Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy

被引:0
作者
Chirdchid, Thamonwan [1 ]
Ruangchan, Sirinya [2 ,3 ]
Sanghangthum, Taweap [1 ,2 ,4 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Fac Med, Bangkok, Thailand
[2] Her Royal Highness Princess Maha Chakri Sirindhor, Bangkok, Thailand
[3] Thai Red Cross Soc, King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, Bangkok, Thailand
[4] Chulalongkorn Univ, Div Radiat Oncol, Dept Radiol, Fac Med, Bangkok, Thailand
关键词
Dual-energy computed tomography; proton therapy; single-energy computed tomography; stoichiometric method; DOSE CALCULATION; CT; CALIBRATION;
D O I
10.4103/jmp.jmp_27_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. Materials and Methods: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose-volume histograms (DVHs) comparison. Results: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% +/- 2.1% and 2.5% +/- 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. Conclusion: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
引用
收藏
页码:292 / 297
页数:6
相关论文
共 14 条
  • [11] Schulte Reinhard W, 2012, Trans Am Nucl Soc, V106, P55
  • [12] Particle radiation therapy using proton and heavier ion beams
    Schulz-Ertner, Daniela
    Tsujii, Hirohiko
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2007, 25 (08) : 953 - 964
  • [13] Khan's the physics of radiation therapy 6th edition
    Semwal, Manoj Kumar
    [J]. JOURNAL OF MEDICAL PHYSICS, 2020, 45 (02) : 134 - 135
  • [14] Clinical Implementation of Dual-energy CT for Proton Treatment Planning on Pseudo-monoenergetic CT scans
    Wohlfahrt, Patrick
    Moehler, Christian
    Hietschold, Volker
    Menkel, Stefan
    Greilich, Steffen
    Krause, Mechthild
    Baumann, Michael
    Enghardt, Wolfgang
    Richter, Christian
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 97 (02): : 427 - 434