Evaluation of the Qvella FAST System and the FAST-PBC cartridge for rapid species identification and antimicrobial resistance testing directly from positive blood cultures

被引:2
作者
Sy, Issa [1 ]
Buehler, Nina [1 ]
Becker, Soeren L. [1 ,2 ,3 ]
Jung, Philipp [1 ]
Richter, Sandra S.
机构
[1] Saarland Univ, Inst Med Microbiol & Hyg, Homburg, Germany
[2] Swiss Trop & Publ Hlth Inst, Allschwil, Switzerland
[3] Univ Basel, Basel, Switzerland
关键词
bloodstream infection; blood culture; diagnosis; antimicrobial susceptibility testing; Qvella FAST System; matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS); MicroScan WalkAway; disk diffusion; STREAM INFECTIONS; FILMARRAY; BACTERIA; BURDEN;
D O I
10.1128/jcm.00569-23
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Blood culture diagnostics require rapid and accurate identification (ID) of pathogens and antimicrobial susceptibility testing (AST). Standard procedures, involving conventional cultivation on agar plates, may take up to 48 hours or more until AST completion. Recent approaches aim to shorten the processing time of positive blood cultures (PBC). The FAST System is a new technology, capable of purifying and concentrating bacterial/fungal pathogens from positive blood culture media and producing a bacterial suspension called "liquid colony" (LC), which can be further used in downstream analyses (e.g., ID and AST). Here, we evaluated the performance of the FAST System LC generated from PBC in comparison to our routine workflow including ID by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry using Sepsityper, AST by automatized MicroScan WalkAway plus and directly inoculated disk diffusion (DD), and MICRONAUT-AM for yeast/fungi. A total of 261 samples were analyzed, of which 86.6% (226/261) were eligible for the comparative ID and AST analyses. In comparison to the reference technique (culture-grown colonies), ID concordance of the FAST System LC and Sepsityper was 150/154 (97.4%) and 123/154 (79.9%), respectively, for Gram positive; 67/70 (95.7%) and 64/70 (91.4%), respectively, for Gram negative. For AST, categorical agreement (CA) of the FAST System LC in comparison to the routine workflow for Gram-positive bacteria was 96.1% and 98.7% for MicroScan and DD, respectively. Similar results were obtained for Gram-negative bacteria with 96.6% and 97.5% of CA for MicroScan and DD, respectively. Taken together, the FAST System LC allowed the laboratory to significantly reduce the time to obtain correct ID and AST (automated MicroScan) results 1 day earlier and represents a promising tool to expedite the processing of PBC.
引用
收藏
页数:13
相关论文
共 22 条
[1]   Identification of microorganisms directly from blood culture bottles with polymicrobial growth: comparison of FilmArray and direct MALDI-TOF MS [J].
Almuhayawi, Mohammed S. ;
Wong, Alicia Y. W. ;
Kynning, Max ;
Luethje, Petra ;
Ullberg, Mans ;
Ozenci, Volkan .
APMIS, 2021, 129 (04) :178-185
[2]   Clinical Evaluation of the FilmArray Blood Culture Identification Panel in Identification of Bacteria and Yeasts from Positive Blood Culture Bottles [J].
Altun, Osman ;
Almuhayawi, Mohammed ;
Ullberg, Mans ;
Ozenci, Volkan .
JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (12) :4130-4136
[3]   Epidemiology of severe sepsis in the United States: Analysis of incidence, outcome, and associated costs of care [J].
Angus, DC ;
Linde-Zwirble, WT ;
Lidicker, J ;
Clermont, G ;
Carcillo, J ;
Pinsky, MR .
CRITICAL CARE MEDICINE, 2001, 29 (07) :1303-1310
[4]   Evaluation of FilmArray and Verigene Systems for Rapid Identification of Positive Blood Cultures [J].
Bhatti, M. M. ;
Boonlayangoor, S. ;
Beavis, K. G. ;
Tesic, V. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2014, 52 (09) :3433-3436
[5]   Mixed bloodstream infections involving bacteria and Candida spp. [J].
Bouza, Emilio ;
Burillo, Almudena ;
Munoz, Patricia ;
Guinea, Jesus ;
Marin, Mercedes ;
Rodriguez-Creixems, Marta .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2013, 68 (08) :1881-1888
[6]  
Canalp HZ, 2021, MICROBIOL SPECTR, V9, DOI [10.1128/Spectrum.01103-21, 10.1128/spectrum.01103-21]
[7]   Overall burden of bloodstream infection and nosocomial bloodstream infection in North America and Europe [J].
Goto, M. ;
Al-Hasan, M. N. .
CLINICAL MICROBIOLOGY AND INFECTION, 2013, 19 (06) :501-509
[8]   Saving Time in Blood Culture Diagnostics: a Prospective Evaluation of the Qvella FAST-PBC Prep Application on the Fast System [J].
Grinberg, Semjon ;
Schubert, Susanne ;
Hochauf-Stange, Kristina ;
Dalpke, Alexander H. ;
Encalada, Marco Narvaez .
JOURNAL OF CLINICAL MICROBIOLOGY, 2022, 60 (05)
[9]   A global accounting of sepsis [J].
Kempker, Jordan A. ;
Martin, Greg S. .
LANCET, 2020, 395 (10219) :168-170
[10]   Analysis of the FASTTM system for expedited identification and antimicrobial susceptibility testing of bloodborne pathogens [J].
Kuo, Peiting ;
LeCrone, Khrissa ;
Chiu, Megan ;
Realegeno, Susan ;
Pride, David T. .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2022, 104 (04)