An alkaline ferrocyanide non-enzymatic electrochemical sensor for creatinine detection

被引:5
作者
Bajpai, Sonal [1 ]
Akien, Geoffrey R. [1 ]
Toghill, Kathryn E. [1 ]
机构
[1] Univ Lancaster, Dept Chem, Lancaster LA1 4YB, England
基金
英国工程与自然科学研究理事会;
关键词
Creatinine; Electrocatalysis; Ferricyanide; Screen printed carbon electrode; Chronic kidney disease; Non-enzymatic; OXIDATION; NANOCOMPOSITE; METHYLGUANIDINE; FERRICYANIDE; COMPLEXES; AMINES; OXIDE;
D O I
10.1016/j.elecom.2023.107624
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Creatinine is a useful metabolite for the evaluation of Chronic Kidney Disease (CKD). Creatinine concentrations in urine outside of the healthy range of 4 to 15 mM may indicate CKD and other kidney conditions. Early detection and monitoring can provide lifesaving intervention, and as such there is a need to develop rapid, cheap and selective electrochemical creatinine sensors. A simple non-enzymatic creatinine sensing platform enabled by the electrocatalytic response of the metabolite to ferricyanide in alkaline conditions is presented. Studies were made on unmodified glassy carbon (GCE) and screen printed carbon electrodes (SPCE), both giving sensitivity of 21 +/- 5 mu A mM-1 cm-2 and limit of detection of ca. 60 mu M.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Non-enzymatic electrochemical glucose sensor based on NiMoO4 nanorods
    Wang, Dandan
    Cai, Daoping
    Huang, Hui
    Liu, Bin
    Wang, Lingling
    Liu, Yuan
    Li, Han
    Wang, Yanrong
    Li, Qiuhong
    Wang, Taihong
    NANOTECHNOLOGY, 2015, 26 (14)
  • [22] Synthesis and electrochemical study of nanoporous palladium-cadmium networks for non-enzymatic glucose detection
    Ahmadalinezhad, Asieh
    Chatterjee, Sanghamitra
    Chen, Aicheng
    ELECTROCHIMICA ACTA, 2013, 112 : 927 - 932
  • [23] Polyaniline-ZnO-NiO Nanocomposite based Non-enzymatic Electrochemical Sensor for Malathion Detection
    Jikamo, Samuel Chufamo
    Habtemariam, Tesfaye Haile
    Dolla, Tarekegn Heliso
    ELECTROANALYSIS, 2023, 35 (04)
  • [24] Innovative Non-Enzymatic Electrochemical Quantification of Cholesterol
    Carp, Oana Elena
    Pinteala, Mariana
    Arvinte, Adina
    SENSORS, 2022, 22 (03)
  • [25] Advances in electrochemical cofactor regeneration: enzymatic and non-enzymatic approaches
    Lee, Yoo Seok
    Gerulskis, Rokas
    Minteer, Shelley D.
    CURRENT OPINION IN BIOTECHNOLOGY, 2022, 73 : 14 - 21
  • [26] Design of a mediator-free, non-enzymatic electrochemical biosensor for glutamate detection
    Zeynaloo, Elnaz
    Yang, Yu-Ping
    Dikici, Emre
    Landgraf, Ralf
    Bachas, Leonidas G.
    Daunert, Sylvia
    NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2021, 31
  • [27] Non-enzymatic electrochemical determination of creatinine using a novel screen-printed microcell
    Fava, Elson Luiz
    do Prado, Thiago Martimiano
    Garcia-Filho, Amauri
    Silva, Tiago Almeida
    Cincotto, Fernando Henrique
    de Moraes, Fernando Cruz
    Faria, Ronaldo Censi
    Fatibello-Filho, Orlando
    TALANTA, 2020, 207
  • [28] Evaluation of a new electrochemical sensor for selective detection of non-enzymatic hydrogen peroxide based on hierarchical nanostructures of zirconium molybdate
    Kumar, J. Vinoth
    Karthik, R.
    Chen, Shen-Ming
    Raja, N.
    Selvam, V.
    Muthuraj, V.
    JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2017, 500 : 44 - 53
  • [29] Magnetic nanofluid based non-enzymatic sensor for urea detection
    Zaibudeen, A. W.
    Philip, John
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 255 : 720 - 728
  • [30] A new non-enzymatic sensor based on TiO2-Ag/polypyrrole for electrochemical detection of tyramine
    Erdogan, Zehra O.
    Akin, Ilker
    Kucukkolbasi, Semahat