摘要
Glaucoma surgery with implantation of aqueous humor draining microstents may compromise long-term corneal health by disrupting aqueous humor circulation. The effect of stent numbers on this circulation was interrogated to determine the number of stents associated with minimal circulation disruption. An in vitro anterior eye model perfusion system was constructed with multiple exit ports. A 3-D model of the anterior eye was imported into ABAQUS CFD, analyzes were carried out for unsteady laminar flow and solved using Navier-Stokes equations. DT Vision Foundry was used to analyze velocity contour plot images. The field variable results output for the CFD model were fluid wall shear, fluid pressure and fluid velocity. In vitro, "aqueous" fluid flow is high through a single stent and "aqueous" stagnation is greatest in the quadrants 180 degrees away. Increasing stent port numbers, results in an exponential decrease in the stagnant flow locations. High wall shear stress was seen in the single stent model and is markedly reduced after a second and subsequent stents are introduced. We identify two factors potentially contributing to corneal compromise post glaucoma drainage surgery: aqueous humor stagnation, remote to the stent site and higher exit flows imparting increased stent exit shear stress (particularly with a single stent). With 4 stents, there is minimal disruption of anterior chamber circulation (mimicking physiological conditions). Furthermore we propose that aqueous humor circulation disruption via the usual single-exit port approach disrupts aqueous humor circulation with long-term consequences for corneal health.
引用
收藏
页码:898 / 907
页数:10
相关论文
共 53 条
  • [1] Investigation of Saccadic Eye Movement Effects on the Fluid Dynamic in the Anterior Chamber
    Abouali, Omid
    Modareszadeh, Amirreza
    Ghaffarieh, Alireza
    Tu, Jiyuan
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2012, 134 (02):
  • [2] Outcome and Prognosis of Pediatric Patients With Delayed Diagnosis of Open-Globe Injuries
    Acuna, Olga M.
    Yen, Kimberly G.
    [J]. JOURNAL OF PEDIATRIC OPHTHALMOLOGY & STRABISMUS, 2009, 46 (04) : 202 - 207
  • [3] Agar A, 2000, J NEUROSCI RES, V60, P495, DOI 10.1002/(SICI)1097-4547(20000515)60:4<495::AID-JNR8>3.0.CO
  • [4] 2-S
  • [5] Retinal ganglion cell line apoptosis induced by hydrostatic pressure
    Agar, Ashish
    Li, Shaojuan
    Agarwal, Neeraj
    Coroneo, Minas T.
    Hill, Mark A.
    [J]. BRAIN RESEARCH, 2006, 1086 : 191 - 200
  • [6] Aminlari Ardalan E, 2013, Middle East Afr J Ophthalmol, V20, P126, DOI 10.4103/0974-9233.110607
  • [7] Maintaining transparency: A review of the developmental physiology and pathophysiology of two avascular tissues
    Beebe, David C.
    [J]. SEMINARS IN CELL & DEVELOPMENTAL BIOLOGY, 2008, 19 (02) : 125 - 133
  • [8] Effect of Obstructive Sleep Apnea on Corneal Morphological Characteristics
    Bojarun, Andrius
    Vieversyte, Zivile
    Jaruseviciene, Ruta
    Galgauskas, Saulius
    Asoklis, Rimvydas
    Zablockis, Rolandas
    [J]. CORNEA, 2019, 38 (12) : 1576 - 1581
  • [9] BRUBAKER RF, 1991, INVEST OPHTH VIS SCI, V32, P3145
  • [10] Variations in active outflow along the trabecular outflow pathway
    Cha, Elliott D. K.
    Xu, Jia
    Gong, Lihua
    Gong, Haiyan
    [J]. EXPERIMENTAL EYE RESEARCH, 2016, 146 : 354 - 360