Magnesium implants in orthopaedic surgery create a diagnostic conundrum: A radiology case series and literature review

被引:4
作者
Anderson, Toni [1 ]
Lee, Jennifer [1 ]
Johnston, Philip [1 ]
Torreggiani, William [1 ]
Ryan, Martin [1 ]
机构
[1] Tallaght Univ Hosp, Dept Radiol, Dublin 24, Ireland
关键词
Bioabsorbable screws; Biodegradable; Magnesium; Orthopaedic screws; FIXATION; ALLOYS;
D O I
10.1007/s11845-022-03071-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
During a routine post-operative orthopaedic radiograph reading session, repeated unusual radiographic soft tissue and bone appearances became evident. It was discovered that these patients had received biodegradable magnesium implants which have recently been introduced into orthopaedic clinical practice. To the untrained eye, the combination of peri-metallic bone resorption with associated soft tissue gas, could easily be mistaken for post-operative infection. The aim of this study is to properly characterise the radiographic post-operative appearances of biodegradable magnesium orthopaedic hardware. We retrospectively evaluated radiographs of all patients who underwent magnesium screw implantation for fractures over a 6 month period. Four patients, mean age of 9.75 (range: 6-15) years who underwent magnesium screw fixation following fracture were included in the study. Follow up duration was 100 days (range: 75-122) with a mean of 2.5 postoperative radiographs being performed per patient during this period. All cases demonstrated post-operative peri-metallic radiolucency which developed around the magnesium screws on x-ray, which subsequently resorbed over time. Peri-metallic soft tissue gas was observed in all patients. In two cases, magnesium implants fractured. As the use of biodegradable metal implants becomes more common, it is important for radiologists to be aware of their imaging characteristics. Prior to reporting a case, it would be prudent to know if biodegradable screws have been utilised and whether there exists a clinical concern for post-operative infection in patients with these particular implants, in which case it would be critical not to dismiss peri-prosthetic radiolucencies and soft tissue gas as merely a sequela of the natural metal degradation process.
引用
收藏
页码:1381 / 1385
页数:5
相关论文
共 38 条
  • [31] Utilization of 3D-Printed Customized Uncemented Stem Prostheses for Revision of Aseptic Loosening in the Distal Femoral Cemented Prostheses: Case Series and Literature Review
    Hou, Zi-Wei
    Zheng, Kai
    Xu, Ming
    Yu, Xiu-Chun
    ORTHOPAEDIC SURGERY, 2025, 17 (03) : 801 - 813
  • [32] Atlantoaxial Stabilization by Posterior C1 and C2 Screw-Rod Fixation for Various Pathologies: Case Series and Comprehensive Review of Literature
    Unni, Chandramohan
    Pettakkandy, Vijayan
    Jaleel, Abdul P.
    Soren, Subrat Kumar
    Vineeth, K. K.
    JOURNAL OF NEUROSCIENCES IN RURAL PRACTICE, 2021, 12 (02) : 228 - 235
  • [33] Spondylolysis Repair Using a Minimally Invasive Modified Buck Technique with Neuronavigation and Neuromonitoring in High School and Professional Athletes: Technical Notes, Case Series, and Literature Review
    Yurac, Ratko
    Bravo, Jose T.
    Silva, Alvaro
    Marre, Bartolome
    WORLD NEUROSURGERY, 2021, 155 : 54 - 63
  • [34] Vulnerability of the L5 nerve root during anterior lumbar interbody fusion at L5-S1: case series and review of the literature
    Dowlati, Ehsan
    Alexander, Hepzibha
    Voyadzis, Jean-Marc
    NEUROSURGICAL FOCUS, 2020, 49 (03) : 1 - 9
  • [35] Augmented ilio-sacral screws for the treatment of fragility pelvic fractures: review of literature, presentation of a novel low-cost technique, and clinical results of a case series
    Castro, Hari
    Carmona, Maximiliano
    Zamora, Tomas
    Klaber, Ianiv
    Botello, Eduardo
    Faundez, Jorge
    Schweitzer, Daniel
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2024, 35 (01)
  • [36] Minimally invasive transforaminal lumbar interbody fusion using bone cement-augmented pedicle screws for lumbar spondylolisthesis in patients with osteoporosis. Case series and review of literature
    Vemula, Venkata Ramesh Chandra
    Prasad, Bodapati Chandramowliswara
    Jagadeesh, M. A.
    Vuttarkar, Jayachandar
    Akula, Sanjeev Kumar
    NEUROLOGY INDIA, 2018, 66 (01) : 118 - 125
  • [37] Relationship of Frontal Plane Rotation of First Metatarsal to Proximal Articular Set Angle and Hallux Alignment in Patients Undergoing Tarsometatarsal Arthrodesis for Hallux Abducto Valgus: A Case Series and Critical Review of the Literature
    Dayton, Paul
    Feilmeier, Mindi
    Kauwe, Merrell
    Hirschi, Jordan
    JOURNAL OF FOOT & ANKLE SURGERY, 2013, 52 (03) : 348 - 354
  • [38] Cervical kyphosis surgery using a cervical pedicle screw placed with a U-shaped wire that enables observation of the lateral edge of the cortical bone of the spinal canal: A case report and literature review
    Ushiku, Chikara
    Akiyama, Shoshi
    Kanai, Tomoaki
    Sawada, Naomu
    Saito, Mitsuru
    MEDICINE, 2023, 102 (46) : E36088