Validation of a drift-corrected 3D MR temperature imaging sequence for breast MR-guided focused ultrasound treatments

被引:6
作者
Adams-Tew, Samuel I. [1 ,2 ,3 ]
Johnson, Sara [1 ]
Od, Henrik
Parker, Dennis L. [1 ]
Payne, Allison [1 ]
机构
[1] Univ Utah, Dept Radiol & Imaging Sci, Salt Lake City, UT USA
[2] Univ Utah, Dept Biomed Engn, Salt Lake City, UT USA
[3] 729 Arapeen Dr, Salt Lake City, UT 84108 USA
关键词
Magnetic resonance guided focused ultrasound; Magnetic resonance thermometry; Breast cancer; Drift correction; MAGNETIC-RESONANCE THERMOMETRY; FREQUENCY-SHIFT; TEMPORAL RESOLUTION; CANCER; HYPERTHERMIA; RECONSTRUCTION; RESPIRATION; ARTIFACT; ABLATION; SURGERY;
D O I
10.1016/j.mri.2022.12.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Real-time temperature monitoring is critical to the success of thermally ablative therapies. This work validates a 3D thermometry sequence with k-space field drift correction designed for use in magnetic resonance-guided focused ultrasound treatments for breast cancer. Fiberoptic probes were embedded in tissue-mimicking phantoms, and temperature change measurements from the probes were compared with the magnetic resonance temperature imaging measurements following heating with focused ultrasound. Precision and accuracy of measurements were also evaluated in free-breathing healthy volunteers (N = 3) under a non-heating condition. MR temperature measurements agreed closely with those of fiberoptic probes, with a 95% confidence interval of measurement difference from -2.0 degrees C to 1.4 degrees C. Field drift-corrected measurements in vivo had a precision of 1.1 +/- 0.7 degrees C and were accurate within 1.3 +/- 0.9 degrees C across the three volunteers. The field drift correction method improved precision and accuracy by an average of 46 and 42%, respectively, when compared to the uncorrected data. This temperature imaging sequence can provide accurate measurements of temperature change in aqueous tissues in the breast and support the use of this sequence in clinical investigations of focused ultrasound treatments for breast cancer.
引用
收藏
页码:126 / 134
页数:9
相关论文
共 51 条
[1]  
American Cancer Society. American Cancer Society, 2021, CANC FACTS FIGURES 2, P2021
[2]   Recurrence rates after treatment of breast cancer with standard radiotherapy with or without additional radiation. [J].
Bartelink, H ;
Horiot, J ;
Poortmans, P ;
Struikmans, H ;
Van den Bogaert, W ;
Barillot, I ;
Fourquet, A ;
Borger, J ;
Jager, J ;
Hoogenraad, W ;
Collette, L ;
Pierart, M .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 345 (19) :1378-1387
[3]   Temperature imaging of laser-induced thermotherapy (LITT) by MRI: evaluation of different sequences in phantom [J].
Bazrafshan, Babak ;
Huebner, Frank ;
Farshid, Parviz ;
Hammerstingl, Renate ;
Paul, Jijo ;
Vogel, Vitali ;
Maentele, Werner ;
Vogl, Thomas J. .
LASERS IN MEDICAL SCIENCE, 2014, 29 (01) :173-183
[4]   Proton Resonance Frequency Shift Thermometry: A Review of Modern Clinical Practices [J].
Blackwell, James ;
Krasny, Marcin J. ;
O'Brien, Aoife ;
Ashkan, Keyoumars ;
Galligan, Josette ;
Destrade, Michel ;
Colgan, Niall .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 55 (02) :389-403
[5]  
Bradley William G Jr, 2009, J Am Coll Radiol, V6, P510, DOI 10.1016/j.jacr.2009.01.004
[6]   Magnetic resonance thermometry during hyperthermia for human high-grade sarcoma [J].
Carter, DL ;
MacFall, JR ;
Clegg, ST ;
Wan, X ;
Prescott, DM ;
Charles, HC ;
Samulski, TV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04) :815-822
[7]   Prostate cancer: MR imaging and thermometry during microwave thermal ablation-initial experience [J].
Chen, JC ;
Moriarty, JA ;
Derbyshire, JA ;
Peters, RD ;
Trachtenberg, J ;
Bell, SD ;
Doyle, J ;
Arrelano, R ;
Wright, GA ;
Henkelman, RM ;
Hinks, RS ;
Lok, SY ;
Toi, A ;
Kucharczyk, W .
RADIOLOGY, 2000, 214 (01) :290-297
[8]   Accuracy of real time noninvasive temperature measurements using magnetic resonance thermal imaging in patients treated for high grade extremity soft tissue sarcomas [J].
Craciunescua, Oana I. ;
Stauffer, Paul R. ;
Soher, Brian J. ;
Wyatt, Cory R. ;
Arabe, Omar ;
Maccarini, Paolo ;
Das, Shiva K. ;
Cheng, Kung-Shan ;
Wong, Terence Z. ;
Jones, Ellen L. ;
Dewhirst, Mark W. ;
Vujaskovic, Zeljko ;
MacFall, James R. .
MEDICAL PHYSICS, 2009, 36 (11) :4848-4858
[9]   Fast PRF-based MR thermometry using double-echo EPI: in vivo comparison in a clinical hyperthermia setting [J].
Dadakova, Tetiana ;
Gellermann, Johanna ;
Voigt, Otilia ;
Korvink, Jan Gerrit ;
Pavlina, John Matthew ;
Hennig, Juergen ;
Bock, Michael .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2015, 28 (04) :305-314
[10]   Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients [J].
Deckers, R. ;
Merckel, L. G. ;
de Senneville, B. Denis ;
Schubert, G. ;
Koehler, M. ;
Knuttel, F. M. ;
Mali, W. P. Th M. ;
Moonen, C. T. W. ;
van den Bosch, M. A. A. J. ;
Bartels, L. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (14) :5527-5542