Evaluation of full thread screw and different fixation configurations in Pauwels type III femoral neck fracture

被引:0
|
作者
Qiu, Jianjun [1 ]
Zou, Hanlin [1 ]
Zhang, Lei [1 ,2 ]
Zhou, Xu [1 ]
机构
[1] Shanghai Univ Tradit Chinese Med, Putuo Hosp, Dept Neurol, Shanghai 200062, Peoples R China
[2] Shanghai Univ Tradit Chinese Med, Putuo Hosp, 164 Lanxi Rd, Shanghai 200062, Peoples R China
关键词
Full thread screw; Different fixation configurations; Femoral neck fracture; Model rigidity; Axial displacement; Limit loads; CANNULATED CANCELLOUS SCREWS; SLIDING HIP SCREW; ADULTS; WEITBRECHT; RETINACULA; SYSTEM; PLATE;
D O I
10.1016/j.heliyon.2023.e21857
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives: The purpose of this study is to evaluate the value of full thread screw and different fixation configurations in Pauwels type III femoral neck fracture.Methods: A total of 40 artificial femoral model specimens were chosen, and Pauwels type III femoral neck fracture was simulated upon osteotomy at 80. According to random number table, models were divided into four groups (10 cases in each group): Group A received the paralleled fixation with three partial thread screws (PTSs), group B received the crossed fixation with three PTSs, group C received the paralleled fixation with two full thread screws (FTSs) and one PTS, and group D received the crossed fixation with two FTSs and one PTS. Changes including the model rigidity, axial displacement in fatigue test and limit loads for Pauwels type III femoral neck fracture models were analyzed through MTS 858 Mini Bionix II test system.Results: Among four groups, the model rigidity, axial displacement in fatigue test and limit loads were the highest in group D, and they were the lowest in group A. However, the model rigidity, axial displacement in fatigue test and limit loads between group B and group C showed no sta-tistically significant difference (P>0.05). Eventually, all the specimens were displaced along the fracture lines while the femoral head was split at varying degrees. After splits, the removal rate of fixation screws in group A (60.0 %) and group C (40.0 %) was significantly higher than that of group B (10.0 %) and group D (0 %) (P<0.05), but it showed no statistically significant difference between group A and group C, and between group B and group D (P>0.05).Conclusions: The crossed fixation configuration with two FTSs and one PTS in group D is proven to be more effective, which can go against the shear stress, tension and introversion in Pauwels type III femoral neck fracture models.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Biomechanical evaluation of different internal fixation methods based on finite element analysis for Pauwels type III femoral neck fracture
    Jiang, Xianbao
    Liang, Ke
    Du, Gang
    Chen, Yuan
    Tang, Yu
    Geng, Kuihua
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (10): : 3115 - 3123
  • [2] Medial femoral plate with cannulated screw for Pauwels type III femoral neck fracture: A meta-analysis
    Su, Zhengbing
    Liang, Linlin
    Hao, Yong
    JOURNAL OF BACK AND MUSCULOSKELETAL REHABILITATION, 2021, 34 (02) : 169 - 177
  • [3] Screw-blade fixation systems in Pauwels three femoral neck fractures: a biomechanical evaluation
    Knobe, Matthias
    Altgassen, Simon
    Maier, Klaus-Jurgen
    Gradl-Dietsch, Gertraud
    Kaczmarek, Chris
    Nebelung, Sven
    Klos, Kajetan
    Kim, Bong-Sung
    Gueorguiev, Boyko
    Horst, Klemens
    Buecking, Benjamin
    INTERNATIONAL ORTHOPAEDICS, 2018, 42 (02) : 409 - 418
  • [4] Finite element analysis of the Femoral neck system for different placement positions in the fixation of Pauwels type III femoral neck fractures
    Zhu, Yongjia
    Wu, Shuai
    Yan, Jiapeng
    Wang, Wenhong
    Huang, Xiao
    Zhang, Hongfei
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2025, 56 (04):
  • [5] Finite element analysis of different internal fixation methods for the treatment of Pauwels type III femoral neck fracture
    Lin Tianye
    Yang Peng
    Xu Jingli
    Wei QiuShi
    Zhou GuangQuan
    He Wei
    Zhang Qingwen
    BIOMEDICINE & PHARMACOTHERAPY, 2019, 112
  • [6] Modified cannulated screw fixation in the treatment of Pauwels type III femoral neck fractures: A biomechanical study
    Zhou, Lin
    Lin, Jiaming
    Huang, Anfei
    Gan, Weidong
    Zhai, Xu
    Sun, Kai
    Huang, Shiqing
    Li, Zhizhong
    CLINICAL BIOMECHANICS, 2020, 74 : 103 - 110
  • [7] Internal fixation system with the pyramid screw head for the treatment of Pauwels type III femoral neck fractures
    Wang, Songyuan
    Deng, Xuan
    Zhong, Hao
    Ma, Lei
    Wang, Xiyu
    Lou, Xinqi
    Fan, Xuanze
    Duan, Wangping
    Zhang, Kai
    Li, Pengcui
    Wei, Xiaochun
    Zhao, Yongwang
    Wang, Yanqin
    Wu, Xiaogang
    Xue, Yanru
    Chen, Weiyi
    ACTA MECHANICA SINICA, 2023, 39 (08)
  • [8] Screw-blade fixation systems in Pauwels three femoral neck fractures: a biomechanical evaluation
    Matthias Knobe
    Simon Altgassen
    Klaus-Jürgen Maier
    Gertraud Gradl-Dietsch
    Chris Kaczmarek
    Sven Nebelung
    Kajetan Klos
    Bong-Sung Kim
    Boyko Gueorguiev
    Klemens Horst
    Benjamin Buecking
    International Orthopaedics, 2018, 42 : 409 - 418
  • [9] Finite element analysis of femoral neck system in the treatment of Pauwels type III femoral neck fracture
    Teng, Yanbin
    Zhang, Yan
    Guo, Chaowei
    MEDICINE, 2022, 101 (28) : E29450
  • [10] Novel screw fixation placement configuration for the treatment of Pauwels type III femoral neck fractures: a finite element analysis
    Petra Adamović
    Janoš Kodvanj
    Dejan Blažević
    Srećko Sabalić
    Medical & Biological Engineering & Computing, 2023, 61 : 1005 - 1015