Photon counting x-ray detectors as scatter probes

被引:1
作者
Lahoud, Elias [1 ,2 ]
Zaffrani, Yael [1 ]
Yagil, Yoad [1 ]
机构
[1] Philips Healthcare, Haifa, Israel
[2] Philips Med Syst Technol Ltd, Adv Technol Ctr, Bldg 34, IL-3100202 Haifa, Israel
关键词
photon counting; scatter; COMPUTED-TOMOGRAPHY; GENERAL FRAMEWORK; FORM-FACTORS; CT; FEASIBILITY; COMPENSATION;
D O I
10.1002/mp.16871
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundDirect conversion x-ray Photon Counting Detectors (PCD) are posed to play a vital role in future medical imaging devices such as Computed Tomography (CT) scanners. PCD are expected to improve current CT technology on several fronts, such as resolution, dose utilization, and spectral performance. However, they are not readily expected to improve the handling of object scatter, one of the major sources of image artifacts in CT technology.PurposeWe explore a potential method for obtaining in-situ object scatter estimation using the same PCD array used in the x-ray imaging system, such as in computed tomography. This unexpected benefit of using PCD has the potential to improve the image quality by providing better input into the scatter estimation and correction algorithms used in image reconstruction.MethodsIn CT scanners the primary method for rejecting scatter signal originating from the scanned object relies on placing anti-scatter grids (ASG) close to the detector plane. This remains the case when transitioning to using PCD instead of energy integration detectors in CT. However, the combination of PCD and ASG opens a possibility to use some of the unique properties of PCD, namely, very low noise and coincidence counters to obtain, in addition to the attenuation data, a simultaneous and instantaneous estimate of the scatter signal reaching every detector element. When a small air gap is introduced between the ASG and the detector surface, the scatter radiation with large angular distribution has a greater probability of producing charge sharing events that can be detected by a coincidence counter. In this work we demonstrate the feasibility of such an approach in a tabletop experiment using PCD detector that lacks coincidence counting capability, instead we use the spectral signature of split charge events as proxy to coincidence counting. For this purpose, we first demonstrate the spectral impact of ASG misalignment using the same experimental setup. In addition, we perform a separate tabletop scattering experiment from a narrow column of water that demonstrates another potential use of the low noise capabilities of PCDs.ResultsWe measured and quantified the high sensitivity of the spectral response to ASG alignment on the PCD detector pixel array, we found that the probability of energy misregistration of 60 keV photons can increase by up to a factor of 3 when the ASG is poorly aligned. We then leveraged these results to obtain an estimate on the expected increase in coincidence counts for a wide range of scatter-to-primary (SPR) ratio and find a good match with expectations from a geometric modeling of the system, where the expected increase in coincidences was of the order of the SPR. Finally, the low noise detector also allowed us to measure the real space scatter signal associated with the coherent molecular form factor of water, revealing the ring-shaped scatter signal with an energy dependent distribution that was well captured by calculation.ConclusionsThe advent of PCD detectors and their imminent use in commercial CT scanners opens new and exciting possibilities for utilizing PCD detectors in unexpected ways. In this proof-of-concept study, we showed how charge sharing, a spectral information degrading effect, can instead be used to obtain in-situ scatter estimation. We also demonstrated the PCD ability to perform extremely sensitive measurements using affordable benchtop setup for investigations normally reserved for synchrotron facilities.
引用
收藏
页码:93 / 102
页数:10
相关论文
共 30 条
  • [1] Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT
    Altunbas, Cem
    Kavanagh, Brian
    Alexeev, Timur
    Miften, Moyed
    [J]. MEDICAL PHYSICS, 2017, 44 (08) : 3952 - 3964
  • [2] Berger J., 2005, XCOM: Photon Cross Sections Database
  • [3] PERFORMANCE OF ANTISCATTER GRIDS IN DIAGNOSTIC-RADIOLOGY - EXPERIMENTAL MEASUREMENTS AND MONTE-CARLO SIMULATION STUDIES
    CHAN, HP
    HIGASHIDA, Y
    DOI, K
    [J]. MEDICAL PHYSICS, 1985, 12 (04) : 449 - 454
  • [4] Multicolor spectral photon-counting computed tomography: in vivo dual contrast imaging with a high count rate scanner
    Cormode, David P.
    Si-Mohamed, Salim
    Bar-Ness, Daniel
    Sigovan, Monica
    Naha, Pratap C.
    Balegamire, Joelle
    Lavenne, Franck
    Coulon, Philippe
    Roessl, Ewald
    Bartels, Matthias
    Rokni, Michal
    Blevis, Ira
    Boussel, Loic
    Douek, Philippe
    [J]. SCIENTIFIC REPORTS, 2017, 7
  • [5] Quantification of multiple mixed contrast and tissue compositions using photon-counting spectral computed tomography
    Curtis, Tyler E.
    Roeder, Ryan K.
    [J]. JOURNAL OF MEDICAL IMAGING, 2019, 6 (01)
  • [6] Photon-counting CT review
    Flohr, Thomas
    Petersilka, Martin
    Henning, Andre
    Ulzheimer, Stefan
    Ferda, Jiri
    Schmidt, Bernhard
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 : 126 - 136
  • [7] Coincidence Counters for Charge Sharing Compensation in Spectroscopic Photon Counting Detectors
    Hsieh, Scott S.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (03) : 678 - 687
  • [8] Knoll G.F., 1999, RAD DETECTION MEASUR
  • [9] Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology
    Leng, Shuai
    Bruesewitz, Michael
    Tao, Shengzhen
    Rajendran, Kishore
    Halaweish, Ahmed F.
    Campeau, Norbert G.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    [J]. RADIOGRAPHICS, 2019, 39 (03) : 729 - 743
  • [10] Spectral Signatures of X-ray Scatter Using Energy-Resolving Photon-Counting Detectors
    Lewis, Cale E.
    Das, Mini
    [J]. SENSORS, 2019, 19 (22)