Simultaneous subcutaneous implantable cardioverter-defibrillator and leadless pacemaker implantation for patients at high risk of infection: a retrospective case series report

被引:2
|
作者
Calvagna, Giuseppe M. [1 ]
Valsecchi, Sergio [2 ]
机构
[1] ARNAS Garibaldi Ctr, Dept Cardiol, Catania, Italy
[2] Boston Sci, Milan, Italy
关键词
Implantable defibrillator; Subcutaneous; Leadless pacemaker; Extraction; Infection; ELECTRONIC DEVICE INFECTION; EUROPEAN-SOCIETY; COMPLICATIONS; ASSOCIATION; MORTALITY; ELIGIBILITY; EXTRACTION; MANAGEMENT; THERAPY; HRS;
D O I
10.1007/s10840-023-01684-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundThe subcutaneous implantable cardioverter defibrillator (S-ICD) and leadless pacemaker (LP) are alternative options for patients at high risk of infection requiring ICD and pacing therapy. In this analysis, we described the simultaneous implantation of S-ICD and LP in patients with high infectious risk.MethodsThe study cohort comprised patients referred to our institution for ICD implantation due to high-risk factors of infection.ResultsBetween 2018 and 2022, 13 patients were referred, including 11 with infected ICD and 2 for first ICD implantation in the presence of high-risk factors. In cases of infected ICD, successful extraction was performed using a mechanical dilatation technique. Reimplantation was delayed until resolution of infection with antibiotic therapy. The devices were implanted during a single procedure, with S-ICD implantation following LP placement for verification of sensing adequacy through surface ECG screening. Suitable vectors for sensing during inhibited and ventricular pacing were identified in all patients. Defibrillation testing was effective, and no issues with double counting or undersensing were observed. The postoperative period was uneventful, and during a median follow-up of 35 months, no complications or infections were reported. The median ventricular pacing percentage was 5%, and a single inappropriate shock episode due to myopotential interference was reported and resolved by reprogramming the sensing vector.ConclusionSimultaneous implantation of S-ICD and LP is feasible and safe in patients at high risk of infection requiring both ICD and pacing therapy. This combined approach provides an effective solution for these patients.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Simultaneous leadless pacemaker and subcutaneous implantable cardioverter-defibrillator implantation-When vascular options have run out
    Ng, Jit Beng
    Chua, Kelvin
    Teo, Wee Siong
    JOURNAL OF ARRHYTHMIA, 2019, 35 (01) : 136 - 138
  • [2] Combined leadless pacemaker and subcutaneous implantable cardioverter-defibrillator to manage recurrent transvenous system failures
    Baroni, Matteo
    Colombo, Giulia
    Testoni, Alessio
    Arupi, Michele
    Lunati, Maurizio
    Cattafi, Giuseppe
    JOURNAL OF ELECTROCARDIOLOGY, 2019, 54 : 43 - 46
  • [3] Subserratus implantation of the subcutaneous implantable cardioverter-defibrillator
    Smietana, Jeffrey
    Frankel, David S.
    Serletti, Joseph M.
    Arkles, Jeffrey
    Pothineni, Naga Venkata K.
    Marchlinski, Francis E.
    Schaller, Robert D.
    HEART RHYTHM, 2021, 18 (10) : 1799 - 1804
  • [4] Patients with Pacemaker or Implantable Cardioverter-Defibrillator
    Schulman, Peter M.
    Rozner, Marc A.
    Sera, Valerie
    Stecker, Eric C.
    MEDICAL CLINICS OF NORTH AMERICA, 2013, 97 (06) : 1051 - +
  • [5] Device orientation of a leadless pacemaker and subcutaneous implantable cardioverter-defibrillator in canine and human subjects and the effect on intrabody communication
    Quast, Anne-Floor B. E.
    Tjong, Fleur V. Y.
    Koop, Brendan E.
    Wilde, Arthur A. M.
    Knops, Reinoud E.
    Burke, Martin C.
    EUROPACE, 2018, 20 (11): : 1866 - 1871
  • [6] Lead Abandonment and Subcutaneous Implantable Cardioverter-Defibrillator (S-ICD) Implantation in a Cohort of Patients With ICD Lead Malfunction
    Russo, Vincenzo
    Viani, Stefano
    Migliore, Federico
    Nigro, Gerardo
    Biffi, Mauro
    Tola, Gianfranco
    Bisignani, Giovanni
    Dello Russo, Antonio
    Sartori, Paolo
    Rordorf, Roberto
    Ottaviano, Luca
    Perego, Giovanni Battista
    Checchi, Luca
    Segreti, Luca
    Bertaglia, Emanuele
    Lovecchio, Mariolina
    Valsecchi, Sergio
    Bongiorni, Maria Grazia
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2021, 8
  • [7] Eligibility for the Subcutaneous Implantable Cardioverter-Defibrillator in Patients With Hypertrophic Cardiomyopathy
    Francia, Pietro
    Adduci, Carmen
    Palano, Francesca
    Semprini, Lorenzo
    Serdoz, Andrea
    Montesanti, Dalma
    Santini, Daria
    Musumeci, Beatrice
    Salvati, Adriano
    Volpe, Massimo
    Autore, Camillo
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (08) : 893 - 899
  • [8] First case of subcutaneous implantable cardioverter-defibrillator extrusion
    Santarpia, Giuseppe
    Polimeni, Alberto
    De Rosa, Salvatore
    Sabatino, Jolanda
    Curcio, Antonio
    Indolfi, Ciro
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2015, 192 : 19 - 20
  • [9] Antithrombotic therapy and the risk of pocket hematoma after subcutaneous implantable cardioverter-defibrillator implantation
    Pepplinkhuizen, S.
    Kors, N.
    de Veld, J. A.
    Dijkshoorn, L. A.
    Bijsterveld, N. R.
    de Weger, A.
    Smeding, L.
    Wilde, A. A. M.
    Nordkamp, L. R. A. Olde
    Knops, R. E.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2025,
  • [10] Infection in patients with subcutaneous implantable cardioverter-defibrillator: Results of the S-ICD Post Approval Study
    Gold, Michael R.
    Aasbo, Johan D.
    Weiss, Raul
    Burke, Martin C.
    Gleva, Marye J.
    Knight, Bradley P.
    Miller, Marc A.
    Schuger, Claudio D.
    Carter, Nathan
    Leigh, Jill
    Brisben, Amy J.
    El-Chami, Mikhael F.
    HEART RHYTHM, 2022, 19 (12) : 1993 - 2001