Development and Validation of Liquid Chromatography-Tandem Mass Spectrometric Method for the Quantification of Diclofenac in Human Plasma

被引:0
|
作者
Grover, Parul [1 ]
Bhardwaj, Monika [2 ]
Mehta, L. [3 ]
Kumar, Sandeep [4 ]
机构
[1] Delhi NCR, KIET Grp Inst, KIET Sch Pharm, Ghaziabad 201206, Uttar Pradesh, India
[2] Indian Inst Integrat Med, Nat Prod Chem Div, Canal Rd, Jammu 180001, Jammu And Kashm, India
[3] Amity Univ, Amity Inst Pharm, Noida 201301, Uttar Pradesh, India
[4] Natl Inst Pharmaceut Educ & Res, Dept Regulatory Affairs, Hyderabad 500037, Telangana, India
关键词
Diclofenac; validation; liquid chromatography-tandem mass spectrophotometry; human; plasma; multiple reaction monitoring; pharmacokinetics; HPLC METHOD; PERFORMANCE; SODIUM;
D O I
暂无
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
According to Food and Drug Administration requirements, a simple, sensitive and specific liquid chromatography-tandem mass spectrophotometry method for quantifying diclofenac in human plasma was developed and fully validated. Electrospray ionisation in the positive mode produced protonated ions, which were utilised to detect analyte and fluconazole (internal standard). Protein precipitation generated by acetonitrile was used to extract the analytes, followed by liquid-liquid extraction with ethyl acetate. The reversed phase separation was performed on a Chromolith speed ROD RP-18e (4.6 mmx50 mm) column with a simple isocratic mobile phase of 10 mM ammonium acetate in water, pH adjusted to 4.5 using acetic acid:acetonitrile in the ratio of (20:80, v/v) at a flow rate of 0.5 ml min-1. On a triple quadrupole mass spectrometer, fragmentation of Diclofenac m/z 296.10- (parent) and 250 (product) and Fluconazole m/z 307.20- (parent) and 220 (product) for internal standard was monitored. The devised method was verified in human plasma spanning a concentration range of 18.75 to 2000.25 ng ml-1, with a correlation coefficient (r2) of 0.9948. The detection was carried out using an electrospray ionisation approach on a triple quadrupole mass spectrometer that was used for multiple reaction monitoring. The intraday precision and accuracy values obtained from six different sets of quality control samples evaluated on different occasions ranged from 96.22-113.46 % and 2.66-9.95 %, respectively. For the analyte and internal standard, the overall recoveries were 61.98 3.97 and 55.01 1.18, respectively. The described approach was used to analyse diclofenac in human plasma samples with great performance.
引用
收藏
页码:740 / 752
页数:13
相关论文
共 50 条
  • [31] Liquid chromatography-tandem mass spectrometric assay for diclofenac and three primary metabolites in mouse plasma
    Sparidans, Rolf W.
    Lagas, Jurjen S.
    Schinkel, Afred H.
    Schellens, Jan H. M.
    Beijnen, Jos H.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 872 (1-2): : 77 - 82
  • [32] Quantification of micafungin in human plasma by liquid chromatography-tandem mass spectrometry
    Cangemi, Giuliana
    Barco, Sebastiano
    Bandettini, Roberto
    Castagnola, Elio
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (06) : 1795 - 1798
  • [33] Development and validation of a hydrophilic interaction liquid chromatography-tandem mass spectrometry method for the quantification of regadenoson in human plasma and its pharmacokinetic application
    Wang, Dun-Jian
    Wang, Da-Wei
    Fang, Qiu-Chen
    Shen, Ye
    Zeng, Nv-Jin
    Yang, Yan-Ling
    Zhang, Hong-Wen
    Wang, Yong-Qing
    Sun, Lu-Ning
    JOURNAL OF SEPARATION SCIENCE, 2022, 45 (06) : 1146 - 1152
  • [34] An ultra high performance liquid chromatography-tandem mass spectrometry method for the quantification of linagliptin in human plasma
    Nannapaneni, Nagaraj Kumar
    Jalalpure, Sunil S.
    Muppavarapu, Rajendraprasad
    Sirigiri, Sunil Kumar
    RSC ADVANCES, 2016, 6 (71): : 66756 - 66766
  • [35] Development and validation of a high-performance liquid chromatography-tandem mass spectrometric method for simultaneous determination of bupropion, quetiapine and escitalopram in human plasma
    Park, Semin
    Park, Chul-Soo
    Lee, Sung Joong
    Cha, Boseok
    Cho, Young Ah
    Song, Yi
    Yu, Eun Ae
    Kim, Gon-Sup
    Jin, Jong Sung
    Abd El-Aty, A. M.
    El-Banna, H. A.
    Hacimuftuoglu, Ahmet
    Shim, Jae-Han
    Shin, Sung Chul
    BIOMEDICAL CHROMATOGRAPHY, 2015, 29 (04) : 612 - 618
  • [36] A simple and accurate liquid chromatography-tandem mass spectrometry method for quantification of zonisamide in plasma and its application to a pharmacokinetic study
    Matar, Kamal M.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 961 : 103 - 109
  • [37] Liquid chromatography-tandem mass spectrometry method for simultaneous quantification of four triazole antifungal agents in human plasma
    Verdier, Marie-Clemence
    Bentue-Ferrer, Daniele
    Tribut, Olivier
    Bellissant, Eric
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (10) : 1515 - 1522
  • [38] Rapid ultraperformance liquid chromatography-tandem mass spectrometry method for quantification of oxcarbazepine and its metabolite in human plasma
    Bhatt, Mitesh
    Shaha, Sanjay
    Shivprakash
    BIOMEDICAL CHROMATOGRAPHY, 2011, 25 (07) : 751 - 759
  • [39] Validation of a liquid chromatography-tandem mass spectrometric method for the quantification of eight quinolones in bovine muscle, milk and aquacultured products
    van Hoop, N
    de Wasch, K
    Okerman, L
    Reybroeck, W
    Poelmans, S
    Noppe, H
    de Brabander, H
    ANALYTICA CHIMICA ACTA, 2005, 529 (1-2) : 265 - 272
  • [40] A Liquid Chromatography-Tandem Mass Spectrometry Bioanalytical Method Development and Validation for the Determination of Pacritinib in Plasma
    Kumar, Mulamraju Aruna
    Bangaraiah, Pagala
    SEPARATION SCIENCE PLUS, 2024, 7 (12):