Early Blood Clot Detection Using Forward Scattering Light Measurements Is Not Superior to Delta Pressure Measurements

被引:0
|
作者
Fischbach, Anna [1 ]
Lamberti, Michael [2 ]
Simons, Julia Alexandra [1 ]
Wrede, Erik [2 ]
Theissen, Alexander [1 ]
Winnersbach, Patrick [1 ]
Rossaint, Rolf [1 ]
Stollenwerk, Andre [2 ]
Bleilevens, Christian [1 ]
机构
[1] Univ Hosp RWTH Aachen, Dept Anesthesiol, Pauwelsstr 30, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen, Informat 11 Embedded Software, D-52074 Aachen, Germany
来源
BIOSENSORS-BASEL | 2023年 / 13卷 / 12期
关键词
forward scattered light measurement; sensor; membrane oxygenator; blood clot detection; in vitro test; EXTRACORPOREAL MEMBRANE-OXYGENATION; BETA-THROMBOGLOBULIN; RESPIRATORY-FAILURE; THROMBUS FORMATION; LIFE-SUPPORT; ADULTS;
D O I
10.3390/bios13121012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The occurrence of thrombus formation within an extracorporeal membrane oxygenator is a common complication during extracorporeal membrane oxygenation therapy and can rapidly result in a life-threatening situation due to arterial thromboembolism, causing stroke, pulmonary embolism, and limb ischemia in the patient. The standard clinical practice is to monitor the pressure at the inlet and outlet of oxygenators, indicating fulminant, obstructive clot formation indicated by an increasing pressure difference (Delta P). However, smaller blood clots at early stages are not detectable. Therefore, there is an unmet need for sensors that can detect blood clots at an early stage to minimize the associated thromboembolic risks for patients. This study aimed to evaluate if forward scattered light (FSL) measurements can be used for early blood clot detection and if it is superior to the current clinical gold standard (pressure measurements). A miniaturized in vitro test circuit, including a custom-made test chamber, was used. Heparinized human whole blood was circulated through the test circuit until clot formation occurred. Four LEDs and four photodiodes were placed along the sidewall of the test chamber in different positions for FSL measurements. The pressure monitor was connected to the inlet and the outlet to detect changes in Delta P across the test chamber. Despite several modifications in the LED positions on the test chamber, the FSL measurements could not reliably detect a blood clot within the in vitro test circuit, although the pressure measurements used as the current clinical gold standard detected fulminant clot formation in 11 independent experiments.
引用
收藏
页数:13
相关论文
共 4 条
  • [1] Blood Pressure Drop Prediction by using HRV Measurements in Orthostatic Hypotension
    Sannino, Giovanna
    Melillo, Paolo
    Stranges, Saverio
    De Pietro, Giuseppe
    Pecchia, Leandro
    JOURNAL OF MEDICAL SYSTEMS, 2015, 39 (11)
  • [2] Algorithm for diagnosing hypertension using out-of-office blood pressure measurements
    Kim, Je Sang
    Rhee, Moo-Yong
    Kim, Chee Hae
    Kim, Yoo Ri
    Do, Ungjeong
    Kim, Ji-Hyun
    Kim, Young Kwon
    Lee, Hyun Jung
    Park, Jee Yeon
    Namgung, June
    Lee, Sung Yun
    Cho, Deok-Kyu
    Choi, Tae-Young
    Kim, Seok Yeon
    JOURNAL OF CLINICAL HYPERTENSION, 2021, 23 (11) : 1965 - 1974
  • [3] Effects of Gas and Steam Humidity on Particulate Matter Measurements Obtained Using Light-Scattering Sensors
    Kim, Hyunsik
    Kim, Jeonghwan
    Roh, Seungjun
    SENSORS, 2023, 23 (13)
  • [4] Comparison between central and ambulatory blood pressure measurements in early detection of end organ damage: a single-center prospective non-randomized controlled trial
    Fouad, Doaa A.
    Al Araby, Hosam Hassan
    Ashraf, Mohammad
    El-Kousy, Ahmed El-Sherif
    EGYPTIAN HEART JOURNAL, 2019, 71 (01)