Identification of screw spacing on pediatric hip locking plate in proximal femoral osteotomy

被引:0
|
作者
Jeong, Byeong Cheol [1 ]
Goh, Tae Sik [2 ,3 ]
Lee, Chiseung [2 ,4 ]
Ahn, Tae Young [2 ,3 ]
Ryu, Dongman [5 ]
机构
[1] Pusan Natl Univ, Grad Sch, Dept Biomed Engn, Busan 49241, South Korea
[2] Pusan Natl Univ Hosp, Biomed Res Inst, Busan 49241, South Korea
[3] Pusan Natl Univ, Sch Med, Dept Orthoped Surg, Busan 49241, South Korea
[4] Pusan Natl Univ, Sch Med, Dept Biomed Engn, Busan 49241, South Korea
[5] Pusan Natl Univ, Med Res Inst, Busan 49241, South Korea
关键词
Developmental dysplasia of the hip; Proximal femoral osteotomy; Pediatric hip locking plate; Screw spacing; Screw angle; Group pile; FIXATION;
D O I
10.1007/s13246-023-01277-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study describes a computational analysis technique for evaluating the effect of screw spacing and angle on the pediatric hip locking plate system in proximal femoral osteotomy in pediatric patients having DDH with an aberrant femoral head and femoral angle. Under static compressive load conditions, the stresses of the screw and bone were examined as the screw spacing and angle changed. The spacing and angle of various screws were specifically considered as variables in this study based on the pile mechanism studied in civil engineering. As with the group pile mechanism, the tighter the screw spacing under static compressive loads, the more the overlapping effect between the bone stresses and the screws develops, increasing the risk of injuring the patient's bone. Therefore, a series of simulations was performed to determine the optimal screw spacing and angles to minimize the overlapping effect of bone stress. In addition, a formula for determining the minimum screw spacing was proposed based on the computational simulation results. Finally, if the outcomes of this study are applied to pediatric patients with DDH in the pre-proximal femoral osteotomy stage, post-operative load-induced femur damage will be reduced.
引用
收藏
页码:1101 / 1114
页数:14
相关论文
共 50 条
  • [1] Identification of screw spacing on pediatric hip locking plate in proximal femoral osteotomy
    Byeong Cheol Jeong
    Tae Sik Goh
    Chiseung Lee
    Tae Young Ahn
    Dongman Ryu
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1101 - 1114
  • [2] Proximal femoral locking plate versus dynamic hip screw for unstable intertrochanteric femoral fractures
    Upadhyay, Sachin
    Raza, H. K. T.
    JOURNAL OF ORTHOPAEDIC SURGERY, 2014, 22 (01) : 130 - 131
  • [3] Proximal femoral locking plate versus dynamic hip screw for unstable intertrochanteric femoral fractures
    Dhamangaonkar, Anoop C.
    Joshi, Deepak
    Goregaonkar, Arvind B.
    Tawari, Akhil A.
    JOURNAL OF ORTHOPAEDIC SURGERY, 2013, 21 (03) : 317 - 322
  • [4] Safety of the LCP Pediatric Hip Plate in Proximal Femoral Osteotomy in Children with Cerebral Palsy
    Sannarah, Omar Q.
    Shaheen, Majd A.
    Tehabsinn, Rana A.
    Shaheen, Bayan A.
    Makahleh, Marah B.
    Almustafa, Mahmoud M.
    Al Hadidi, Fadi A.
    Hussein, Lutfi A.
    Hammad, Yazan S.
    JOURNAL OF MULTIDISCIPLINARY HEALTHCARE, 2020, 13 : 779 - 784
  • [5] Proximal femoral locking compression plate for proximal femoral fractures
    Lee, Wei Ting
    Murphy, Diarmuid
    Kagda, Fareed H. Y.
    Thambiah, Joseph
    JOURNAL OF ORTHOPAEDIC SURGERY, 2014, 22 (03) : 287 - 293
  • [6] Which implant for proximal femoral osteotomy in children?: A comparison of the AO (ASIF) 90° fixed-angle blade plate and the Richards intermediate hip screw
    Hau, R
    Dickens, DRV
    Nattrass, GR
    O'Sullivan, M
    Torode, IP
    Graham, HK
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2000, 20 (03) : 336 - 343
  • [7] Scattering and clustering the proximal screw construct in unilateral locking plate osteosynthesis of distal femoral fractures
    Jang, Jae Hoon
    Rhee, Seung Joon
    Jun, Se Bin
    Choi, Yoon Young
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (09) : 2193 - 2203
  • [8] Adult Proximal Humerus Locking Plate for the Treatment of a Pediatric Subtrochanteric Femoral Nonunion: A Case Report
    Cortes, Luis E.
    Triana, Miguel
    Vallejo, Francisco
    Slongo, Theddy F.
    Streubel, Philipp N.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (07) : E63 - E67
  • [9] Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs
    Jain, Amit
    Thompson, John M.
    Brooks, Jaysson T.
    Ain, Michael C.
    Sponseller, Paul D.
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2016, 36 (01) : E1 - E5
  • [10] Planning a Proximal Femoral Varus Osteotomy in Pediatric Orthopedics
    Mifsud, Maximillian
    Buckingham, Rachel
    Theologis, Tim
    TECHNIQUES IN ORTHOPAEDICS, 2021, 36 (02) : 157 - 161