Partial femoral head replacement: a new innovative hip-preserving approach for treating osteonecrosis of the femoral head and its finite element analysis

被引:4
作者
Yang, Yanjiang [1 ,2 ,3 ]
Cheng, Xiaodong [1 ,2 ,3 ]
Chen, Wei [1 ,2 ,3 ]
Li, Guimiao [1 ,2 ,3 ]
Wang, Yuchuan [1 ,2 ,3 ]
Sun, Weiyi [1 ,2 ,3 ]
An, Wen [1 ,2 ,3 ]
Zhang, Qi [1 ,2 ,3 ]
Zhang, Yingze [1 ,2 ,3 ,4 ,5 ]
机构
[1] Hebei Med Univ, Dept Emergency, Hosp 3, Shijiazhuang, Hebei, Peoples R China
[2] Orthopaed Res Inst Hebei Prov, Shijiazhuang, Hebei, Peoples R China
[3] Key Lab Biomech Hebei Prov, Shijiazhuang, Hebei, Peoples R China
[4] NHC Key Lab Intelligent Orthopaed Equipment, Shijiazhuang, Hebei, Peoples R China
[5] Hebei Orthopaed Clin Res Ctr, Shijiazhuang, Hebei, Peoples R China
基金
中国国家自然科学基金;
关键词
femoral head replacement; hip-preserving; osteonecrosis of the femoral head; finite element analysis; von Mises stress; CORE DECOMPRESSION; AVASCULAR NECROSIS;
D O I
10.3389/fbioe.2024.1352882
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Purpose: Controversy remains regarding the optimal treatment for stage III Osteonecrosis of the femoral head (ONFH). This study presents, for the first time, the precise treatment of stage III ONFH using the "substitute the beam for a pillar" technique and performs a comparative finite element analysis with other hip-preserving procedures.Methods: A formalin-preserved femur of male cadavers was selected to obtain the CT scan data of femur. The proximal femur model was reconstructed and assembled using Mimics 20.0, Geomagic, and UG-NX 12.0 software with four different implant types: simple core decompression, fibula implantation, porous tantalum rod implantation, and partial replacement prosthesis. The finite element simulations were conducted to simulate the normal walking gait, and the stress distribution and displacement data of the femur and the implant model were obtained.Results: The peak von Mises stress of the femoral head and proximal femur in the partial replacement of the femoral head (PRFH) group were 22.8 MPa and 37.4 MPa, respectively, which were 3.1%-38.6% and 12.8%-37.4% lower than those of the other three surgical methods.Conclusion: The PRFH group exhibits better mechanical performance, reducing stress and displacement in the ONFH area, thus maintaining femoral head stability. Among the four hip-preserving approaches, from a biomechanical perspective, PRFH offers a new option for treating ONFH.
引用
收藏
页数:9
相关论文
共 50 条
[41]   Location or size? A finite element analysis study of necrotic lesion impact on femoral head collapse [J].
Lu, Shun ;
Lin, Tianye ;
Han, Longfei ;
Li, Ziqi ;
He, Mingcong ;
Wei, Qiushi .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2025, 20 (01)
[42]   Biomechanical supporting effect of tantalum rods for the femoral head with various sized lesions: a finite-element analysis [J].
Liu Wen-guang ;
Wang Shao-jin ;
Yin Qing-feng ;
Liu Sheng-hou ;
Guan Yan-jin .
CHINESE MEDICAL JOURNAL, 2012, 125 (22) :4061-4065
[43]   Does spinopelvic alignment affect femoral head cartilage and the proximal femoral physis in slipped capital femoral epiphysis? A finite element analysis [J].
Kumaran, Yogesh ;
Mumtaz, Muzammil ;
Quatman, Carmen ;
Balch-Samora, Julie ;
Soehnlen, Sophia ;
Hoffman, Brett ;
Tripathi, Sudharshan ;
Nishida, Norihiro ;
Goel, Vijay K. .
CLINICAL BIOMECHANICS, 2024, 116
[44]   Evaluation of the long-term patient-reported outcomes after hip arthroplasty or joint preserving with Sugioka femoral osteotomy in patients with femoral head osteonecrosis [J].
Shoji, Takeshi ;
Shozen, Hideki ;
Ueki, Shinichi ;
Kaneta, Hiroki ;
Yaunaga, Yuji ;
Adachi, Nobuo .
INTERNATIONAL ORTHOPAEDICS, 2024, 48 (05) :1201-1208
[45]   Evaluation of the long-term patient-reported outcomes after hip arthroplasty or joint preserving with Sugioka femoral osteotomy in patients with femoral head osteonecrosis [J].
Takeshi Shoji ;
Hideki Shozen ;
Shinichi Ueki ;
Hiroki Kaneta ;
Yuji Yaunaga ;
Nobuo Adachi .
International Orthopaedics, 2024, 48 :1201-1208
[46]   Risk Prediction of Femoral Head Necrosis: A Finite Element Analysis Based on Fracture Mechanics [J].
Li, Honggeng ;
Li, Fan ;
Liu, Niansong ;
Li, Pengfei .
INTERNATIONAL JOURNAL OF COMPUTATIONAL METHODS, 2020, 17 (06)
[47]   Lateral pillar is the key in supporting pre-collapse osteonecrosis of the femoral head: a finite element model analysis of propensity-score matched cohorts [J].
Bahk, Ji Hoon ;
Jo, Woo-Lam ;
Kim, Seung-Chan ;
Kwon, Soon-Yong ;
Lim, Young Wook .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
[48]   Lateral pillar is the key in supporting pre-collapse osteonecrosis of the femoral head: a finite element model analysis of propensity-score matched cohorts [J].
Ji Hoon Bahk ;
Woo-Lam Jo ;
Seung-Chan Kim ;
Soon-Yong Kwon ;
Young Wook Lim .
Journal of Orthopaedic Surgery and Research, 16
[49]   A new modification of the Trapdoor Procedure without surgical hip dislocation in the management of patients with osteonecrosis of the femoral head: A retrospective case series [J].
Akman, Budak ;
Guven, Melih ;
Kadioglu, Baris ;
Ozkan, Namik Kemal ;
Yonga, Omer ;
Ozturan, Burak .
ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2021, 55 (04) :299-305
[50]   Numerical studies on alternative therapies for femoral head necrosisA finite element approach and clinical experience [J].
André Lutz ;
Udo Nackenhorst ;
Gabriela von Lewinski ;
Henning Windhagen ;
Thilo Floerkemeier .
Biomechanics and Modeling in Mechanobiology, 2011, 10 :627-640