwu
    Chi, Yi
    Frerichs, Inez
    Long, Yun
    Zhao, Zhanqi
    BMC PULMONARY MEDICINE, 2024, 24 (01):
  • [48] Soft Nanocomposite Based Multi-point, Multi-directional Strain Mapping Sensor Using Anisotropic Electrical Impedance Tomography
    Lee, Hyosang
    Kwon, Donguk
    Cho, Haedo
    Park, Inkyu
    Kim, Jung
    SCIENTIFIC REPORTS, 2017, 7
  • [49] Bedside personalized methods based on electrical impedance tomography or respiratory mechanics to set PEEP in ARDS and recruitment-to-inflation ratio: a physiologic study
    Pavlovsky, Bertrand
    Desprez, Christophe
    Richard, Jean-Christophe
    Fage, Nicolas
    Lesimple, Arnaud
    Chean, Dara
    Courtais, Antonin
    Mauri, Tommaso
    Mercat, Alain
    Beloncle, Francois
    ANNALS OF INTENSIVE CARE, 2024, 14 (01)
  • [50] Bedside Selection of Positive End Expiratory Pressure by Electrical Impedance Tomography in Patients Undergoing Veno-Venous Extracorporeal Membrane Oxygenation Support: A Comparison between COVID-19 ARDS and ARDS from Other Etiologies
    Di Pierro, Michela
    Giani, Marco
    Bronco, Alfio
    Lembo, Francesca Maria
    Rona, Roberto
    Bellani, Giacomo
    Foti, Giuseppe
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (06)