Fiberscope-Based Measurement of Coaptation Height for Intraoperative Assessment of Mitral Valve Repair

被引:2
|
作者
Recco, Dominic P. [1 ,2 ]
Kneier, Nicholas E. [1 ]
Earley, Patrick D. [1 ]
Kizilski, Shannen B. [1 ,2 ]
Hammer, Peter E. [1 ,2 ]
Hoganson, David M. [1 ,2 ,3 ]
机构
[1] Boston Childrens Hosp, Dept Cardiac Surg, Boston, MA USA
[2] Harvard Med Sch, Boston, MA USA
[3] Boston Childrens Hosp, Dept Cardiac Surg, 300 Longwood Ave,Bader,2nd Floor, Boston, MA 02215 USA
关键词
mitral valve; mitral regurgitation; mitral valve coaptation; coaptation height; coaptation measurement; congenital valve disease; SYSTOLIC ANTERIOR MOTION; REGURGITATION; REPLACEMENT; SURGERY; RECONSTRUCTION; ANNULOPLASTY; DURABILITY; MANAGEMENT; EVOLUTION; POSTERIOR;
D O I
10.1177/21501351231221459
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Restoring adequate coaptation height is a key principle of mitral valve (MV) repair. This study aimed to evaluate the utility of fiberscope (FS) technology to assess MV coaptation height for intraoperative use.Methods Ex-vivo testing was performed on five adult porcine hearts. The left atrium (LA) was resected, and the left ventricle (LV) was pressurized retrograde to 27 +/- 1mm Hg. An endoscope was inserted into the LV apex, centered under the MV orifice. An FS system (Milliscope II camera, LED light source, and 0.7 mm diameter x 15 cm long) 90 degrees semirigid scope with 1.2 mm focal length) was mounted above the MV annulus in a custom alignment and measuring fixture. Three blinded measurements were taken at two locations on each MV, A2 and P2 segment, from the top of coaptation to the leaflet edge identified by the FS. Accurate positioning was verified using the LV endoscope. A control (metal rod of similar thickness) was used for comparison, with coaptation height recorded when the control was seen via the endoscope.Results Coaptation heights were similar for the control and FS methods across all hearts at A2 (11.6 +/- 2.6 mm control vs 11.8 +/- 2.2 mm FS) and P2 (13.3 +/- 2.6 mm control vs 13.4 +/- 2.9 mm FS) segments, with similar measurement variability (control SD 0.1-1.0 mm; FS SD 0.1-0.9 mm). One outlier was excluded from analysis (n = 19/20). The maximum absolute difference and percent error between measurement methods were less than 1.1 mm (median [IQR], 0.6 [0.3-0.9] mm) and less than 14% (4.1 [2.2-7.6]%).Conclusions Utilization of a miniaturized FS enabled precise and accurate quantification of MV coaptation. This technique is promising for evaluating post-repair valve competence and coaptation height.
引用
收藏
页码:371 / 379
页数:9
相关论文
共 50 条
  • [31] EVALUATION OF MITRAL-VALVE REPAIR BY INTRAOPERATIVE TRANSESOPHAGEAL ECHOCARDIOGRAPHY
    KALMAN, JM
    JONES, EF
    LUBICZ, S
    BUXTON, BB
    TONKIN, AM
    CALAFIORE, P
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1993, 23 (05): : 463 - 469
  • [32] Dobutamine stress echocardiography and intraoperative assessment of mitral valve
    Mahmood, Feroze
    Lerner, Adam B.
    Matyal, Robina
    Karthik, Swaminathan
    Maslow, Andrew D.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2006, 20 (06) : 867 - 871
  • [33] Three-Dimensional Echocardiographic Assessment of Coaptation After Aortic Valve Repair
    Szymanski, Trevor
    Maslow, Andrew
    Mahmood, Feroze
    Singh, Arun
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2017, 31 (03) : 993 - 1000
  • [34] Interobserver and intraobserver variability among different vendors for mitral valve assessment: implications for transcatheter mitral valve repair
    Challa, Apurva B.
    Radike, Monika
    Rizvi, Asim
    Weber, Nikkole M.
    Wamil, Malgorzata
    Arunachalam, Shivaram Poigai
    Sheedy, Emily
    Leng, Shuai
    Williamson, Eric E.
    RADIOLOGIA MEDICA, 2025, 130 (03): : 296 - 301
  • [35] Midterm Results of Mitral Valve Repair Using Loop Technique With Simple Height Reduction of the Large Posterior Leaflet
    Sumii, Yosuke
    Morisaki, Akimasa
    Kawase, Takumi
    Doi, Masataro
    Nagao, Munehide
    Nishimoto, Yukihiro
    Sumii, Yosuke
    Inno, Gouki
    Nishiya, Kenta
    Sakon, Yoshito
    Aoyama, Takanobu
    Shibata, Toshihiko
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2023, 18 (05) : 435 - 444
  • [36] Intraoperative Identification of an Anomalous Cardiac Structure During Mitral Valve Repair
    Codere-Maruyama, Takumi
    Mullen, John
    Berube, Joel
    Yukawa, Masaru
    Sidhu, Surita
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2017, 31 (02) : 773 - 775
  • [37] Measurement of Aortic Valve Coaptation and Effective Height Using Echocardiography in Patients with Ventricular Septal Defects and Aortic Valve Prolapse
    Satoru Iwashima
    Hiroki Uchiyama
    Takamichi Ishikawa
    Kiyohiro Takigiku
    Ken Takahashi
    Manatomo Toyono
    Nao Inoue
    Masaki Nii
    Pediatric Cardiology, 2017, 38 : 608 - 616
  • [38] Machine-learning-based prediction of survival and mitral regurgitation recurrence in patients undergoing mitral valve repair
    Kang, Yoonjin
    Sohn, Suk Ho
    Choi, Jae Woong
    Hwang, Ho Young
    Kim, Kyung Hwan
    INTERDISCIPLINARY CARDIOVASCULAR AND THORACIC SURGERY, 2023, 37 (05):
  • [39] MECHANISM OF OUTFLOW TRACT OBSTRUCTION CAUSING FAILED MITRAL-VALVE REPAIR - ANTERIOR DISPLACEMENT OF LEAFLET COAPTATION
    LEE, KS
    STEWART, WJ
    LEVER, HM
    UNDERWOOD, PL
    COSGROVE, DM
    CIRCULATION, 1993, 88 (05) : 24 - 29
  • [40] Augmenting mitral valve repair evaluation with intraoperative left ventricle pressure measurements
    Issa, Hugo
    Deng, Mimi
    Rahmouni, Kenza
    Chan, Vincent
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2022, 35 (05)