Simultaneous subcutaneous implantable cardioverter-defibrillator and leadless pacemaker implantation for patients at high risk of infection: a retrospective case series report

被引:2
|
作者
Calvagna, Giuseppe M. [1 ]
Valsecchi, Sergio [2 ]
机构
[1] ARNAS Garibaldi Ctr, Dept Cardiol, Catania, Italy
[2] Boston Sci, Milan, Italy
关键词
Implantable defibrillator; Subcutaneous; Leadless pacemaker; Extraction; Infection; ELECTRONIC DEVICE INFECTION; EUROPEAN-SOCIETY; COMPLICATIONS; ASSOCIATION; MORTALITY; ELIGIBILITY; EXTRACTION; MANAGEMENT; THERAPY; HRS;
D O I
10.1007/s10840-023-01684-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThe subcutaneous implantable cardioverter defibrillator (S-ICD) and leadless pacemaker (LP) are alternative options for patients at high risk of infection requiring ICD and pacing therapy. In this analysis, we described the simultaneous implantation of S-ICD and LP in patients with high infectious risk.MethodsThe study cohort comprised patients referred to our institution for ICD implantation due to high-risk factors of infection.ResultsBetween 2018 and 2022, 13 patients were referred, including 11 with infected ICD and 2 for first ICD implantation in the presence of high-risk factors. In cases of infected ICD, successful extraction was performed using a mechanical dilatation technique. Reimplantation was delayed until resolution of infection with antibiotic therapy. The devices were implanted during a single procedure, with S-ICD implantation following LP placement for verification of sensing adequacy through surface ECG screening. Suitable vectors for sensing during inhibited and ventricular pacing were identified in all patients. Defibrillation testing was effective, and no issues with double counting or undersensing were observed. The postoperative period was uneventful, and during a median follow-up of 35 months, no complications or infections were reported. The median ventricular pacing percentage was 5%, and a single inappropriate shock episode due to myopotential interference was reported and resolved by reprogramming the sensing vector.ConclusionSimultaneous implantation of S-ICD and LP is feasible and safe in patients at high risk of infection requiring both ICD and pacing therapy. This combined approach provides an effective solution for these patients.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Dual Antiplatelet Therapy and Heparin "Bridging" Significantly Increase the Risk of Bleeding Complications After Pacemaker or Implantable Cardioverter-Defibrillator Device Implantation
    Tompkins, Christine
    Cheng, Alan
    Dalal, Darshan
    Brinker, Jeffrey A.
    Leng, Charles T.
    Marine, Joseph E.
    Nazarian, Saman
    Spragg, David D.
    Sinha, Sunil
    Halperin, Henry
    Tomaselli, Gordon F.
    Berger, Ronald D.
    Calkins, Hugh
    Henrikson, Charles A.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (21) : 2376 - 2382
  • [42] Defibrillation testing during implantation of the subcutaneous implantable cardioverter-defibrillator: a necessary standard or becoming redundant?
    W. van der Stuijt
    A. B. E. Quast
    R. E. Knops
    Netherlands Heart Journal, 2020, 28 : 122 - 127
  • [43] Defibrillation testing during implantation of the subcutaneous implantable cardioverter-defibrillator: a necessary standard or becoming redundant?
    van der Stuijt, W.
    Quast, A. B. E.
    Knops, R. E.
    NETHERLANDS HEART JOURNAL, 2020, 28 (SUPPL 1) : 122 - 127
  • [44] Are We Improving Care of Medicare Patients Undergoing Primary Prevention Implantable Cardioverter-Defibrillator Implantation?
    Russo, Andrea M.
    CIRCULATION, 2014, 130 (10) : 808 - 810
  • [45] Unexpected Shocks From a Subcutaneous Implantable Cardioverter-Defibrillator Despite Attempted Reprogramming and Magnet Use: A Case Report
    McFaul, Colleen M.
    Lombaard, Stefan
    Arora, Vivek
    Van Cleve, William C.
    Rooke, G. Alec
    Prutkin, Jordan M.
    A & A PRACTICE, 2020, 14 (06): : e01178
  • [46] Simultaneous transvenous lead extraction and implantation of leadless cardiac pacemaker: A case report
    Wu, Mengwei
    Su, Hao
    CLINICAL CASE REPORTS, 2023, 11 (06):
  • [47] Safety and Performance of the Subcutaneous Implantable Cardioverter Defibrillator Detection Algorithm INSIGHT™ in Pacemaker Patients
    Weipert, Kay F.
    Kostic, Srdjan
    Goekyildirim, Timur
    Johnson, Victoria
    Chasan, Ritvan
    Gemein, Christopher
    Rosenbauer, Josef
    Erkapic, Damir
    Schmitt, Joern
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (01)
  • [48] Leadless Pacemaker Infection Risk in Patients with Cardiac Implantable Electronic Device Infections: A Case Series and Literature Review
    Bertolino, Lorenzo
    Florio, Letizia Lucia
    Patauner, Fabian
    Gallo, Raffaella
    Peluso, Anna Maria
    Scafuri, Antonio
    De Vivo, Stefano
    Corrado, Carmelina
    Zampino, Rosa
    Ruocco, Giuseppe
    D'Onofrio, Antonio
    Durante-Mangoni, Emanuele
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (24)
  • [49] Sudden Cardiac Death Risk over Time in HCM Patients with Implantable Cardioverter-Defibrillator
    Klopotowski, Mariusz
    Kukula, Krzysztof
    Jamiolkowski, Jacek
    Oreziak, Artur
    Dabrowski, Maciej
    Chmielak, Zbigniew
    Witkowski, Adam
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (06)
  • [50] High rate of subcutaneous implantable cardioverter-defibrillator sensing screening failure in patients with Brugada syndrome: a comparison with other inherited primary arrhythmia syndromes
    Conte, Giulio
    Kawabata, Mihoko
    de Asmundis, Carlo
    Taravelli, Erika
    Petracca, Francesco
    Ruggiero, Diego
    Caputo, Maria Luce
    Regoli, Francois
    Chierchia, Gian-Battista
    Chiodini, Alessandra
    Del Bufalo, Alessandro
    Moccetti, Tiziano
    Goya, Masahiko
    Hirao, Kenzo
    Vicentini, Alessandro
    De Ferrari, Gaetano M.
    Brugada, Pedro
    Auricchio, Angelo
    EUROPACE, 2018, 20 (07): : 1188 - 1193