High time-resolution MRPC detector for TOF-PET

被引:2
|
作者
Liu, J. [1 ]
Wang, Y. [1 ]
Guo, B. [1 ]
Yu, Y. [1 ]
Han, D. [1 ]
Li, Y. [1 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Key Lab Particle & Radiat Imaging, Beijing 100084, Peoples R China
关键词
MRPC detector; TOF-PET; Time resolution; Detection efficiency;
D O I
10.1016/j.nima.2023.168453
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The traditional TOF-PET (Positron emission tomography) mainly uses the structure of a scintillator detector connected to a photomultiplier tube. The structure has the disadvantages that time resolution is not good enough (about 300 ps FWHM) for improving the signal-to-noise ratio of the acquisition system, which is not conducive to the detection of microtumors in the clinic and reduction in imaging time. The thickness of the scintillator is large (greater than 3 cm). Traditional PET has low geometric efficiency and sensitivity. It requires a long time to complete a full-body scan. However, the MRPC can be made into a 2.4-meter-long whole body PET. The high cost of traditional TOF-PET limits its widespread use in medical and other fields. The excellent time resolution of MRPC makes MRPC TOF-PET a good application prospect. This paper shows the time resolution of two 32-gap MRPCs with 128 & mu;m gap thickness on cosmic rays and 0.511 MeV photons. By using the fast front-end amplifier and waveform acquisition system, the time resolution for cosmic rays is 27 ps and the rms time accuracy per 0.511 MeV photon is 72 ps. The time resolution in terms of FWHM of the time difference between photon pairs is 239 ps FWHM. The reason why MRPC detectors have different time resolutions for cosmic rays and 0.511 MeV photons is related to the different ways different particles act and the thickness of the detector. We propose to use a Cerenkov radiator to convert gamma into ultraviolet photons and use a composite photocathode to detect, improving the detection efficiency to 6.4% for RPC.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] A very thin MRPC developed for TOF-PET
    Liu, J.
    Wang, Y.
    Guo, B.
    Han, D.
    Li, Y.
    JOURNAL OF INSTRUMENTATION, 2024, 19 (02):
  • [2] Evaluation of a TOF-PET detector design that achieves ≤100 ps coincidence time resolution
    Cates, J. W.
    Levin, C. S.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [3] Advantage of the Four-Layer DOI Information in the Time Resolution for a TOF-PET Detector
    Tsuda, Tomoaki
    Kitamura, Keishi
    Ohi, Junichi
    Tonami, Hiromichi
    Satoh, Masanobu
    Kumazawa, Yoshihiko
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3201 - 3204
  • [4] Evaluation of a clinical TOF-PET detector design that achieves ≤100 ps coincidence time resolution
    Cates, Joshua W.
    Levin, Craig S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (11):
  • [5] Timing resolution improved by DOI information in an LYSO TOF-PET detector
    Shibuya, Kengo
    Nishikido, Furnihiko
    Inadama, Naoko
    Yoshida, Eiji
    Lam, Chihfung
    Tsuda, Tomoaki
    Yamaya, Taiga
    Murayama, Hideo
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3678 - 3680
  • [6] Studies on the Cherenkov Effect for Improved Time Resolution of TOF-PET
    Brunner, S. E.
    Gruber, L.
    Marton, J.
    Suzuki, K.
    Hirtl, A.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (01) : 443 - 447
  • [7] Study of optical reflectors for a 100ps coincidence time resolution TOF-PET detector design
    Gonzalez-Montoro, Andrea
    Pourashraf, Shirin
    Lee, Min Sun
    Cates, Joshua W.
    Levin, Craig S.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (06)
  • [8] High time-resolution photodetectors for PET applications
    Ronzhin, Anatoly
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 809 : 53 - 57
  • [9] Simultaneous time-skew and time-walk correction for TOF-PET detector
    Yoshida, Eiji
    Obata, Fujino
    Yamaya, Taiga
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1049
  • [10] Sparse SiPM Pixel Arrangement for a TOF-PET Detector Design that Achieves 100 ps Coincidence Time Resolution
    Cates, J. W.
    Levin, C. S.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2023, 7 (07) : 665 - 672