A phantom for the quantitative determination and improvement of the spatial resolution in slice-selective 2D-FT magnetic resonance micro-imaging and -microscopy based on Deep X-ray Lithography (DXRL)

被引:0
作者
Berg, Andreas Georg [1 ,2 ]
Boerner, Martin [3 ,4 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, MR Phys, Vienna, Austria
[2] Med Univ Vienna, High Field MR Ctr, Vienna, Austria
[3] Karlsruhe Inst Technol KIT, Karlsruhe Nanomicro Facil KNMFi, Karlsruhe, Germany
[4] KIT, Inst Microstruct Technol IMT, Karlsruhe, Germany
关键词
imaging; resolution; quantification; phantom; modulation-transfer-function; MTF; microscopy; XRL; MODULATION TRANSFER-FUNCTION; POINT-SPREAD-FUNCTION; DEGENERATIVE CHANGES; FINGER; TISSUE; MRI;
D O I
10.3389/fphy.2023.1144112
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Introduction: The most important assessed quality-control (QC) criteria for improvements in high-resolution imaging are represented by the contrast-to-noise-ratio and spatial resolution. Ultra-High-Field (UHF) Magnetic-Resonance-scanners (B >= 7 T) for medical research allowed for the improvement in spatial resolution up to the microimaging and nominal microscopy range [pixel-size: ps < (100 mu m)(2)], even in-vivo on humans just recently. Preclinical MRI- and dedicated MR-microscopy (MRM) scanners already allow for microimaging and MRM (1-256 mu m) but lack a sensible spatial resolution phantom for QC and performance improvements in hardware, pulse-sequencing and MRprotocols. In most scientific MRI articles, the spatial resolution is characterized by the ps, though this measurement parameter only limits the actual resolution. Methods: Here the Modulation-Transfer-Function (MTF) is used as evaluation concept for the determination of the spatial resolution in MRM using simple intensity profiles. The resolution limit is defined using a critical modulation-level. In approaching visual impressions on spatial resolution an additional criterion derived from the Modulation-depth-to-Noise-Ratio (MNR) is proposed. A practical method for assessment based on a concrete phantom design and its realization is shown. Results: The phantom design consists of several sets of fine grids, specifically featuring high structural anisotropy for optimum SNR and CNR, with different spatial periods ranging from a(1) = 256 mu m down to a(8) = 2 mu m, not only for a quick visual qualitative check, but also for quantification of resolution using the MTF for two different spatial encodings in two orthogonal in-plane directions. The challenging demands on the manufacturing technology especially with regard to the aspect-ratio are approached using Deep-X-Ray-Lithography (DXRL) relying on the high brilliance of Synchroton-radiation. Smallest grid plates with width of 4 mu m corresponding to 125 line pairs/mm at a plate depth of 100 mu m were achieved. Discussion: MR-microscopic images, originating from a microscopy insert on a human UHF-MR-scanner, were used for demonstration of the evaluation process with two independent resolution-criteria. The developed prototype offers unique possibilities for quantitative resolution QC on UHF human and preclinical MR-scanners. Such a resolution-phantom might be very important for the improvement of MR-pulse-sequences, MR-protocols and even hardware. In principle the phantom can also be used for other microscopic imaging-modalities as for instance mu CT and Optical-Coherence-Tomography (OCT).
引用
收藏
页数:17
相关论文
共 38 条
[1]  
American College of Radiolog, 2022, MR ACCR PROGR PHANT
[2]   Uncertainty analysis in polymer gel dosimetry [J].
Baldock, C ;
Murry, P ;
Kron, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) :N243-N246
[3]   High-resolution diffusivity Imaging at 3.0 T for the detection of degenerative changes - A trypsin-based arthritis model [J].
Berg, A ;
Singer, T ;
Moser, E .
INVESTIGATIVE RADIOLOGY, 2003, 38 (07) :460-466
[4]   Diffusivity- and T2 imaging at 3 Tesla for the detection of degenerative changes in human-excised tissue with high resolution -: Atherosclerotic arteries [J].
Berg, A ;
Sailer, J ;
Rand, T ;
Moser, E .
INVESTIGATIVE RADIOLOGY, 2003, 38 (07) :452-459
[5]  
Berg A., 2010, P ISMRM ESMRMB 2010, P1048
[6]  
Berg A., 2020, P INT SOC MAG RESON, V28, P4267
[7]  
Berg A., 2001, PROC ISMRMESMRMB JOI, P925
[8]  
Berg A., 1999, BERG A
[9]  
Bruker, 2022, PROD SOL PRECL IM
[10]  
Callaghan P., 2003, Principles of Nuclear Magnetic Resonance Microscopy"