Fiberscope-Based Measurement of Coaptation Height for Intraoperative Assessment of Mitral Valve Repair

被引:2
|
作者
Recco, Dominic P. [1 ,2 ]
Kneier, Nicholas E. [1 ]
Earley, Patrick D. [1 ]
Kizilski, Shannen B. [1 ,2 ]
Hammer, Peter E. [1 ,2 ]
Hoganson, David M. [1 ,2 ,3 ]
机构
[1] Boston Childrens Hosp, Dept Cardiac Surg, Boston, MA USA
[2] Harvard Med Sch, Boston, MA USA
[3] Boston Childrens Hosp, Dept Cardiac Surg, 300 Longwood Ave,Bader,2nd Floor, Boston, MA 02215 USA
关键词
mitral valve; mitral regurgitation; mitral valve coaptation; coaptation height; coaptation measurement; congenital valve disease; SYSTOLIC ANTERIOR MOTION; REGURGITATION; REPLACEMENT; SURGERY; RECONSTRUCTION; ANNULOPLASTY; DURABILITY; MANAGEMENT; EVOLUTION; POSTERIOR;
D O I
10.1177/21501351231221459
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Restoring adequate coaptation height is a key principle of mitral valve (MV) repair. This study aimed to evaluate the utility of fiberscope (FS) technology to assess MV coaptation height for intraoperative use.Methods Ex-vivo testing was performed on five adult porcine hearts. The left atrium (LA) was resected, and the left ventricle (LV) was pressurized retrograde to 27 +/- 1mm Hg. An endoscope was inserted into the LV apex, centered under the MV orifice. An FS system (Milliscope II camera, LED light source, and 0.7 mm diameter x 15 cm long) 90 degrees semirigid scope with 1.2 mm focal length) was mounted above the MV annulus in a custom alignment and measuring fixture. Three blinded measurements were taken at two locations on each MV, A2 and P2 segment, from the top of coaptation to the leaflet edge identified by the FS. Accurate positioning was verified using the LV endoscope. A control (metal rod of similar thickness) was used for comparison, with coaptation height recorded when the control was seen via the endoscope.Results Coaptation heights were similar for the control and FS methods across all hearts at A2 (11.6 +/- 2.6 mm control vs 11.8 +/- 2.2 mm FS) and P2 (13.3 +/- 2.6 mm control vs 13.4 +/- 2.9 mm FS) segments, with similar measurement variability (control SD 0.1-1.0 mm; FS SD 0.1-0.9 mm). One outlier was excluded from analysis (n = 19/20). The maximum absolute difference and percent error between measurement methods were less than 1.1 mm (median [IQR], 0.6 [0.3-0.9] mm) and less than 14% (4.1 [2.2-7.6]%).Conclusions Utilization of a miniaturized FS enabled precise and accurate quantification of MV coaptation. This technique is promising for evaluating post-repair valve competence and coaptation height.
引用
收藏
页码:371 / 379
页数:9
相关论文
共 50 条
  • [1] The correlation between the coaptation height of mitral valve and mitral regurgitation after mitral valve repair
    Wei, Dan
    Han, Jie
    Zhang, Haibo
    Li, Yan
    Xu, Chunlei
    Meng, Xu
    JOURNAL OF CARDIOTHORACIC SURGERY, 2017, 12
  • [2] The correlation between the coaptation height of mitral valve and mitral regurgitation after mitral valve repair
    Dan Wei
    Jie Han
    Haibo Zhang
    Yan Li
    Chunlei Xu
    Xu Meng
    Journal of Cardiothoracic Surgery, 12
  • [3] Improved mitral valve coaptation and reduced mitral valve annular size after percutaneous mitral valve repair (PMVR) using the MitraClip system
    Patzelt, Johannes
    Zhang, Yingying
    Magunia, Harry
    Ulrich, Miriam
    Jorbenadze, Rezo
    Droppa, Michal
    Zhang, Wenzhong
    Lausberg, Henning
    Walker, Tobias
    Rosenberger, Peter
    Seizer, Peter
    Gawaz, Meinrad
    Langer, Harald F.
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2018, 19 (07) : 785 - 791
  • [4] Echocardiographic Assessment of the Mitral Valve for Suitability of Repair: An Intraoperative Approach From a Mitral Center
    Mahmood, Feroze
    Sharkey, Aidan
    Maslow, Andrew
    Mufarrih, Syed Hamza
    Qureshi, Nada Qaisar
    Matyal, Robina
    Khabbaz, Kamal R.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2022, 36 (07) : 2164 - 2176
  • [5] Adjusting the neochords height for better leaflets coaptation following mitral valve repair
    Alameddine, Abdallah K.
    DiBiasio-White, Eric
    Binnall, Brian
    Conlin, Frederick
    JOURNAL OF CARDIAC SURGERY, 2019, 34 (07) : 610 - 613
  • [6] MitraClip® catheter-based mitral valve repair system
    Jonssonn, Anders
    Settergren, Magnus
    EXPERT REVIEW OF MEDICAL DEVICES, 2010, 7 (04) : 439 - 447
  • [7] Percutaneous increase of mitral leaflet coaptation length after mitral valve repair: results from a preclinical study
    Tozzi, Piergiorgio
    Siniscalchi, Giuseppe
    Ferrari, Enrico
    Kirsch, Matthias
    Hayoz, Daniel
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (04) : 681 - 686
  • [8] Intraoperative transesophageal echocardiography following mitral valve repair: a systematic review
    Zamper, Raffael
    Prempeh, Agya
    Iglesias, Ivan
    Fayad, Ashraf
    BRAZILIAN JOURNAL OF ANESTHESIOLOGY, 2022, 72 (03): : 379 - 397
  • [9] Assessment of the mitral valve coaptation zone with 2D and 3D transesophageal echocardiography before and after mitral valve repair
    Guo, Yong
    He, Yihua
    Zhang, Ye
    Ge, Shuping
    Sun, Lin
    Liu, Wenxu
    Han, Jiancheng
    Gu, Xiaoyan
    JOURNAL OF THORACIC DISEASE, 2018, 10 (01) : 283 - 290
  • [10] Effects of Mechanical Ventilation on Heart Geometry and Mitral Valve Leaflet Coaptation During Percutaneous Edge-to-Edge Mitral Valve Repair
    Patzelt, Johannes
    Zhang, Yingying
    Seizer, Peter
    Magunia, Harry
    Henning, Andreas
    Riemlova, Veronika
    Patzelt, Tara A. E.
    Hansen, Marc
    Haap, Michael
    Riessen, Reimer
    Lausberg, Henning
    Walker, Tobias
    Reutershan, Joerg
    Schlensak, Christian
    Grasshoff, Christian
    Simon, Dan I.
    Rosenberger, Peter
    Schreieck, Juergen
    Gawaz, Meinrad
    Langer, Harald F.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2016, 9 (02) : 151 - 159