Comparison of Intraocular Pressure Measurements Obtained Using Three Tonometers after Descemet Membrane Stripping Endothelial Keratoplasty

被引:0
作者
Choi, Jeong Mun [1 ]
Kang, Yong Koo [1 ]
Kim, Hong Kyun [1 ]
Kim, Dai Woo [1 ,2 ]
机构
[1] Kyungpook Natl Univ, Sch Med, Dept Ophthalmol, Daegu, South Korea
[2] Kyungpook Natl Univ, Sch Med, Dept Ophthalmol, 130 Dongdeok Ro, Daegu 41944, South Korea
来源
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY | 2023年 / 64卷 / 07期
关键词
Descemet-stripping endothelial keratoplasty; Goldmann applanation tonometer; Icare ic200; Noncontact tonometer; Rebound tonometer; GOLDMANN APPLANATION; CORNEAL THICKNESS; REBOUND TONOMETER; ICARE TONOMETER; RISK-FACTORS; GLAUCOMA; NONCONTACT; ELEVATION; IC200; EYES;
D O I
10.3341/jkos.2023.64.7.605
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the level of agreement between the Goldmann applanation tonometer (GAT), iCare IC200 rebound tonometer (IRT), and noncontact tonometer (NCT) in patients who underwent Descemet membrane stripping endothelial keratoplasty (DSEK), and to identify factors contributing to variations in intraocular pressure (IOP) measurements among the three tonometers.Methods: We retrospectively analyzed the medical records of 41 patients who underwent DSEK. IOP was measured using NCT, IRT, and GAT, in this order. We evaluated the level of agreement among IOP measurements using the three tonometers, and analyzed whether clinical factors affected the results.Results: We analyzed 49 eyes of 41 patients (average age: 62.0 years). The IOP values measured by IRT and NCT were lower than those measured by GAT, although the difference was not significant (p = 0.098 and p = 0.320, respectively). A Bland-Altman plot showed greater agreement between IOP measurements obtained by IRT and GAT than those obtained by NCT and GAT. In multivariate regression analysis, the IOP measured by GAT (8 = 0.215, p = 0.022), corneal curvature (8 =-1.692, p = 0.037), and postoperative duration (8 = 0.042, p = 0.018) affected the difference in IOPs measured by GAT and IRT. The IOP measured by GAT (8 = 0.301, p = 0.013) and corneal curvature (8 =-2.670, p = 0.010) affected the difference in IOP measurements obtained by GAT and NCT.Conclusions: In DSEK eyes, IRT showed good agreement and high correlation with GAT, suggesting that it is useful for IOP measurement. However, the IOP measured by GAT, corneal curvature, and postoperative duration should be considered when measuring IOP with an IRT.
引用
收藏
页码:605 / 612
页数:8
相关论文
共 27 条
[1]   Intraocular pressure measurement after DSAEK by iCare, Goldmann applanation and dynamic contour tonometry: A comparative study [J].
Achiron, A. ;
Blumenfeld, O. ;
Avizemer, H. ;
Karmona, L. ;
Leybowich, G. ;
Man, V. ;
Bartov, E. ;
Burgansky-Eliash, Z. .
JOURNAL FRANCAIS D OPHTALMOLOGIE, 2016, 39 (10) :822-828
[2]   Agreement of Intraocular Pressure Measurement of Icare ic200 with Goldmann Applanation Tonometer in Adult Eyes with Normal Cornea [J].
Badakere, Swathi Vallabh ;
Chary, Raghava ;
Choudhari, Nikhil S. ;
Rao, Harsha L. ;
Garudadri, Chandrasekhar ;
Senthil, Sirisha .
OPHTHALMOLOGY GLAUCOMA, 2021, 4 (01) :89-94
[3]   Risk factors for endothelial cell loss post-keratoplasty [J].
Bertelmann, Eckart ;
Pleyer, Uwe ;
Rieck, Peter .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (06) :766-770
[4]   Glaucoma [J].
Coleman, AL .
LANCET, 1999, 354 (9192) :1803-1810
[5]   Corneal Transplantation and Glaucoma [J].
Haddadin, Ramez I. ;
Chodosh, James .
SEMINARS IN OPHTHALMOLOGY, 2014, 29 (5-6) :380-396
[6]   The Influence of Central Corneal Thickness and Corneal Curvature Radius on The Intraocular Pressure as Measured By Different Tonometers: Noncontact and Goldmann Applanation Tonometers [J].
Harada, Yukinori ;
Hirose, Naofumi ;
Kubota, Toshiaki ;
Tawara, Akihiko .
JOURNAL OF GLAUCOMA, 2008, 17 (08) :619-625
[7]   Clinical Evaluation of a Rebound Tonometer in Patients Who Underwent Penetrating Keratoplasty [J].
Kim, Jong Woo ;
Jung, Ji Won .
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2018, 59 (12) :1122-1128
[8]   Usefulness of the Icare ic200 Rebound Tonometer in Korean [J].
Lee, Jong-uk ;
Park, Kee Sup ;
Kim, Jaeyoung ;
Lee, Sung Bok ;
Lee, Yeon Hee ;
Kim, Kyoung Nam .
JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2021, 62 (05) :665-673
[9]   Intraocular pressure elevation and post-DSEK glaucoma after Descemet's stripping endothelial keratoplasty [J].
Maier, Anna-Karina B. ;
Klamann, Matthias K. J. ;
Torun, Necip ;
Gonnermann, Johannes ;
Schroeter, Jan ;
Joussen, Antonia M. ;
Rieck, Peter .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 251 (04) :1191-1198
[10]   Relationship between intraocular pressure and glaucoma onset and progression [J].
Miglior, Stefano ;
Bertuzzi, Francesca .
CURRENT OPINION IN PHARMACOLOGY, 2013, 13 (01) :32-35