Evaluation of full thread screw and different fixation configurations in Pauwels type III femoral neck fracture

被引:0
|
作者
Qiu, Jianjun [1 ]
Zou, Hanlin [1 ]
Zhang, Lei [1 ,2 ]
Zhou, Xu [1 ]
机构
[1] Shanghai Univ Tradit Chinese Med, Putuo Hosp, Dept Neurol, Shanghai 200062, Peoples R China
[2] Shanghai Univ Tradit Chinese Med, Putuo Hosp, 164 Lanxi Rd, Shanghai 200062, Peoples R China
关键词
Full thread screw; Different fixation configurations; Femoral neck fracture; Model rigidity; Axial displacement; Limit loads; CANNULATED CANCELLOUS SCREWS; SLIDING HIP SCREW; ADULTS; WEITBRECHT; RETINACULA; SYSTEM; PLATE;
D O I
10.1016/j.heliyon.2023.e21857
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives: The purpose of this study is to evaluate the value of full thread screw and different fixation configurations in Pauwels type III femoral neck fracture.Methods: A total of 40 artificial femoral model specimens were chosen, and Pauwels type III femoral neck fracture was simulated upon osteotomy at 80. According to random number table, models were divided into four groups (10 cases in each group): Group A received the paralleled fixation with three partial thread screws (PTSs), group B received the crossed fixation with three PTSs, group C received the paralleled fixation with two full thread screws (FTSs) and one PTS, and group D received the crossed fixation with two FTSs and one PTS. Changes including the model rigidity, axial displacement in fatigue test and limit loads for Pauwels type III femoral neck fracture models were analyzed through MTS 858 Mini Bionix II test system.Results: Among four groups, the model rigidity, axial displacement in fatigue test and limit loads were the highest in group D, and they were the lowest in group A. However, the model rigidity, axial displacement in fatigue test and limit loads between group B and group C showed no sta-tistically significant difference (P>0.05). Eventually, all the specimens were displaced along the fracture lines while the femoral head was split at varying degrees. After splits, the removal rate of fixation screws in group A (60.0 %) and group C (40.0 %) was significantly higher than that of group B (10.0 %) and group D (0 %) (P<0.05), but it showed no statistically significant difference between group A and group C, and between group B and group D (P>0.05).Conclusions: The crossed fixation configuration with two FTSs and one PTS in group D is proven to be more effective, which can go against the shear stress, tension and introversion in Pauwels type III femoral neck fracture models.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Finite element analysis of a new plate for Pauwels type III femoral neck fractures
    Wang, Gang
    Tang, Yong
    Wu, Xuhua
    Yang, Huilin
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2020, 48 (02)
  • [42] Comparison of three different internal fixation methods in the treatment of femoral neck fracture
    Zhang, Xinzhao
    Zheng, Changling
    Huang, Jin
    Chen, Hui
    Lei, Jie
    Huang, Cong
    HELIYON, 2024, 10 (14)
  • [43] Biomechanical evaluation of the expansive cannulated screw for fixation of femoral neck fractures
    Zhang, Yang
    Tian, Li
    Yan, Yabo
    Sang, Hongxun
    Ma, Zhensheng
    Jie, Qiang
    Lei, Wei
    Wu, Zixiang
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2011, 42 (11): : 1372 - 1376
  • [44] The effect of anteromedial support plate with three cannulated screws in the treatment of Pauwels type III femoral neck fracture in young adults
    Lin, Shanghui
    Li, Jinmei
    Wang, Renkai
    Ou, Yongliang
    Jia, Zhenyu
    Zhang, Ying
    Xia, Hong
    Li, Baofeng
    Chen, Bei
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2022, 48 (05) : 4011 - 4017
  • [45] Internal Fixation of Garden Type III Femoral Neck Fractures with Sliding Hip Screw and Anti-Rotation Screw: Does Increased Valgus Improve Healing?
    Hackl, Simon
    von Rueden, Christian
    Weisemann, Ferdinand
    Kloepfer-Kraemer, Isabella
    Stuby, Fabian M.
    Hoegel, Florian
    MEDICINA-LITHUANIA, 2022, 58 (11):
  • [46] MECHANICAL ANALYSIS OF FEMORAL NECK FRACTURE FIXATION WITH DYNAMIC CONDYLAR SCREW IN SYNTHETIC BONE
    Freitas, Anderson
    Maciel, Rafael Almeida
    Lima, Renato de Almeida
    de Macedo Souto, Diogo Ranier
    Ferrer, Marcelo de Almeida
    ACTA ORTOPEDICA BRASILEIRA, 2014, 22 (05): : 264 - 268
  • [47] The role of the medial plate for Pauwels type III femoral neck fracture: a comparative mechanical study using two fixations with cannulated screws
    Giordano, Vincenzo
    Alves, Danilo Diniz
    Paes, Roger Pletsch
    Amaral, Arthur Bonfim
    Giordano, Marcos
    Belangero, William
    Freitas, Anderson
    Koch, Hilton A.
    do Amaral, Ney Pecegueiro
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2019, 6 (01)
  • [48] Global trends of Pauwels type III femoral neck fractures: bibliometric analysis and visualized study
    Wang, Mengyu
    Zhang, Ze
    Sun, Fengpo
    Zhu, Yi
    Han, Ruining
    Pei, Zijie
    He, Zhaoshuo
    Liu, Junzhi
    Wen, Liangyuan
    FRONTIERS IN SURGERY, 2024, 11
  • [49] The role of the medial plate for Pauwels type III femoral neck fracture: a comparative mechanical study using two fixations with cannulated screws
    Vincenzo Giordano
    Danilo Diniz Alves
    Roger Pletsch Paes
    Arthur Bonfim Amaral
    Marcos Giordano
    William Belangero
    Anderson Freitas
    Hilton A. Koch
    Ney Pecegueiro do Amaral
    Journal of Experimental Orthopaedics, 6
  • [50] Biomechanical comparison of distal femoral fracture fixation: Polyaxial vs. uniaxial locking screws with different blocking screw configurations in retrograde femoral nailing
    Batar, Suat
    Soylemez, Mehmet Salih
    Bozdag, Sureyya Ergun
    Yaman, Enis
    Kemah, Bahattin
    Poyanli, Oguz Sukru
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2024, 110 (01)