High-sensitivity atomic magnetometer realized by weak-value-amplification effect

被引:5
|
作者
Lin, Shudong [1 ]
Tang, Junjian [1 ,2 ]
Yuan, Ziqi [1 ]
Huang, Binyue [1 ]
Wang, Yuhao [1 ]
Zhai, Yueyang [1 ,3 ]
机构
[1] Beihang Univ, Sch Instrumentat & Optoelect Engn, Key Lab Ultraweak Magnet Field Measurement Technol, Minist Educ, Beijing 100191, Peoples R China
[2] Beihang Univ, Sch Phys, Beijing 100191, Peoples R China
[3] Hefei Natl Lab, Hefei 230088, Peoples R China
基金
中国国家自然科学基金; 中国博士后科学基金;
关键词
MAGNETOENCEPHALOGRAPHY; SPIN;
D O I
10.1063/5.0168990
中图分类号
O59 [应用物理学];
学科分类号
摘要
The weak-value-amplification (WVA) effect, which is also called weak measurement, has been developed extensively in various sensing systems. Here, we report the actual realization of the WVA effect in spin-exchange relaxation-free atomic magnetometer system, wherein the slight separation of transverse momentum of the probe light is amplified by introducing orthogonal pre- and post-selection states. A differential detector is used to obtain the transverse position of the probe light accurately in real time. The weak coupling of the magneto-optical interaction with atoms will be reflected in the differential signal. The WVA effect is observed and demonstrated directly and a high sensitivity of 8 fT / root H z is achieved. Also, we obtain the stable and distinct simulated magnetocardiography signal through our system. The present successful implementation of this probe method paves the way for further technical noise suppression and sensitivity improvement of quantum sensors.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] A dual-axis, high-sensitivity atomic magnetometer
    李茹杰
    全伟
    范文峰
    邢力
    王卓
    翟跃阳
    房建成
    Chinese Physics B, 2017, (12) : 200 - 204
  • [2] A dual-axis, high-sensitivity atomic magnetometer
    Li, Rujie
    Quan, Wei
    Fan, Wenfeng
    Xing, Li
    Wang, Zhuo
    Zhai, Yueyang
    Fang, Jiancheng
    CHINESE PHYSICS B, 2017, 26 (12)
  • [3] Continuous High-Sensitivity and High-Bandwidth Atomic Magnetometer
    Li, Rujie
    Baynes, Fred N.
    Luiten, Andre N.
    Perrella, Christopher
    PHYSICAL REVIEW APPLIED, 2020, 14 (06)
  • [4] Weak-value-amplification metrology without spectral analysis
    Huang, Jing-Zheng
    Fang, Chen
    Zeng, Guihua
    PHYSICAL REVIEW A, 2018, 97 (06)
  • [5] Analysis for weak-value-amplification measurement in trapped ion systems
    Qin, Lupei
    Liu, Yingxin
    PHYSICS LETTERS A, 2022, 448
  • [6] Human MCG measurements with a high-sensitivity potassium atomic magnetometer
    Kamada, K.
    Ito, Y.
    Kobayashi, T.
    PHYSIOLOGICAL MEASUREMENT, 2012, 33 (06) : 1063 - 1071
  • [7] A high-sensitivity Hall magnetometer
    Rostami, Kh. R.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2016, 59 (02) : 273 - 277
  • [8] A HIGH-SENSITIVITY MAGNETORESISTIVE MAGNETOMETER
    SMITH, N
    JEFFERS, F
    FREEMAN, J
    JOURNAL OF APPLIED PHYSICS, 1991, 69 (08) : 5082 - 5084
  • [9] HIGH-SENSITIVITY CESIUM MAGNETOMETER
    ANDRIANOV, BA
    GRINKO, IE
    LUKOSHIN, AF
    OVCHARENKO, PS
    MEASUREMENT TECHNIQUES, 1976, 19 (10) : 1519 - 1522
  • [10] A HIGH-SENSITIVITY GRADIOMETRIC MAGNETOMETER
    CHAPLYGIN, YA
    VOLKOV, SI
    GRIGORASHVILI, YE
    MATVEEV, IA
    MEASUREMENT TECHNIQUES USSR, 1994, 37 (12): : 1341 - 1344