Spectral separation of evoked and spontaneous cortical activity, Part 1: Delta to high gamma band

被引:1
作者
Fischer, G. [1 ]
Haueisen, J. [2 ]
Baumgarten, D. [1 ]
Kofler, M. [3 ]
机构
[1] UMIT TIROL Private Univ Hlth Sci & Hlth Technol, Inst Elect & Biomed Engn, Eduard Wallnoefer Zentrum 1, A-6060 Hall In Tirol, Austria
[2] Tech Univ Ilmenau, Inst Biomed Engn & Informat, Ilmenau, Germany
[3] Hochzirl Hosp, Dept Neurol, Zirl, Austria
关键词
Somatosensory evoked potentials; Fourier transform; Short latency SEP; Evoked to background ratio; COHERENT AVERAGING TECHNIQUE; OSCILLATION; RESPONSES; SIGNALS;
D O I
10.1016/j.bspc.2024.106094
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Spectral analysis of repeatedly evoked potentials (EPs) is challenging since recordings contain a superposition of evoked signals and spontaneous activity. We developed a novel approach, N-interval Fourier Transform Analysis (N-FTA), which allows for reliable separation and simultaneous assessment of triggered and background spectral components. Median nerve stimulation data from a total of eleven volunteers recorded in two labs with different experimental settings were investigated. Consistently, short latency spectral components were mainly contained in the gamma and high gamma bands. In contrast, spontaneous activity displayed a 1/f spectral profile with distinct alpha and beta peaks. Spontaneous power spectral densities (PSDs) obtained for real and sham stimulation were highly comparable. The low frequency background PSD was more than two orders of magnitude above the spectral short latency peaks. Within the 30 Hz to 90 Hz band, the evoked peaks were -17 dB to -4 dB below the background suggesting that target band filtered short latency deflection might be extracted using less than 100 trials. SEPs following tibial nerve stimulation (3 subjects) displayed a narrower spectral band at about half the bandwidth as compared to median nerve stimulation. Evoked peaks were between 30 Hz and 37 Hz at PSD levels being -10 dB to -4 dB below the background activity. These spectral peaks were related to the short latency response of typical W-morphology. Cortical short latency responses are contained in distinct spectral target bands which are much narrower than the standard bandwidth recommendations for routine recordings. In particular, the high pass corner frequency may be selected about one order of magnitude above the current standard. This might render SEP recordings more robust since it eases the suppression of spontaneous activity and movement artifacts such as eye-blinks. Real-time zero -phase filters are required for translating these findings into improved recording systems.
引用
收藏
页数:11
相关论文
共 26 条
  • [1] Long-latency gamma modulation after median nerve stimulation delineates the central sulcus and contrasts the states of consciousness
    Asman, Priscella
    Pellizzer, Giuseppe
    Tummala, Sudhakar
    Tasnim, Israt
    Bastos, Dhiego
    Bhavsar, Shreyas
    Prabhu, Sujit
    Ince, Nuri F.
    [J]. CLINICAL NEUROPHYSIOLOGY, 2023, 145 : 1 - 10
  • [2] Online denoising of eye-blinks in electroencephalography
    Barthelemy, Quentin
    Mayaud, Louis
    Renard, Yann
    Kim, Daekeun
    Kang, Seung-Wan
    Gunkelman, Jay
    Congedo, Marco
    [J]. NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 2017, 47 (5-6): : 371 - 391
  • [3] Towards non-invasive multi-unit spike recordings: Mapping 1 kHz EEG signals over human somatosensory cortex
    Fedele, T.
    Scheer, H. J.
    Waterstraat, G.
    Telenczuk, B.
    Burghoff, M.
    Curio, G.
    [J]. CLINICAL NEUROPHYSIOLOGY, 2012, 123 (12) : 2370 - 2376
  • [4] Intraoperative subdural low-noise EEG recording of the high frequency oscillation in the somatosensory evoked potential
    Fedele, Tommaso
    Schonenberger, Claudio
    Curio, Gabriel
    Serra, Carlo
    Krayenbuhl, Niklaus
    Sarnthein, Johannes
    [J]. CLINICAL NEUROPHYSIOLOGY, 2017, 128 (10) : 1851 - 1857
  • [5] A Lead Field Two-Domain Model for Longitudinal Neural Tracts-Analytical Framework and Implications for Signal Bandwidth
    Fischer, G.
    Kofler, M.
    Handler, M.
    Baumgarten, D.
    [J]. COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2020, 2020
  • [6] Fischer G., 2023, Biomed. Signal Process
  • [7] Primary somatosensory contextual modulation is encoded by oscillation frequency change
    Goetz, T.
    Milde, T.
    Curio, G.
    Debener, S.
    Lehmann, T.
    Leistritz, L.
    Witte, O. W.
    Witte, H.
    Haueisen, J.
    [J]. CLINICAL NEUROPHYSIOLOGY, 2015, 126 (09) : 1769 - 1779
  • [8] SPHARA - A Generalized Spatial Fourier Analysis for Multi-Sensor Systems with Non-Uniformly Arranged Sensors: Application to EEG
    Graichen, Uwe
    Eichardt, Roland
    Fiedler, Patrique
    Strohmeier, Daniel
    Zanow, Frank
    Haueisen, Jens
    [J]. PLOS ONE, 2015, 10 (04):
  • [9] Guerrero Federico N., 2017, Journal of Medical Engineering & Technology, V41, P546, DOI 10.1080/03091902.2017.1366562
  • [10] Multiplicity in the high-frequency signals during the short-latency somatosensory evoked cortical activity in humans
    Haueisen, J
    Schack, B
    Meier, T
    Curio, G
    Okada, Y
    [J]. CLINICAL NEUROPHYSIOLOGY, 2001, 112 (07) : 1316 - 1325