Stem alignment with short tapered-wedge cementless stems affects bone mineral density continuously for at least 5 years after primary total hip arthroplasty

被引:3
作者
Masuda, Sho [1 ,2 ]
Iwakiri, Kentaro [1 ]
Ohta, Yoichi [3 ]
Minoda, Yukihide [3 ]
Kobayashi, Akio [1 ]
Nakamura, Hiroaki [3 ]
机构
[1] Shiraniwa Hosp Joint Arthroplasty Ctr, Dept Orthopaed Surg, 6-10-1 Shiraniwadai Ikoma City, Nara 6300136, Japan
[2] Osaka Saiseikai Nakatsu Hosp, Dept Orthopaed Surg, Osaka, Japan
[3] Osaka Metropolitan Univ, Dept Orthopaed Surg, Grad Sch Med, Osaka, Japan
关键词
Bone mineral density; flexed insertion; short tapered wedge cementless stem; 3D stem alignment; total hip arthroplasty; varus insertion; FOLLOW-UP; REPLACEMENT; PERFORMANCE; COMPONENT; INCISION; FIXATION;
D O I
10.1177/11207000231178984
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aim: This study aimed to assess the relationship between the three-dimensional (3D) alignment of short tapered-wedge cementless stems and bone mineral density (BMD) changes in patients followed up for 5 years after total hip arthroplasty (THA). Methods: We retrospectively analysed the hips of 52 patients who underwent THA using short tapered-wedge cementless stems at our institution from 2013 to 2016 with complete 5-year follow-up data. We evaluated the relationship between stem alignment, measured using a 3D-templating software, and BMD changes in the 7 Gruen zones. Results: After 1 year, significant negative correlations between varus insertion and a decrease in BMD in zone 7 and between flexed insertion and decreases in BMD in zones 3 and 4 were noted. After 5 years, significant negative correlations between varus insertion and a decrease in BMD in zone 7 and between flexed insertion and decreases in BMD zones 2, 3, and 4 were observed. With increased amounts of varus/flexion stem alignment, the amount of BMD loss decreased. There was no correlation between anteverted stem insertion and changes in BMD levels. Conclusions: Our data showed that stem alignment affects BMD based on 5-year follow-up data after surgery. Careful observation is necessary, especially when using short tapered-wedge cementless stems, as stem alignment may affect changes in BMD levels more than 5 years after surgery.
引用
收藏
页码:57 / 65
页数:9
相关论文
共 31 条
  • [1] Favorable Results of a Short, Tapered, Highly Porous, Proximally Coated Cementless Femoral Stem at a Minimum 4-Year Follow-Up
    Albers, Anthony
    Aoude, Ahmed A.
    Zukor, David J.
    Huk, Olga L.
    Antoniou, John
    Tanzer, Michael
    [J]. JOURNAL OF ARTHROPLASTY, 2016, 31 (04) : 824 - 829
  • [2] Efficient computational method for assessing the effects of implant positioning in cementless total hip replacements
    Bah, Mamadou T.
    Nair, Prasanth B.
    Taylor, Mark
    Browne, Martin
    [J]. JOURNAL OF BIOMECHANICS, 2011, 44 (07) : 1417 - 1422
  • [3] The Short and "Shorter" of It: >1750 Tapered Titanium Stems at 6-to 88-Month Follow-Up
    Barrington, John W.
    Emerson, Roger H., Jr.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (08) : 38 - 40
  • [4] Importance of precision in bone density measurements
    Bonnick, SL
    Johnston, CC
    Kleerekoper, M
    Lindsay, R
    Miller, P
    Sherwood, L
    Siris, E
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2001, 4 (02) : 105 - 110
  • [5] Periprosthetic Bone Remodeling Using a Triple-Taper Polished Cemented Stem in Total Hip Arthroplasty
    Buck, Aaron J.
    Dowsey, Michelle M.
    Stoney, James D.
    Hardidge, Andrew J.
    Ng, Kong Wah
    Choong, Peter F. M.
    [J]. JOURNAL OF ARTHROPLASTY, 2010, 25 (07) : 1083 - 1090
  • [6] Distal Fixation of Proximally Coated Tapered Stems May Predispose to a Failure of Osteointegration
    Cooper, H. John
    Jacob, Anitha P.
    Rodriguez, Jose A.
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (06) : 78 - 83
  • [7] de Beer J, 2006, CAN J SURG, V49, P407
  • [8] STRUCTURAL AND CELLULAR ASSESSMENT OF BONE QUALITY OF PROXIMAL FEMUR
    DORR, LD
    FAUGERE, MC
    MACKEL, AM
    GRUEN, TA
    BOGNAR, B
    MALLUCHE, HH
    [J]. BONE, 1993, 14 (03) : 231 - 242
  • [9] A QUANTITATIVE-EVALUATION OF PERIPROSTHETIC BONE-REMODELING AFTER CEMENTLESS TOTAL HIP-ARTHROPLASTY
    ENGH, CA
    MCGOVERN, TF
    BOBYN, JD
    HARRIS, WH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) : 1009 - 1020
  • [10] POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS
    ENGH, CA
    BOBYN, JD
    GLASSMAN, AH