Telemedicine in the Era of a Pandemic: Usefulness of a Novel Three-Lead ECG

被引:3
作者
Babic, Milos D. D. [1 ]
Veljkovic, Stefan [1 ]
Lakcevic, Jovana [1 ]
Babic, Rade [1 ,2 ]
Ostojic, Miodrag [1 ,2 ]
Petrovic, Masa [1 ,2 ]
Boljevic, Darko [1 ]
Tomic, Stanko [3 ]
Bojic, Milovan [1 ]
Nikolic, Aleksandra [1 ,2 ]
机构
[1] Dedinje Cardiovasc Inst, Belgrade 11000, Serbia
[2] Univ Belgrade, Fac Med, Belgrade 11000, Serbia
[3] Vinca Inst Nucl Sci, Belgrade 11000, Serbia
关键词
electrocardiogram; COVID-19; pandemic; telemedicine; medical devices; signal processing; 12-LEAD ELECTROCARDIOGRAMS; ATRIAL-FIBRILLATION; FEASIBILITY; ACCURACY; REST;
D O I
10.3390/diagnostics13152525
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The 12-lead electrocardiogram (ECG) is a first-line diagnostic tool for patients with cardiac symptoms. As observed during the COVID-19 pandemic, the ECG is essential to the initial patient evaluation. The novel KardioPal three-lead-based ECG reconstructive technology provides a potential alternative to a standard ECG, reducing the response time and cost of treatment and improving patient comfort. Our study aimed to evaluate the diagnostic accuracy of a reconstructed 12-lead ECG obtained by the KardioPal technology, comparing it with the standard 12-lead ECG, and to assess the feasibility and time required to obtain a reconstructed ECG in a real-life scenario. A prospective, nonrandomized, single-center, adjudicator-blinded trial was conducted on 102 patients during the COVID-19 pandemic at the Dedinje Cardiovascular Institute in Belgrade. The KardioPal system demonstrated a high feasibility rate (99%), with high specificity (96.3%), sensitivity (95.8%), and diagnostic accuracy (96.1%) for obtaining clinically relevant matching of reconstructed 12-lead compared to the standard 12-lead ECG recording. This novel technology provided a significant reduction in ECG acquisition time and the need for personnel and space for obtaining ECG recordings, thereby reducing the risk of viral transmission and the burden on an already overwhelmed healthcare system such as the one experienced during the COVID-19 pandemic.
引用
收藏
页数:17
相关论文
共 36 条
[21]   Assessment of derived 12-lead electrocardiograms using general and patient-specific reconstruction strategies at rest and during transient myocardial ischemia [J].
Nelwan, SP ;
Crater, SW ;
Green, CL ;
Johanson, P ;
van Dam, TB ;
Meij, SH ;
Simoons, ML ;
Krucoff, MW .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (12) :1529-1533
[22]   Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion [J].
Nelwan, Stefan P. ;
Kors, Jan A. ;
Crater, Suzanne W. ;
Meij, Simon H. ;
van Dam, Teus B. ;
Simoons, Maarten L. ;
Kmcoff, Mitchell W. .
JOURNAL OF ELECTROCARDIOLOGY, 2008, 41 (03) :230-237
[23]   Diagnostic accuracy of multi-lead ECGs obtained using a pocket-sized bipolar handheld event recorder [J].
Nigolian, Ani ;
Dayal, Nicolas ;
Nigolian, Haig ;
Stettler, Carine ;
Burri, Haran .
JOURNAL OF ELECTROCARDIOLOGY, 2018, 51 (02) :278-281
[24]   The diagnosis of cardiac arrhythmias: A prospective multi-center randomized study comparing mobile cardiac outpatient telemetry versus standard loop event monitoring [J].
Rothman, Steven A. ;
Laughlin, James C. ;
Seltzer, Jonathan ;
Walia, Jasjit S. ;
Baman, Rakesh I. ;
Siouffi, Samer Y. ;
Sangrigoli, Robert M. ;
Kowey, Peter R. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 (03) :241-247
[25]   Derivation of the 12-lead electrocardiogram and 3-lead vectorcardiogram [J].
Schreck, David M. ;
Fishberg, Robert D. .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2013, 31 (08) :1183-1190
[26]   Comparison of EASI-derived 12-lead electrocardiograms versus paramedic-acquired 12-lead electrocardiograms using Mason-Likar limb lead configuration in patients with chest pain [J].
Sejersten, M ;
Pahlm, O ;
Pettersson, J ;
Zhou, S ;
Maynard, C ;
Feldman, CL ;
Wagner, GS .
JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (01) :13-21
[27]   The relative accuracies of ECG precordial lead waveforms derived from EASI leads and those acquired from paramedic applied standard leads [J].
Sejersten, M ;
Pahlm, O ;
Pettersson, J ;
Clemmensen, PM ;
Rautaharju, F ;
Zhou, O ;
Maynard, C ;
Feldman, CL ;
Wagner, GS .
JOURNAL OF ELECTROCARDIOLOGY, 2003, 36 (03) :179-185
[28]   Feasibility of CardioSecur®, a Mobile 4-Electrode/22-Lead ECG Device, in the Prehospital Emergency Setting [J].
Spaich, Sebastian ;
Kern, Hanna ;
Zelniker, Thomas A. ;
Stiepak, Jan ;
Gabel, Michael ;
Popp, Erik ;
Katus, Hugo A. ;
Preusch, Michael R. .
FRONTIERS IN CARDIOVASCULAR MEDICINE, 2020, 7
[29]   2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiacmonitoring/telemetry [J].
Steinberg, Jonathan S. ;
Varma, Niraj ;
Cygankiewicz, Iwona ;
Aziz, Peter ;
Balsam, Pawel ;
Baranchuk, Adrian ;
Cantillon, Daniel J. ;
Dilaveris, Polychronis ;
Dubner, Sergio J. ;
El-Sherif, Nabil ;
Krol, Jaroslaw ;
Kurpesa, Malgorzata ;
La Rovere, Maria Teresa ;
Lobodzinski, Suave S. ;
Locati, Emanuela T. ;
Mittal, Suneet ;
Olshansky, Brian ;
Piotrowicz, Ewa ;
Saxon, Leslie ;
Stone, Peter H. ;
Tereshchenko, Larisa ;
Turitto, Gioia ;
Wimmer, Neil J. ;
Verrier, Richard L. ;
Zareba, Wojciech ;
Piotrowicz, Ryszard .
HEART RHYTHM, 2017, 14 (07) :E55-E96
[30]  
Stroobandt RX., 2016, ECG from Basics to Essentials